SITE PREPARATION GUIDE QuantStudio 3 and 5 Real-Time PCR ™ Systems Publication Number MAN0010405 Revision C.0 ■ Site preparation overview ..........2 ■...
Site preparation overview Site preparation overview Customers have the option to install the instrument themselves or purchase installation service from the field services team. Note: Optional installation by a Thermo Fisher Scientific service representative includes some basic operator training and a review of data during installation. If you opt for a service representative to install the instrument, contact Thermo Fisher Scientific to schedule the installation.
Site preparation checklist Site preparation checklist Complete all items in the following checklist before you install the instrument. IMPORTANT! For installation by a Thermo Fisher Scientific service representative, complete, date, and initial the following checklist before the scheduled installation date. If the site preparation checklist is not complete when the service representative arrives, the scheduled installation may be postponed.
Customer responsibilities Customer responsibilities Personnel Responsibilities • Receives and inspects the instrument. Site preparation/ installation • Unpacks and stores the reagents and plates as specified. coordinator • Reviews the site preparation guide for safety information and the instrument requirements. • Coordinates personnel and tasks. •...
Site requirements Site requirements Dimensions and Note: All measurements are rounded to the nearest whole- or half-unit. weights Components (packaged) WARNING! PHYSICAL INJURY HAZARD. The packages are heavy. Do not attempt to lift or move the packages without professional assistance. Any incorrect lifting or moving of the packages can cause serious injury.
Page 6
Site requirements Instrument and computer connections Figure 1 Instrument back panel USB ports RS232 Port—For service use only WiFi USB port—Connect USB wireless Fuse Cover adapter for wireless network access Power Switch (ordered separately) Power Port—100 to 240 VAC Ethernet Port—RJ45 port for 100/1,000 Mbps Ethernet communication with the instrument Figure 2 Instrument‑to‑computer connections (minitower configuration)
Page 7
Site requirements Environmental Table 1 Environmental requirements requirements Condition Acceptable range Installation site Indoor use only Electromagnetic Do not use this device in close proximity to sources of strong interference electromagnetic radiation (for example, unshielded intentional RF sources). Strong electromagnetic radiation may interfere with the proper operation of the device.
Page 8
Site requirements Electrical WARNING! For safety, the power outlet used for powering the instrument requirements must be accessible at all times. See “Instrument clearances“ on page 5 for information about the space needed between the wall and the instrument. In case of emergency, you must be able to immediately disconnect the main power supply to all the equipment.
Page 9
Site requirements Electrical We recommend the use of electrical protective devices to protect the system in environments with large voltage and power fluctuations. protective devices Recommended devices Power line regulator • 1.5-kVA power line regulator • Use in areas where the supplied power fluctuates in excess of ±10% of the normal voltage. •...
Instrument-to-computer connections Instrument-to-computer connections Supported options We support the following direct, networked (LAN– l ocal area network), or Cloud configurations. Configurations other than those listed are not recommended. Select a for instrument configuration that meets the needs of your laboratory's instrument, software, and and computer workflow requirements.
Page 11
Instrument-to-computer connections Networked Note: Networked (LAN) configurations can be internal configurations that do not require internet access. However, if you are networking an instrument for Cloud (local area access, the LAN must have internet access. network– LAN) configuration IMPORTANT! For all wired configurations using an Ethernet cable, we support DHCP or static IP only.
Safety requirements Safety requirements Safety practices A safety representative from your facility must ensure that: • Personnel establish and follow all applicable safety practices and policies to protect laboratory personnel from potential hazards. • All applicable safety devices and equipment are available at all times. Required safety Your laboratory has specific safety practices and policies designed to protect laboratory personnel from potential hazards that are present.
Receive and inspect the shipment Receive and inspect the shipment 1. Verify that the items on the shipping list are the same items that you ordered at the time of purchase. 2. Carefully inspect the shipping containers. Report any damage to customer support.
Materials for installation and operation Materials for installation and operation Installation Ensure the following materials are available before installation of the instrument. materials • Safety glasses, lab coats, and chemical-resistant, disposable gloves (powder-free) • Easily accessible specified power outlet • (Optional) Electrical protective devices (universal power supply unit, surge protector, and/or power line regulator) •...
Download and install the desktop software Download and install the desktop software Computer If you purchased a computer provided by Thermo Fisher Scientific, you can install the desktop software and use it to control the instrument. requirements for the desktop Thermo Fisher Scientific does not support the use of customer-provided computers to software control the instrument.
Register to access the Thermo Fisher Cloud Antivirus No antivirus software is provided because customer preferences and network requirements vary. You are responsible for installing antivirus software of your choice requirements to protect the computer against viruses. Compatible USB The system supports USBs with formats: FAT, FAT32, and NTFS. formats IMPORTANT! Do not use a USB with exFAT formatting.
Documentation and support Documentation and support Related documentation Document Pub. No. QuantStudio ™ 3 and 5 Real-Time PCR Systems Installation, Use, and Maintenance Guide MAN0010407 QuantStudio Design and Analysis desktop Software User Guide ™ MAN0010408 QuantStudio 3 and 5 Real-Time PCR Systems SAE Admin Console User Guide ™...
Page 18
Manufacturer's address: Life Technologies Holdings Pte Ltd | Block 33 | Marsiling Industrial Estate Road 3 | #07-06, Singapore 739256 The information in this guide is subject to change without notice. DISCLAIMER: TO THE EXTENT ALLOWED BY LAW, LIFE TECHNOLOGIES AND/OR ITS AFFILIATE(S) WILL NOT BE LIABLE FOR SPECIAL, INCIDENTAL, INDIRECT, PUNITIVE, MULTIPLE, OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH OR ARISING FROM THIS DOCUMENT, INCLUDING YOUR USE OF IT.
Need help?
Do you have a question about the QuantStudio 3 and is the answer not in the manual?