108
Flag CALC (result error)
Flag CALC (result error)
Flag value
0
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
21
22
23
24
25
26
27
28
29
30
31
255
y List of flag CALC
Roche Diagnostics
COBAS INTEGRA 400 plus analyzer/Cedex Bio HT Analyzer · Software version 3.6/5.1 · Host Interface Manual · Version 1.24
Priority
Name
1
HIGH ABS
2
DARK ERR
3
OUTLIERS
4
FP UNSTAB
5
POL ERR
6
LOW SIGNAL
7
NO FLUID
8
NO ISECAL
9
SEG FLUID
10
SEG ISECAL
11
AIR FLUID
12
AIR ISECAL
13
ISE UNSTAB
14
< REAG RNG
15
> REAG RNG
16
EP UNSTAB
17
HIGH ACT
18
LOW ACT
19
CURVE DIR
20
NON LINEAR
21
< KIN READ
22
HIGH BKG
23
> REPL DEV
24
CALC ERROR
25
NON MONO
26
OUT OF RNG
27
> STD DEV
28
SOL 1 F DEV
29
AG EXCESS
30
< TEST RNG
31
> TEST RNG
33
TEST BLOCKED
Flag CALC provides information about the checks
performed during the calculation of the result.
q
A flag with a higher priority (lower number)
supersedes all other flags.
Meaning
No flag
High absorbency
Dark error
Outliers
FP unstable
Polarization error
Low signal intensity
No sample
No calibrator bottle
No end of sample segment detected
No end of standard segment detected
Not enough sample / too many bubbles
Not enough standard / too many bubbles
ISE measurement unstable
Low reagent range exceeded
High reagent range exceeded
Endpoint unstable
High activity (please refer to following note)
Low activity (please refer to following note)
Curve direction
Non-linear kinetic
Insufficient kinetic readings
High background noise (please refer to following note)
Replicate deviation check failed
Calculation error (please refer to following note)
Curve not monotone
Curve / slope out of range
Standard deviation exceeded
Solution 1 factor deviation
Antigen excess
Lower test range exceeded (please refer to following note)
Upper test range exceeded (please refer to following note)
Used with collated results if a test is blocked (see note)