7.6.3

Replace the injector needle and seal pack

Replace the injector needle and seal pack when these conditions apply:
The needle is bent.
The tip of the needle is damaged.
Contaminant peaks appearing in the chromatogram cannot be corrected by changing the
wash solvent in the needle wash system.
Tip:
For information about isolating problems in the sample management system, see
The procedure for replacing the needle and seal pack involves these tasks:
Removing the injector
Removing the seal pack and needle
Removing and replacing the needle
Installing the new seal pack and/or needle
Installing the injector
Adjusting the seal pack
Required materials
Screwdrivers, T-10 and T-20 TORX
Wrenches, 5/16-inch, open-end, two
Squirt bottle with 100% alcohol (methanol or isopropanol)
Screwdriver, #2 cross-head
Required kits
To replace the seal pack, you need one of the following kits:
Seal Pack Assembly and Needle Kit (includes an assembled seal pack, replacement
needle, and instructions)
Seal Pack Seal Replacement Kit (includes seal pack seals, frits, replacement needle, and
instructions)