Number Of Decimal Digits; Password-Protected Data; Hardware Key - Siemens SITRANS F FUS380 Operating Instructions Manual

Ultrasonic flowmeters
Table of Contents

Advertisement

Functions

7.4 Hardware key

7.2

Number of decimal digits

For the display values (totalizers in menu 1 and 2 and flow in menu 3) Auto adjust decimal
point is the default setting. This means that the number of digits after the decimal point
automatically will be reduced with increasing number of digits before the decimal point.
7.3

Password-protected data

In the Modbus communication via SIMATIC PDM, the flowmeter information is protected by a
password. The default password is "1000" and it can be changed after gaining access to the
flowmeter or via the PDM menu [Device → Change Password (Online)]. The password can be
changed without the use of the HW key.
Figure 7-2
Click [Write the new password to the device] and [Close].
7.4
Hardware key
HW key
To gain access to protected parameters of the transmitter, a hardware jumper (HW key) must
be installed as shown below. The HW key place is located internally on the right top part of
PCB behind the display as shown.
Note
For type-approved and verified FUE380 flowmeter the HW key is protected by a verification
sealing on the display frame (see Sealing of FUE380). The verification sealing can only be
opened by the user with the acceptance of the local authorities.
60
Change password
Operating Instructions, 09/2016, A5E00730100-AB
FUS380/FUE380