L672 Dementia Controller Menu Settings
The menu structure allows the unit configuration to be changed at any time. The Menu enable jumper must be fitted
to the rear of the unit to gain access to the following menus which proceed after the Profile choice.
Profile
SELECT PROFILE: 00
BED SENSOR DELAY
ROOM ALARM TYPE
X4 ALARM TYPE
X5 ALARM TYPE
ROOM USER ID
X4 USER ID
X5 USER ID
ALARM RELAY
AUTO OFF TIMER
BED SENSOR CPR
MIN ENSUITE TIME
AUTO RESET
RELAY 4
CALL POINT ID
BED OCCUPIED
ALARM TIMER
[Profile 17+18 only]
LIGHT TIMER
[Profile 17 only]
BED LIGHT OFF DELY
[Profile 18 only]
Intercall 600/700 Installation & Operation Guide Issue 4.59
Settings
Sixteen preset profiles which can be selected to suit the individual occupants normal
00 to 18
behaviour patterns. Please refer to the profile table for more information on specific
timings for each profile
OFF / 02 / 04 / 08
This setting allows a delay to be introduced from the bed sensor before the dementia
controller detects an out of bed condition. The range is none (00) 2,4 or 8 seconds.
(Seconds)
CALL / PRIORITY /
Selects the type of call generated by the room profile alarm. The menu structure
prevents the doors from being set to a lower call level than the room.
EMERGENCY
Selects the type of call generated by the X4 CORRIDOR DOOR contact & X5 PATIO
*CALL / *PRIORITY /
DOOR contact. *The menu structure prevents the doors from being set to a lower
EMERGENCY
alarm type than the room. If the Room Alarm Type is set to Call, the door alarm may
be set to Call, Priority or Emergency. If the Room Alarm Type is set to Priority, the door
*CALL / *PRIORITY /
alarm may be set to Priority or Emergency. If the Room Alarm Type is set to
EMERGENCY
Emergency, the door alarm can only be set to Emergency.
Selects the User ID generated by the room profile alarm. This can be used to generate
OFF – 254
a secondary address from the room for example "BEDROOM 10 DEMENTIA ALARM"
Selects the User ID generated by the X4 Door Contact input. This can be used to
OFF – 254
generate a secondary address from the room for example "BEDROOM 10 CORRIDOR
DOOR"
Selects the User ID generated by the X5 Door Contact input. This could be used to
OFF – 254
generate a secondary address from the room for example "BEDROOM 10 PATIO
DOOR"
The unit is fitted with an auxiliary relay to trigger third party alarm systems and pager
ON / OFF
systems. This is disabled by default.
The unit can automatically switch off after a number of hours. For example it is
OFF, 5- 12 Hours
switched on when the resident is put to bed but automatically switches off in the
morning. The range is between 5 and 12 hours.
When used in conjunction with later versions of the Intercall Call Management
ON / OFF
Software, the out of bed activity is silently monitored by the software. However, this
function is not compatible with earlier systems and must be disabled.
Selects the minimum length of time that the Ensuite PIR will need to be activated
00,02,04,08 Secs
before the light switch-over & cue timer is enabled. This allows alteration of the
ensuite/room detection performance.
Once the room alarm is activated, if the resident returns to be the bed the unit can be
ON / OFF
configured to automatically reset the room alarm. This function can be disabled so a
member of staff must visit to reset the room alarm at the Dementia Controller.
DOOR ONLY /ROOM ONLY
Relay 4 operates when part or all of the Dementia Controller is enabled as options on
/DOOR & ROOM/
left. This can be used to switch the light control automatically between normal light
switches (ie during the day) and the Dementia Controller (ie at night time)
DOOR OR ROOM
Selects the User ID generated by the X6 and X7 Slave Call Point. This can be used to
OFF – 254
generate a secondary address from the room for example "BEDROOM 10 CALL POINT"
Select the polarity of the X1 Bed Sensor input. Bed Mats are normally Contact Closed
CONTACT CLOSED
when the bed is occupied. The input can be set to Contact Open for different bed
CONTACT OPEN
sensor types.
If the PIR's are continuously activated for this period of time, the unit generates a
02 - 30 Minutes
Room Alarm. This setting is only available and only applies to Profile 17+18.
The Light Timer is the length of time the lights remain on after the last PIR activation is
10 – 100 Seconds
detected. This setting is only available and only applies to Profile 17.
The Bed Light Off Delay sets the length of time the lights remain on after the resident
0 – 120 Seconds
is back in the bed. This setting only applies to Profile 18.