9
Revision D

Technical Specifications

Microcontroller
AT90USB1286
Storage capacity
128 kBytes Flash
Measuring range
Systolic pressure:
60 ... 260 mmHg (8.0 ... 34.8 kPa)
Diastolic pressure: 40 ... 220 mmHg (5.3 ... 29.3 kPa)
Mean pressure:
50 ... 250 mmHg (6.7 ... 33.3 kPa)
Pulse rate (HR):
35 ... 240 min-1
Recording period
30 h or 200 measurements max.
Battery
2 Mignon (size AA) NiMH rechargeable batteries, 1.2 V, >1500 mAh or
2 Mignon (size AA) alkaline batteries
Charging time for rechargeable batteries: 2 ... 3 h
Max. cuff pressure
300 mmHg
Measuring method
oscillometric
Ambient conditions
Operation
Temperature +10...+ 40 °C
Relative humidity 15...85 %, no condensation
Atmospheric pressure 700 hPa (altitude +2800 m relative to sea level) to 1060 hPa 
(-400 m relative to sea level)
Transport and storage
Temperature -20...+ 70 °C
Relative humidity 10...90 %, no condensation
Atmospheric pressure 500 hPa (altitude +4500 m relative to sea level) to 1060 hPa 
(-400 m relative to sea level)