Digital color doppler palm ultrasound system (76 pages)
Summary of Contents for Chison ECO 6
Page 1
Digital Color Doppler Ultrasound System Model ECO 6 V1.2 Sep 12, 2017 OPERATION MANUAL Direction: 57-00594-00 CHISON Medical Technologies Co., Ltd. We reserve the rights to make changes to this manual without prior notice. WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL...
Page 3
Essential Requirements of the Medical Device Directive. This manual is a reference for the ECO 6. Please verify that you are using the latest revision of this document. If you need the latest revision, contact your distributor.
Page 4
9. This manual contains warnings regarding foreseeable potential dangers, but user shall always be alert to dangers other than those indicated as well. CHISON shall not be liable for damage or loss that results from negligence or from ignoring the precautions and operating instructions described in this operation manual.
Page 5
15. While the system or probe is over life time, please refer to operation manual section 9.4 16. Important data must be backed up on external memory media. CHISON shall not be liable for loss of data stored in the memory of this system caused by operator error or accidents.
ECO 6 Digital Color Doppler Ultrasound System Content Chapter 1 Introduction ..........................1 1.1 System Overview ........................1 1.2 Contact Information ........................ 1 Chapter 2 System Safety........................... 2 2.1 Safety Overview ........................2 2.2 Electrical Safety ........................3 2.3 Label............................5 2.3.1 Warning Symbols ......................
Page 7
ECO 6 Digital Color Doppler Ultrasound System 4.2 Alphanumeric Keyboard ....................... 25 4.3 Function Keys/Knob ......................26 4.4 Parameter Control key ......................28 4.5 STC ............................28 4.6 Central Control ........................28 4.7 Information Area Indicating Machine Status ..............29 4.7.1 Indicator Light ......................
Page 8
ECO 6 Digital Color Doppler Ultrasound System 5.5.14 Biospy and Super Needle .................... 36 5.5.15 PW Mode ........................36 5.6 B Image Adjustment ......................37 5.6.1 Parameters Adiustment ....................37 5.6.2 B Menu Adjustment ...................... 38 5.63 Accessories........................40 5.7 CFM Image Adjustment ......................40 5.8 PW / CW Image Adjustment ....................
Page 9
ECO 6 Digital Color Doppler Ultrasound System 5.14.2 Manual Loop ....................... 52 5.14.3 Automatic Loop ......................52 5.14.4 Save and Recall Image ....................52 5.14.5 Save and Recall Cine ....................52 5.14.6 Delete images ......................52 5.14.7 Send images ........................ 52 5.15 Image Browse ........................
Page 10
ECO 6 Digital Color Doppler Ultrasound System 6.4 B General Measurement ......................62 6.5 ABD Measurement ......................... 63 6.6 OB Measurement ........................65 6.6.1 Twins Measurement ...................... 67 6.6.2 EDD (estimated date of delivery) Estimation ............... 67 6.6.3 Growth curves ....................... 67 6.7 Pediatric Measurement ......................
Page 11
ECO 6 Digital Color Doppler Ultrasound System 6.22.4 Auto Trace ........................79 6.22.5 Manual Trace ......................79 6.23 PW Fast Measurement ......................79 6.24 PW Genearl Measurement ....................81 6.25 PW Abdomen Measurement ....................83 6.26 PW OB Measurement ......................84 6.27 PW GYN Measurement .......................
Page 12
ECO 6 Digital Color Doppler Ultrasound System 7.6 Keyboard ..........................108 7.7 DICOM ..........................109 7.7.1 Add/Edit DICOM Function ..................110 7.8 NET Work ..........................110 7.9 System ........................... 111 7.9.1 System information ..................... 111 7.9.2 Update ......................... 111 7.9.3 Function Setting ......................111 7.9.4 Installment setting .......................
Page 13
ECO 6 Digital Color Doppler Ultrasound System Appendix C: Guidance and Manufacturer’s Declaration ..............179 Appendix D: Measurement Results Summary..................183 Appendix E: Display Accuracy and Acoustic Measurement Uncertainties .......... 184 Appendix F: Transducer Maximum Surface Temperature ..............185 Appendix G: Procedures of set network sharing in ECO series ............186...
ECO 6 Digital Color Doppler Ultrasound System Chapter 1 Introduction This manual contains necessary information for safe system operation. Read and understand all instructions in this manual before operating the system. Always keeping this manual with the equipment, and periodically review the procedures for operation and safety precautions.
ECO 6 Digital Color Doppler Ultrasound System Chapter 2 System Safety 2.1 Safety Overview This section discusses measures to ensure the safety of both the operator and patient. To ensure the safety of both operator and patient, please read the relevant details in this chapter carefully before operating this system.
ECO 6 Digital Color Doppler Ultrasound System 2.2 Electrical Safety Type of protection against electric shock Class I Equipment CLASS I EQUIPMENT in which protection against electric shock does not rely on BASIC INSULATION only, but includes a protective earth ground. This additional safety precaution prevents exposed metal parts from becoming LIVE in the event of an insulation failure.
