7.6.3
Replace the injector needle and seal pack
Replace the injector needle and seal pack when these conditions apply:
•
The needle is bent.
•
The tip of the needle is damaged.
•
Contaminant peaks appearing in the chromatogram cannot be corrected by changing the
wash solvent in the needle wash system.
Tip:
For information about isolating problems in the sample management system, see
The procedure for replacing the needle and seal pack involves these tasks:
•
Removing the injector
•
Removing the seal pack and needle
•
Removing and replacing the needle
•
Installing the new seal pack and/or needle
•
Installing the injector
•
Adjusting the seal pack
Required materials
•
Screwdrivers, T-10 and T-20 TORX
•
Wrenches, 5/16-inch, open-end, two
•
Squirt bottle with 100% alcohol (methanol or isopropanol)
•
Screwdriver, #2 cross-head
Required kits
To replace the seal pack, you need one of the following kits:
•
Seal Pack Assembly and Needle Kit (includes an assembled seal pack, replacement
needle, and instructions)
•
Seal Pack Seal Replacement Kit (includes seal pack seals, frits, replacement needle, and
instructions)