Reagents for Measuring Total Chlorine

Reagents needed:
 1 x 50 ml of Oxycon On-line DPD
 1 x 240 gr of Oxycon On-line Buffer
 1 x 60 gr of Reagent Oxycon On-line KI
 4 liters demineralized water
1 Rinse the canister [E] labelled OXYCON ON LINE DPD-Re-
agent with demineralized water.
2 Fill the canister up to the 2 liter mark with demineralized water.
3 Slowly pour the content of one bottle of concentrate Oxycon
On-line DPD (50 ml) into the canister.
Avoid splashing!
4 Close the canister with the screw cover and tighten it well.
5 Mix the demineralized water and the reagents well.
6 Put the canister [E] into the holder [G].
7 Remove the screw cover, insert suction lance [A] and tighten
the screw cover.
1 Rinse the canister [F] labelled "OXYCON ON LINE Buffer" solu-
tion with demineralized water.
2 Fill the canister up to the 2 liter mark with demineralized water.
3 Slowly pour the content of one bag of Oxycon On-line Buffer
into the canister.
Avoid splashing!
4 Add the content of one bag Oxycon On-line KI to the same can-
ister.
5 Close the canister with the screw cover and tighten it well.
6 Mix the demineralized water and the reagents well.
7 Put the canister [F] into the holder [G].
8 Remove the screw cover and insert suction lance [B]. Tighten
the screw cover.
9 Fill reagent system. See
Fill or Flush Reagent System, p. 38.
49