Page 17
Locate the equipment as far away as possible from other electronic equipment. Be sure to use only the cables provided by or designated by CHISON. Connect these cables following the installation procedures (e.g. wire power cables separately from signal cables).
ECO 6 Digital Color Doppler Ultrasound System 2.3 Label Fig. 1 Real panel label 2.3.1 Warning Symbols Caution, consult accompanying documents. Dangerous electric voltage. Unplug the main This symbol advises the reader to consult the plug before opening the system!
ECO 6 Digital Color Doppler Ultrasound System information concerning the decommissioning of your equipment. AUTHORIZED REPRESENTATIVE Type-BF applied part IN THE EUROPEAN COMMUNITY: This symbol is accompanied by the name and the address of the authorized representative in the European Community.
Page 20
The Patient Environmental devices shown above are specified to be suitable for use within the PATIENT ENVIRONMENT. CAUTION: DO NOT connect any probes or accessories without approval by CHISON within the PATIENT ENVIRONMENT. DO NOT touch patient and devices without IEC/EN 60601-1 approval to avoid the leakage current risk within the PATIENT ENVIRONMENT.
Such malfunction may impact the stability, effectiveness and safety of the system and probes, and users should immediately stop using the system and probes until CHISON authorized engineer has checked the system and confirm the effectiveness and safety of the system and probes.
ECO 6 Digital Color Doppler Ultrasound System Heating: Elevating tissue temperature during obstetrical examinations creates medical concerns. At the embryo development stage, the rise in temperature and the length of time exposed to heat combine to determine potential detrimental effects. Exercise caution particularly during Doppler/Color exams. The Thermal Index (TI) provides a statistical estimate of the potential temperature elevation (in centigrade) of tissue temperature.
Page 23
ECO 6 Digital Color Doppler Ultrasound System Select a low setting for each new patient. Output should only be increased during the examination if penetration is still required to achieve a satisfactory result, and after the Gain control has been moved to its maximum value.
ECO 6 Digital Color Doppler Ultrasound System Non-diagnostic use of ultrasound equipment is not generally recommended. Examples of non-diagnostic uses of ultrasound equipment include repeated scans for operator training, equipment demonstration using normal subjects, and the production of souvenir pictures or videos of a fetus. For equipment of which the safety indices are displayed over their full range of values, the TI should always be less than 0.5 and the MI...
Page 25
ECO 6 Digital Color Doppler Ultrasound System Where a is the attenuation coefficient in dB cm-1 MHz-1, f is the transducer center frequency, and z is the distance along the beam axis between the source and the point of interest.
Page 26
ECO 6 Digital Color Doppler Ultrasound System TI is defined as the ratio of the In Situ acoustic power (W.3) to the acoustic power required to raise tissue temperature by 1° C (Wdeg), TI = W.3/Wdeg. Three TIs corresponding to soft tissue (TIS) for abdominal; bone (TIB) for fetal and neonatal cephalic; and cranial bone (TIC) for pediatric and adult cephalic, have been developed for applications in different exams.
ECO 6 Digital Color Doppler Ultrasound System Operator Control Features: The user should be aware that certain operator controls may affect the acoustic output. It is recommended to use the default (or lowest) output power setting and compensate using Gain control to acquire an image. Other than the output power setting in the soft-menu, which has the most direct impact on the power;...
Page 28
ECO 6 Digital Color Doppler Ultrasound System Do not pierce the battery casing with a nail or other sharp object, break it open with a hammer, or step on it. Do not strike, throw or subject the battery to sever physical shock.
ECO 6 Digital Color Doppler Ultrasound System Chapter 3 System Introduction 3.1 Console Overview Fig. 2 Console Overview 3.2 Physical Specification Dimensions of main unit (approx.): 335mm(Length)× 155mm(Width)× 350mm(Height) Net weight of main unit (approx.): 6.5kg WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL...
ECO 6 Digital Color Doppler Ultrasound System 3.3 System View in Different Views Fig. 3 System Front View Fig. 4 System Side View 1. Ethernet 2. VGA 3.USB 4.FOOT SWITCH 5.DC IN Fig. 5 Console Overview 1. Probe Holder 2.Probe 3.Keyboard...
ECO 6 Digital Color Doppler Ultrasound System LAN port output LAN for DICOM and image review station USB 2.0 for flash drive Foot switch AC/DC adapter: MDS-060AAS19 B Input: 100-250V~1.5-0.75A, 50-60Hz Output: 19V 3.15A DELTA ELECTRONICS, INC. Battery Pack: BT-2500, 4400mAh, DONGGUAN POWER INC 3.5 Installation Procedures...
Page 34
ECO 6 Digital Color Doppler Ultrasound System Power the equipment from a source different from that of the affected device. Consult the point of purchase or service representative for further suggestions. 3.5.1.2 Transport and Storage Environmental Requirement The following environmental transport and storage conditions are within system tolerances: Temperature: -5º...
Page 35
ECO 6 Digital Color Doppler Ultrasound System 3.5.1.4 Operation Space Please leave enough free space from the back of the system to ensure well ventilation. Caution: Leave enough free space from the back of the system; otherwise, with the increasing of the temperature inside the unit, malfunction may occur.
