TEST METHOD Binocular Visual Function Test
 Vertical phoria correction
1
Present the vertical coincidence chart.
Press
The polarizing filters are placed in the measuring windows. (The red and green filters are
placed for the SC-2000, SC-1600.)
The base up/down prism powers can be entered.
2
Ask the patient, "Can you see a square with a dot in the center? Are the left and right
frames aligned? Or, is one higher than the other?"
When they are not aligned, correct them as follows:
<1> The left frame
is higher.
<2> The right frame
is higher.
182
.
Appearance of
chart
Right eye
hyperphoria
hyperphoria
• The rotary prism lenses are removed from the measuring windows when the already
selected button is pressed again in prism mode.
ex.When
• Use
or
• When the prism power is added, the prism power can be added in 0.5 D increments by
adding the prism power for single-eye compared to both eyes.
Dial
,
Phoria
Press
and right frames are aligned.
(Add BD to the right eye and BU to the left eye.)
Press
Left eye
right frames are aligned.
(Add BU to the right eye and BD to the left eye.)
is pressed in the base up/down prism, the rotary prism lenses are removed.
instead of the dial for fine adjustment.
1.0 0.5
0.20.1
161221687130 承認済
Correction
. Turn the dial counterclockwise until the left
. Turn the dial clockwise until the left and

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RT-5100 and is the answer not in the manual?

Questions and answers

Prudence Tam
February 20, 2025

Hello! I **** currently unable to use the touch screen to toggle between sphere, cylinder, and axis when refracting. The other functions on the touch screen are functional. I also cannot switch between OD and OS using the touch screen when refracting. The only way I **** able to switch selection between the sphere, cylinder, and axis is to either use the physical buttons or center button on the dial. The only way I **** able to switch from OD to OS is to use the "R" and "L" physical buttons.

User image 67b78a1e82325
1 comments:
Mr. Anderson
February 20, 2025

If the Nidek Medical RT-5100 touch screen is not allowing toggling between sphere, cylinder, and axis, or switching between OD and OS, perform the position adjustment of the touch-screen panel (see section 7.13). If the issue persists, further troubleshooting or servicing may be required.

This answer is automatically generated

Caterina
April 10, 2025

having trouble syncing each machine. after scanning patients eyes, id number wont pull through to the touch screen machine. Also when doing final, print, print, it won't send info to my computer.