Troubleshooting

Fault and Condition
1
Steady
2
Steady
3
Flashing
4
Flashing
The LED changed from green to red and the battery
symbol is blinking during treatment
5
Flashing
Lamira turns on but LED flashes green-red and LCD
display flashes this symbol
6
Flashing
Lamira beeps 1x, then beeps high-low with the LED
flashing and the LCD display flashing this symbol
7
Flashing
Lamira beeps low-high 3 times, the LED flashed green-
red, the LCD displays this symbol, and then turns off
8
Steady
Lamira beeps 1 time and LCD shows this symbol
9
Steady
10
Flashing
11
Brief
Reference ID: 4327567
17
Questions about Lamira? Contact the Arikares Support Program at 1-833-ARIKARE (1-833-274-5273)
Possible Cause and Solution
Controller running on AC Power.
Controller running on battery power.
Empty battery, replace batteries or switch to A/C power.
Low battery, replace batteries or switch to A/C power.
Bad or missing Connection Cord. Check Connection Cord between
Handset and Controller. Bad or missing Aerosol Head. Correct then
restart Controller.
No ARIKAYCE detected. Add ARIKAYCE. If you have already added
ARIKAYCE, gently tap the Handset, hold the Handset level and restart
the Controller.
Have reached 20 minute maximum time and will shut down. If
ARIKAYCE remains, restart the Controller. After treatment is
complete, clean and disinfect the Handset including the Aerosol Head.
Misting and working properly.
Controller has paused. To resume press On/Off button.
Press button to start treatment.
Functioning properly, the Controller has started properly.