ECO 6 Digital Color Doppler Ultrasound System 3.5.2 Powering the System 3.5.2.1 Acclimation Time After being transported, the unit requires one hour for each 2.5 º increment if its temperature is below 10 º C or above 40 º C.
ECO 6 Digital Color Doppler Ultrasound System When installing and disassembling probe, please put the probe head inside the probe holder, it can prevent the probe falling down to the ground. 3.5.3.1 Probe Disassembly Turn the dead lock switch 90 degree in counterclockwise direction, extract probe connector plug vertically.
ECO 6 Digital Color Doppler Ultrasound System Chapter 4 Control Panel 4.1 Keyboard Appearance Fig. 7 Keyboard Appearance 4.2 Alphanumeric Keyboard Fig. 8 Alphanumeric Keyboard The alphanumeric keys are used for inputting patient’s number, name, character and figure etc. WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL...
ECO 6 Digital Color Doppler Ultrasound System 4.3 Function Keys/Knob Key/knob Icon Name Function description Switch Turn on or off the device. Press this key for selecting probe. It can only select the Probe connected probe. Set up a new patient data, input name and other Patient information.
Page 40
ECO 6 Digital Color Doppler Ultrasound System Display B mode. Display CFM mode. Display PW mode. Press this key to change the mode between B/M and M.. Press this key to open or close the THI function.. Only press this key for automatic optimization image.
ECO 6 Digital Color Doppler Ultrasound System 4.4 Parameter Control key Increase/decrease the corresponding parameters of the screen, or open/close the function. 4.5 STC STC can be used for adjusting gain compensation in different image depth. 4.6 Central Control 1. UPDATE 2.EXIT 3.Cursor 4.Change 5.MENU 6.
ECO 6 Digital Color Doppler Ultrasound System Key/knob’s name Function description This multifunction key is work with trackball. The function switches with the unit UPDATE status. Such as, call the annotation and back in measuring. Press this key can exit measurement, dialog, and menu.
Page 43
ECO 6 Digital Color Doppler Ultrasound System From left to right: Adapter Indicator, Charge Indicator, and Sleep Indicator. Adapter Indicator: when the main unit connects to the adapter with power supply, the indicator lights, otherwise extinguishes. Charge Indicator: When the battery is charging, the indicator lights, once the battery is charged, the indicator goes out.
ECO 6 Digital Color Doppler Ultrasound System Chapter 5 Operation and Exam Mode This chapter mainly describes the process of the normal operation of the device, including the preparation before examination, how to get the image, optimize the image, add comments, body mark and so on.
ECO 6 Digital Color Doppler Ultrasound System 5.3 Patient Data Entry Press the to display the Patient screen Fig. 9 Patient’s Information Screen Function Buttons on Patient screen: [Archive]: Operation on the patient information which has already existed; [Wordlist]: Recall patient information in worklist. And need to open the DICOM function;...
ECO 6 Digital Color Doppler Ultrasound System before saving measurement or image; otherwise, it will be stored in the wrong patient records. After checking the patient, press the [END] key to save the patient’s information in the system. 5.4 Image Interface Display Fig.
ECO 6 Digital Color Doppler Ultrasound System real-time status. Pressing it again can switch the real-time status among four images. 5.5.4 B/M Mode Press [M] button, a real time B-mode image and a real-time M-mode image will be displayed at the same time.
ECO 6 Digital Color Doppler Ultrasound System Color Gain: amplify the overall strength of echoes processed in the CROI Persistence: affect temporal smoothing and color Doppler ‘robustness’. 5.5.7 B/BC Mode In active color mode, press the [MENU] to enter into the C Menu and select [B/BC] option to turn on the [B/BC] item to display a real B mode image at the right side of the screen and active Color mode image at the left side of the screen.
ECO 6 Digital Color Doppler Ultrasound System 5.5.14 Biospy and Super Needle 1. How to enter into Biopsy Activate the [Biopsy] and press the [MENU] knob to show or hide biopsy line. 2. How to adjust the biopsy After the biopsy line shows, press the [ENTER] key to activate the adjustment function of biopsy line, horizontal rolling the trackball can translate the biopsy line, vertical rolling the trackball can adjust the line angle, press the [UPDATE] key to set the default biopsy line position.
ECO 6 Digital Color Doppler Ultrasound System function between Sample Volume Gate position and size. Press the [UPDATE] key after the Sample Volume Gate is defined to activate the Spectral Doppler mode. Press the [UPDATE] key for second time to toggle back to 2D (B or Color) update and deactivate the Spectral Doppler.
ECO 6 Digital Color Doppler Ultrasound System Turn ON/OFF THI function. Only press this key to optimize image. B mode image and B/M mode image can be reversed horizontally and vertically. Press the key, the displayed image is reversed in the right-left horizontal direction.
Page 52
ECO 6 Digital Color Doppler Ultrasound System In real status, move the cursor to [Persistence] selection to adjust. Dynamic range is used for adjusting the contrast resolution of B mode image and mode image, compressing or enlarging the display Dynamic range of gray scale.
