Error
E-03
Reading flashes
followed by
E-04
E-06
E-07
E-09
E-10
8
The value that flashes will be 10% over the upper test range limit.
22 English
Description
Standard adjust
error
The reading is
more or less than
the instrument
8
range.
Absorbance error The absorbance value is not correct or the
Standard value
error
Flash error
Environment
temperature too
high or too low
Solution
• The measured value of the standard
solution is more than the adjustment
limits. Prepare a fresh standard.
• The standard solution is not within the
concentration range that can be used
for standard calibration adjust. Prepare
a standard with a value at or near the
recommended concentrations given in
the procedure.
• Make sure that the concentration of the
standard solution is entered correctly.
If the reading is less than the instrument
range, make sure that the instrument cap
is fully installed on the cell holder.
Measure a blank. If the blank reading is
not zero, set the instrument to zero again.
If the reading is more than the instrument
range, identify if there is a light blockage in
the cell holder. Dilute the sample. Do the
test again.
user-entered calibration curve has fewer
than two points. Enter or measure the
absorbance value again.
The standard solution concentration is
equal to another standard solution
concentration that is already entered in the
user-entered calibration curve. Enter the
correct standard concentration.
The instrument is not able to save data.
Push and hold
the instrument.
The ambient temperature is out of range.
Use the instrument only in the specified
operating conditions. Refer to

Specifications

on page 3.
for 5 seconds to reset

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DR300 and is the answer not in the manual?

Questions and answers

dave
March 18, 2025

Display shows E-21

1 comments:
Mr. Anderson
March 18, 2025

The E-21 error code on the Hach DR300 indicates an unstable signal measurement. This occurs when the light detector receives an inconsistent signal due to excessive or unstable ambient light. To resolve this issue, ensure that the instrument cover is fully placed over the cell holder and repeat the test. If the error persists, contact technical support.

This answer is automatically generated

Table of Contents

Save PDF