  • Page 1 Rev. 14/11/17 R-030425 EN ......4...
  • Page 3 Tank Keypad Lid opening Pressure Gauge Water tap Water outlet Air inlet Air outlet Electrical connection Main switch...
  • Page 4: Installation

    Automatic polymerisation pot  NOTE: Read this manual carefully. It contains impor-  Avoid excessive traction, twisting or folding of the tant information about safety during installa- cable for connection to the electricity grid. tion, use and maintenance. If the equipment is ...
  • Page 5: Use Of The Keys

    Polymerisation programs There are three types of program: short, long and con- tinuous. Short programs: Consist of a consigned polymerisation temperature ºC1 and a maintenance time t ºC1. 230 V Long programs: Consist of a preheating temperature ºC1, Water outlet Air inlet a maintenance time t ºC1, a temperature rise slope r ºC2,...
  • Page 6: Operation

    2. Use keys to select one of the 5 pro- 5. Press the key again to modify the rise slope to grams (Pc, P1, P2, P3 or P4-L). the second temperature r ºC2. This is the continuous program (see in its sec- Time for temperature rise slope to the tion).
  • Page 7 6. When the assigned temperature is reached, a beep Manual pressurisation will sound and the machine will show the message If the polymerisation temperature is higher than 60 °C, LoAd LoAd. LoAd LoAd the user will have to carry out the pressurisation manual- Insert the material for polymerisation and close the lid of the pot.
  • Page 8 (there are safety valves inside).  Replace the rubber seal of the lid periodically. For this operation, we recommend using original MESTRA Error messages replacements.  If you feel excessive resistance when turning the lids, E-01 Broken interior thermal probe.
  • Page 9 Talleres Mestraitua S.L. Txori-erri Etorbidea, 60 T: (+34)944530388 - F: (+34)944711725 Your trusted brand mestra@mestra.es - www.mestra.es 48150 SONDIKA - BILBAO - ESPAÑA...