ECO 6 Digital Color Doppler Ultrasound System the left or right image and enlarge the area without to rotate the probes. Press the [MENU] to show or hide biopsy line. After the biopsy line shows, press the [ENTER] key to activate the...
ECO 6 Digital Color Doppler Ultrasound System Activate [Color Invert] and rotate [MENU] to turn on or off. Or press realize Color invert. Steering Angle Activate [Steering Angle] and rotate [MENU] to adjust Steering Angle. Change type of color map.
ECO 6 Digital Color Doppler Ultrasound System Freq In real status, press the corresponding key of [Freq] to adjust. Adjust the position of baseline. Base Line arameter control key Press p on [Baseline] to change position of baseline. Activate [PRF] and rotate Menu to Adjust PRF.
ECO 6 Digital Color Doppler Ultrasound System spectrum. 5.9 M Image Menu & Parameters Adjustment Menu & Parameters Function description Menu This function includes post processing, slide show and other items. Press Utility [MENU] knob, then choose Utility selection, it will appear Utility options.
ECO 6 Digital Color Doppler Ultrasound System Parameter keys Function description Freq In real status, press the corresponding key of [Freq] to adjust. Adjust the position of baseline. Base Line arameter control key Press p on [Baseline] to change position of baseline.
ECO 6 Digital Color Doppler Ultrasound System better image quality while low density image has higher frame rate. Utility Please refer to the Utility of B Menu adjustment. 5.11 Full Screen Show Full screen the image area. Press [GAIN] knob to activate the function; Press [EXIT] button or [GAIN] knob again to exit full screen show.
ECO 6 Digital Color Doppler Ultrasound System 5.12.3 Input Comment Library Characters 1. In comment status, move trackball to image area to edit; 2. Press [Font size] to adjust front size of comments; 3. Rotate [MENU] to select needed comments, then press [MENU] to exit ;...
ECO 6 Digital Color Doppler Ultrasound System 5.12.9 Set the Position of Default Comment Operation: 1. Press [Save Home Pos.] to move cursor to the initial position; 2. Press [Load Home Pos.] to set the initial position; 5.13 Set Body Mark 5.13.1 General Description...
Page 61
ECO 6 Digital Color Doppler Ultrasound System Fig. 13 Obstetric Mark Fig. 14 Gynecology Mark WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL...
Page 62
ECO 6 Digital Color Doppler Ultrasound System Fig. 15 Small Part Mark Fig. 16 Urology Mark WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL...
Page 63
ECO 6 Digital Color Doppler Ultrasound System Fig. 17 Vascular Mark Fig. 18 Nerve Mark WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL...
ECO 6 Digital Color Doppler Ultrasound System Fig. 19 Fast Mark 5.13.2 Body Mark Operation Operation: 1. Press to enter into the body status. 2. Select the body mark that you need. 3. Move trackball after adding the body mark image and then adjust the position of probe. Rotate [MENU] or [ANGLE] knob can adjust the probe direction.
ECO 6 Digital Color Doppler Ultrasound System Fig. 20 Cine loop indicate diagram 5.14.2 Manual Loop Press [FREEZE] to freeze image, pop cine playback bar, at this time, move cursor to play by hand; Trackball rolling to the right, the loop play with ascending order side by side. Or press parameter button on [Next/Pre] to play.
ECO 6 Digital Color Doppler Ultrasound System 5.15 Image Browse Press to enter image information browsing interface. Press [ENTER] to any function in the image. Fig. 21 Review interface ID: ID of Current patient. Name: Current patient’s name.
ECO 6 Digital Color Doppler Ultrasound System Fig. 22Archive interface Item: Type selection, select Patient’s ID or Name; Keyword: Search for key words; Period: Time filter, select today, one week, one month, three months, six months, recent one year and all;...
Page 68
ECO 6 Digital Color Doppler Ultrasound System Reports contain normal report, abdominal report, cardiac report, small part report etc. Move the cursor to the required report page and press [ENTER] to choose. Press the [Report] button and the page pop up the report page of the current exam mode. Change the other exam mode report by the drop-down box.
ECO 6 Digital Color Doppler Ultrasound System 5.18 DICOM 5.18.1 DICOM Worklist Press [Worklist] in Patient Interface, pop up the following dialog box. Fig. 24 DICOM Worklist Interface ID: input ID or some characters, fuzzy query needs server; Search By: select term, ID or name;...
ECO 6 Digital Color Doppler Ultrasound System Fig. 27 DICOM Send Interface Select DICOM Storage in left, choose DICOM server and press Export button to DICOM storage. Enter Task Sequence and watch or edit DICOM process. 5.18.3 DICOM Print DICOM Print operation is the same as DICOM storage.
ECO 6 Digital Color Doppler Ultrasound System Chapter 6 Measurement and Calculation Main content of this chapter: Normal calculation and measurement on B mode image and M mode image, OB calculation and Urology measurement etc., system can enter into corresponding measurement mode depend on current exam mode, and enter into the corresponding report depend on the measurement mode.
ECO 6 Digital Color Doppler Ultrasound System 6.1.4 [DEL] Main functions are as follows: In frozen status, press [DEL], delete all the measurement results, comments and traces. 6.1.5 [Change] Press [Change] to switch other menu; 6.1.6 [Exit] Press [Exit] to exit the measurement menu.
