Cleaning and Disinfecting Procedures
NOTE
- Clean and disinfect the outside of the device only. Do not remove battery cover when
cleaning and disinfecting.
To clean the meter:
1. Throughly wipe the entire surface of the meter with disinfecting wipes
listed above to clean any possible dirt, dust, blood and other body
fluids.
To disinfect the meter:
2. Take another disinfecting wipe and wipe the meter thoroughly. (Note: All
blood and body fluids should be cleaned from surface before
performing the disinfecting procedure)
3. Allow the surface to remain wet for 1 minute.
4. Allow to air dry.
NOTE
- Your GE006 Blood Glucose Meter has been tested to ensure that there is no change in the
performance or external materials of the device after 550 cleaning cycles and 550
disinfecting cycles. The testing simulates 2 cleaning and disinfecting cycles per week over
the typical life of the meter (5 years).
48
To clean the lancing device:
1. Throughly wipe the entire surface of the lancing device with
disinfecting wipes listed above to clean any possible dirt, dust, blood
and other body fluids.
1
To disinfect the lancing device:
2. Take another disinfecting wipe and wipe the Lancing device
thoroughly. (Note: All blood and body fluids should be cleaned from
surface before performing disinfecting procedure)
3. Allow the surface to remain wet for 1 minute.
Allow to remain
wet for 1 minute
Cleaning and Disinfecting Procedures
NOTE
- Your
Rightest™
GD500 Lancing Device has been tested to ensure that there is no
change in the performance of the device after 550 cleaning cycles and 550 disinfecting
cycles. The testing simulates 2 cleaning and disinfecting cycles per week over the typical
life of the meter (5 years).
1
2
3
Allow to remain
wet for 1 minute
49