Figure 5. OAS Pump control panel
______________________________________________________________________
A. Low saline red LED indicator
B. Prime button
C. Start button and green LED indicator
D. Status yellow LED indicator
______________________________________________________________________
2. Prepare a full 1000 mL bag of normal saline solution with lubricant. Refer to the
VIPERSLIDE Lubricant Instructions for Use for lubricant preparation instructions.
3. Hang the prepared saline bag with lubricant from the low saline level sensor on the
standard IV pole.
Caution: Do not use glass bottles for the saline solution with lubricant or hang
multiple saline bags from the low saline level sensor as this will disable the Low Saline
Information signal.
11.4. OAD Set Up
Carefully remove the OAD (or individual OAD handle and cartridge) from the tray(s) and
set onto a stable surface.
Note: Removal of the OAD from the packaging by tipping the tray and allowing the OAD
to fall out can result in damage to the OAD or ancillary devices underneath the OAD.
11.4.1. Connect the Cartridge to the Handle