ECO 6 Digital Color Doppler Ultrasound System 4. Move the cursor to the end-point of the ellipse, press [ENTER] to fix the axis, at the same time, the next axis can be updated, and can change the size of the axis by the trackball.
ECO 6 Digital Color Doppler Ultrasound System The horizontal axis represents the gray scale of the image ranging from 0 to 250. The vertical axis represents the distribution ratio of each gray scale. The value shown on the top of vertical axis represents the percentage of the maximally distributed gray in the whole gray distribution.
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Refer to distance Meas. D1: the longer distance D2: Volume (2 Distance) V = Formula: the shorter distance (π/6)×D1×D22 Refer to distance Meas. Volume (3 Distance) D1, D2, D3: Distance V =...
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Refer to distance Volume Meas. D1, D2, D3: Distance V = (3 Distance) Formula: (π/6)×D1×D2×D3 Refer to distance A: Long Axis Volume(1 Dis 1 and ellipse Meas.
Page 77
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Refer to distance Meas. D1: Length of Normal StD% Formula: D2: Length of Stenosis ((D1-D2)÷ D1)× 1 Refer to ellipse Meas. A1: Area of Normal...
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Refer to ellipse Press [UPDATE] Vessel Area and trace Meas. change ellipse and trace. Refer to distance Vessel Dis Meas. 6.6 OB Measurement Choose OB exam mode. Freeze the required image, and then Press [Calc] to enter into OB measurement status.
Page 79
ECO 6 Digital Color Doppler Ultrasound System Meas. Menu Submenu Unit Meas. Method Comment Refer to distance Formula: CFEF Meas. Formula to choose: Bessis, CFEF, Campbell, Chitty, FTA<Ellipse>/< Refer to distance Doubilet, Hadlock, Trace>/<Auto Meas. Hansmann, Hohler, Jeanty, Cubic Spline>...
ECO 6 Digital Color Doppler Ultrasound System Meas. Menu Submenu Unit Meas. Method Comment Refer to distance Meas. Refer to distance LtKid Meas. Refer to distance RtKid Meas. Refer to distance LtRenalAP Meas. OB Others Refer to distance RtRenalAP Meas.
ECO 6 Digital Color Doppler Ultrasound System 3. Choose the growth curve need to display, and check it to show the growth curve on report. 4. Click [× ] icon on the dialog box to exit. Tips: The abscissa of Growth curves is the gestational weeks calculated according to the LMP in patient information.
ECO 6 Digital Color Doppler Ultrasound System 6.8 GYN Measurement GYN measurement includes measurement of UT-D (uterus diameter), ENDO (endometrium), CX-L (Uterine cervix length), LEFT OV and RIGHT OV (volume of left and right ovary) and LEFT FO and RIGHT FO (left and right follicle).
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula SPSA: input the SPSA Formula: PSAD ng/ml When create a new PSAD=SPSA/V Patient Residual Vol. Refer to distance Width: Residual_W Width Meas. Residual Thick: Residual _T Thick Formula: V=...
Page 85
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Comment Meas. Formula Refer to Volume of A4CTrace method EDV(A2C) Formula: ESV(A2C) SV=|EDV-ESV| Simpson BP EDV(A4C) CO=SV*HR/1000 ESV(A4C) EF=SV/EDV*100 SI=SV/BSA CI=CO/BSA Refer to Distance Meas. & Ellipse Mea.
Page 86
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Comment Meas. Formula Refer to Distance Meas. Formula: EDV=π/6*(0.98*LVIDd+5.9)*LVID d*LVIDd ESV=π/6*(0.98*LVIDs+5.9)*LVIDs LVIDd *LVIDs Gibson LVIDs SV=|EDV-ESV| CO=SV*HR/1000 EF=SV/EDV*100 FS=(LVIDd-LVIDs)/LVIDd*100 SI=SV/BSA CI=CO/BSA Refer to Distance Meas. & Ellipse MV Diam Mea.
ECO 6 Digital Color Doppler Ultrasound System yellow short line on it represents it is in active status. The distance between the two straight lines stands for time you want to measure. You can drag the active straight line to anywhere you want to change the measured time.
ECO 6 Digital Color Doppler Ultrasound System 6.16 M OB Measurement Refer to M mode general measurement. 6.17 M GYN Measurement Refer to M mode general measurement. 6.18 M Mode Cardiac Measurement Normally Cardiology measurements are performed in M and B/M mode.
Page 90
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Refer to M Distance Meas. Formula: EDV=7*LVIDd /(2.4+ LVIDd) ESV=7*LVIDs /(2.4+ LVIDs) Teichholz(LVIDd, SV=|EDV-ESV| LVIDs) CO=SV*HR/1000 EF=SV/EDV*100 FS=( LVIDd-LVIDs)/LVIDd *100 SI=SV/BSA CI=CO/BSA Refer to M Distance Meas.
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Refer to M Distance Meas. LVPEP/LVET Refer to M Time Meas. Refer to M Distance Meas. RVOT Diam Refer to M Distance Meas. RA Diam Refer to M Distance Meas.
