Input - Endress+Hauser Proline Promass F 300 Operating Instructions Manual

Coriolis flowmeter ethernet / ip
Table of Contents

Advertisement

Proline Promass F 300 EtherNet/IP
Measured variable
Measuring range
16.3

Input

Direct measured variables
• Mass flow
• Density
• Temperature
Calculated measured variables
• Volume flow
• Corrected volume flow
• Reference density
Measuring ranges for liquids
DN
[mm]
8
15
25
40
50
80
100
150
250
Measuring ranges for gases
The full scale values depend on the density of the gas and can be calculated with the
formula below:
= 
· ρ
: x
max(G)
max(F)
G
Maximum full scale value for gas [kg/h]
max(G)
Maximum full scale value for liquid [kg/h]
max(F)
< 
max(G)
max(F)
ρ
Gas density in [kg/m³] at operating conditions
G
x
Constant dependent on nominal diameter
DN
[mm]
8
15
25
Measuring range full scale values 
[in]
³⁄₈
½
1
2
3
0 to 180 000
4
0 to 350 000
6
0 to 800 000
10
0 to 2 200 000
can never be greater than 
max(G)
[in]
³⁄₈
½
1
Technical data
min(F)
[kg/h]
[lb/min]
0 to 2 000
0 to 73.50
0 to 6 500
0 to 238.9
0 to 18 000
0 to 661.5
0 to 45 000
0 to 1 654
0 to 70 000
0 to 2 573
0 to 6 615
0 to 12 860
0 to 29 400
0 to 80 850
max(F)
x
3
[kg/m
]
60
80
90
to 
max(F)
223