ECO 6 Digital Color Doppler Ultrasound System 6.22.1 Velocity Refer to velocity of M mode general measurement. 6.22.2 Time Refer to time of M mode general measurement. 6.22.3 HR Refer to HR of M mode general measurement. 6.22.4 Auto Trace Measurement step: 1.
Page 93
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Formula: = Pressure(d) mmHg Pressure 4× Vd× Vd/10000 Formula:SD=Vs/Vd = Formula:SD (Vs-Vd)/Vs Time Refer to M time meas. HR(Single Single wave Refer to M HR meas.
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Comment Meas. Formula Long Volume(1Ellips Refer to ellipse meas. Axis Formula: V=(π/6)*A*B Short Axis longer Refer to Distance Meas. Volume(2Dista distance Formula: nce) V=(π/6)*D1*D2 shorter distance Refer to Distance Meas.
Page 95
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Meas. Comment Formula Formula:SD = Vs/Vd Formula:SD = (Vs-Vd)/Vs Refer Time time meas. Refer cm/s speed meas. Formula: = Pressure Pressure(s) mmHg 4× Vs× Vs/1000 Refer cm/s speed meas.
ECO 6 Digital Color Doppler Ultrasound System 6.26 PW OB Measurement Press [Calc] to enter PW measurement in PW OB mode. Or press the corresponding parameter control key to switch the OB measurement item. Meas. Method/ Meas. Menu Submenu Unit Comment Meas.
ECO 6 Digital Color Doppler Ultrasound System Meas. Method/ Meas. Menu Submenu Unit Meas. Comment Formula Refer VMean cm/s speed meas. Formula: = Pressure Pressure(VMean) mmHg 4× VMean× VM ean/10000 Formula:SD = Vs/Vd Formula:SD = (Vs-Vd)/Vs Formula:SD = (Vs-Vd)/VMea HR(Single wave) 6.28 PW Cardiology Measurement...
Page 99
ECO 6 Digital Color Doppler Ultrasound System Refer to PW Velocity Meas. Formular: MV A PG=4*MV A Vel*MV A MV A Vel cm/s Vel/10000 E/A=MV E Vel/MV A Vel A/E=MV A Vel/MV E Vel Refer to PW Trace Meas. Formular:...
Page 100
ECO 6 Digital Color Doppler Ultrasound System Refer to PW Trace Meas. Formular: PGmax=4*AV Vmax*AV Vmax/10000 AV PGmax=4*AV Vmean*AV AV VTI Vmean/10000 SV=0.785*AV Diam*AV Diam*|AV VTI| Fraction=AR Flow/AV SV*100 Formular: π AVA(VTI)= *|LVOT AVA(VTI) VTI*LVOT Diam*LVOT Diam|/AV VTI/4 Refer to PW Velocity Meas.
Page 101
ECO 6 Digital Color Doppler Ultrasound System Refer to PW Velocity Meas. Formular: TV A PG=4*TV A Vel*TV A TV A Vel cm/s Vel/10000 E/A=TV E Vel/TV A Vel A/E=TV A Vel/TV E Vel Formular: π TVA(VTI)= *|RVOT TVA(VTI) VTI*LVOT...
Page 102
ECO 6 Digital Color Doppler Ultrasound System Refer to PW Velocity Meas. Formular: LPA Vmax cm/s LPA PGmax=4*LPA Vmax*LPA Vmax/10000 RVPEP Refer to PW Time Meas. RVET Refer to PW Time Meas. Formular: PGmax=4*PR Vmax*PR RAEDP Vmax/10000 RAEDP=RAP+4*PR Vmax*PR Vmax/10000 Refer to PW Velocity Meas.
ECO 6 Digital Color Doppler Ultrasound System Refer to PW Time Meas. Refer to PW Time Meas. Formular: Tei Index=(ICT+IRT)/ET MV medial Refer to PW Velocity Meas. & (Sa,Ea,Aa,ARa,DRa) Slope Meas. MV lateral Refer to PW Velocity Meas. & (Sa,Ea,Aa,ARa) Slope Meas.
ECO 6 Digital Color Doppler Ultrasound System 6.30 PW Urology Measurement Refer to PW general measurement. 6.31 PW Small parts Measurement Refer to PW general measurement. 6.32 PW Pediatric Measurement Refer to PW general measurement. WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL...
ECO 6 Digital Color Doppler Ultrasound System Chapter 7 Preset This chapter introduces the operation to make settings of the system through preset menu at preset mode. Preset function is used to set up working environment and status, parameters of each examination mode. The setting will be stored in the memory of system and not be lost even after the system is switched off.
ECO 6 Digital Color Doppler Ultrasound System Function name Setting method Function description Time zone Set up freely Set up the working clock of the system. Enable screen saver, user can custom screen saver Screen Select the function images, a custom picture named "screensaver", JPG,...
Page 107
ECO 6 Digital Color Doppler Ultrasound System Fig. 30 General Measurement settings interface Distance: cm, mm Area: cm , mm Volume: ml, l Time: s, ms Speed: cm/s, mm/s Acceleration: cm/s , mm/s Angle: deg, rad ...
ECO 6 Digital Color Doppler Ultrasound System 7.2.2 Measurement formula setting 7.2.2.1 Interface Description-Measurement Menu Fig. 31 Interface for Edition of measurement Formula Region: pull down and select needed measurement menu. Manage: pop up measurement software edition interface, add modify delete change marshalling sequence in measurement menu.
Page 109
ECO 6 Digital Color Doppler Ultrasound System Measure Option: None, Repeated, & Next Order After choosing the Repeat and Sequential, choose one Default item measurement or calculation to activate the measurement rule 7.2.2.2 Interface Description-Measurement Manipulation Fig. 32 Interface for Edition of Measurement Formula...
Page 110
ECO 6 Digital Color Doppler Ultrasound System Average The average rule of data Method Press this button to pop up interface to modify measurement Modify operation Remove Press this button to delete selected measurement operation 7.2.2.3 Interface Description-Measurement Calculation Fig. 33 Interface for Edition of Measurement Formal...
Page 111
ECO 6 Digital Color Doppler Ultrasound System Display In Whether display in the report or not. Report Press this button to move measurement operation up. Down Press this button to move measurement operation down. Average The average rule of data.
Page 112
ECO 6 Digital Color Doppler Ultrasound System Edit Enter into interface to edit formula when selecting calculation item The minimum value displays in result zone and report The maximum value displays in result zone and report Unit Data unit which measurement operation produces...
Page 113
ECO 6 Digital Color Doppler Ultrasound System 7.2.2.6 Formula Edit-OB It is necessary to call built-in OB formula sheet when creating OB measurement operation the following function interface is required. Caution: the results of GA and EDD don’t require unit, the unit of this class has been built-in.
Page 114
ECO 6 Digital Color Doppler Ultrasound System Fig. 37 OB Table Setup Interface Measure Item: OB Measurement Item. GA Tables: Gestational list for the current measuring project. Growth Table: Growth table for the current measuring. Fetal Weight: Fetal Weight calculation formula.
ECO 6 Digital Color Doppler Ultrasound System 7.3 Comment Fig. 38 Comment Setting Interface 7.3.1 Comment Library The annotation database of the system is classified as: Abdomen, OB, GYN, Cardiac, Small part, Pathological change Annotation can be made by inputting characters from the soft keyboard or recalling the terms saved in annotation database.
ECO 6 Digital Color Doppler Ultrasound System 7.3.2 Edit Comment Operator uses only current annotation instead of all content annotation status provides inlay common annotation. If necessary, user can import annotation or self-compiled annotation into common annotation. 7.3.2.1 Add annotation from comment library Operation: 1.
ECO 6 Digital Color Doppler Ultrasound System 7.4 Body marks Fig. 39 Interface for Body Mark Setting 7.4.1 Body Marks Library Built-in body marks: Abdomen, OB, GYN, Cardiac, Small part, Urology, Vessel. Press [BodyMark Lib] pull down button, pop up name of built-in body marks, show needed body marks via Trackball and [ENTER] button.
ECO 6 Digital Color Doppler Ultrasound System [>] button to move selected body marks in [BodyMark Selected] column into source body marks. 4. Press [ENTER] on [>>] button to import all body marks in source into user-selected status; press [ENTER] on [>>] button to move all body marks in [BodyMarks Selected] column into source.
ECO 6 Digital Color Doppler Ultrasound System control area. Fig. 41 Interface for ExamMode Setting Probe: select needed probe and ExamMode fit for it. ExamMode Lib: show all existed Exam modes. ExamMode Selected: show Exam modes in probe column.
ECO 6 Digital Color Doppler Ultrasound System Fig. 42 Interface for ExamMode config ExamMode: show all existed Exam modes in system. Comment: double press [ENTER] to activate widget box, can choose existed annotation status name. After setting, the default of the Exam mode is user-selected.
ECO 6 Digital Color Doppler Ultrasound System Fig. 43 Image Freeze Config Interface Status After Freeze: options for status after freezing. Status After Real Time saving Cine Frames: whether unfreezing upon save or not. 7.6 Keyboard Users can set the number buttons and the store button on the keyboard, easy to use.
DICOM has been activated. In the system page of setting interface, you can check whether DICOM is open or not. If you want to activate DICOM, please contact with CHISON. There must be DICOM SCP server which has been installed with PACS or other relative DICOM server software.
ECO 6 Digital Color Doppler Ultrasound System Set as default: set one DICOM service as default. Send while saving: check this item and enable DICOM storage while saving image or cine, send clip or image according to activated function.
Show the status of current functions, and click the related button to turn on/off this function (need the key provided by CHISON). 7.9.4 Installment setting Input relevant key to open trial function and the detail please contact CHISON Company. WWW.CFS.IT - CFS PRODOTTI MEDICALI SRL...
ECO 6 Digital Color Doppler Ultrasound System 7.9.5 Video VGA Choose the video data: NTSC, PAL-M, and PAL-D. Video opened: Choose the item to open this function. VGA opened: Choose the item to open this function. 7.9.6 Hardware function Export hardware SN and import hardware key, it only for engineer use.
3. When it is necessary to clean the probe connector and peripheral instrument, please contact Sales office contact customer service or agent of CHISON. Any self-cleaning may result in malfunction or degrading the function of device.
ECO 6 Digital Color Doppler Ultrasound System Cleaning procedure: 1. Wipe the remaining coupling gel and blot with flowing clear water. Avoid the joint part between cable and probe touching the water or others. 2. Use wet gauze or other soft cloth with a little liquid soap to clean the probe totally. Don’t use cleaning agent and cleaner with abrasiveness.
ECO 6 Digital Color Doppler Ultrasound System Customer Service department or Agents of manufacturer. If no image or menu but image, please check following malfunction table. If the malfunction can’t be solved, please contact with Sale Office, Customer Service department or Agents of manufacturer.
ECO 6 Digital Color Doppler Ultrasound System Chapter 9 Probes 9.1 General Description Cable Stress Relief Handle Binding line Transducer Scanhead (Lens) Convex Probe Overview The probes provide high spatial and contrast ultrasound imaging of frequencies from 2.0MHz to 11.0MHz.
ECO 6 Digital Color Doppler Ultrasound System The probes are designed to be used only with this ultrasound system. Use of these probes on any other system or a non-qualified probe may cause electrical shock or damage on the system/transducer.
Page 131
Before inserting the connector into the probe port, inspect the probe connector pins. If a pin is bent, DO NOT use the probe until it has been inspected and repaired/replaced by a CHISON Service Representative. Electrical leakage checks should be performed on a routine basis by CHISON Service or qualified hospital personnel.
Page 132
ECO 6 Digital Color Doppler Ultrasound System manual. Sterilant Exposure from Probe connector to Patient (e.g. Cidex): DO NOT allow the sterilant to contact the patient. Only immerse the probe to its specified level. Ensure that no solution has entered the probe’s connector before scanning the patient.
Page 133
ECO 6 Digital Color Doppler Ultrasound System WARNING To avoid electrical shock, always turn off the system and disconnect the probe before cleaning the probe. CAUTION Take extra care when handling the lens face of the Ultrasound transducer. The lens face is especially sensitive and can easily be damaged by rough handling.
Page 134
CAUTION Please do not use any gel or other materials which are not provided by CHISON. Un-authorized gel, lubricants and other materials may corrode probes and other parts of the device, for example the keyboard.
Page 135
ECO 6 Digital Color Doppler Ultrasound System Returning/Shipping Probes and Repair Parts Transportation dept. and our policy require that equipment returned for service MUST be clean and free of blood and other infectious substances. When you return a probe or part for service, you need to clean and disinfect the probe or part prior to packing and shipping the equipment.
Page 136
ECO 6 Digital Color Doppler Ultrasound System should also review the Centers for Disease Control and Prevention document on sterilization and disinfection of medical devices to be certain that their procedures conform to the CDC principles for disinfection of patient care equipment.
Page 137
ECO 6 Digital Color Doppler Ultrasound System the transducer’s performance periodically. Other agents such as quaternary ammonium compounds are not considered high level disinfectants and should not be used. Isopropanol is not a high level disinfectant when used as a wipe and probe manufacturers generally do not recommend soaking probes in the liquid.
ECO 6 Digital Color Doppler Ultrasound System 1998;69:409-11. Hignett M, Claman P. High rates of perforation are found in endovaginal ultrasound probe covers before and after oocyte retrieval for in vitro fertilization-embryo transfer. J Assist Reprod Genet 1995;12:606-9. Sterilization and Disinfection of Medical Devices: General Principles. Centers for Disease Control, Division of Healthcare Quality Promotion.
ECO 6 Digital Color Doppler Ultrasound System Put on medical sterile glove Get the condom for the package. Unfold the condom. Load some ultrasound gel into condom. Take the condom with one hand, and put the probe head into the condom.
If users install, use and maintain the system fully according to CHISON’s installation manual, operation manual and service manual, then CHISON ECO main unit has a life time of 5 years and CHISON ECO probes have life time of 5 years after ex-work.
Page 141
But if user still wants to continue using the system and probes, user should first contact CHISON service center at CHISON headquarter to arrange the necessary safety check and calibration by CHISON’s authorized service engineer.
ECO 6 Digital Color Doppler Ultrasound System Appendix B: Acoustic Output Report Table Acoustic Output Table System: ECO 6 Transducer Model: C3-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.92 Acoustic min of [P α...
Page 144
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: C3-A Operating Mode: THI-B Non-scan Index Label Aaprt > Scan Non-scan Aaprt≤1c Global Maximum Index Value Associated 0.94 Acoustic 2.01 min of [P α Parameter ta.α...
Page 145
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: C3-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.66 Acoustic min of [P α Parameter 10.27 ta.α...
Page 146
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: C3-A Operating Mode: THI-B+M Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.66 Acoustic 23.55 min of [P α Parameter 10.27...
Page 147
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: C3-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.64 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 148
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: C3-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated Acoustic min of [P α Parameter 22.03 ta.α 2.61 4.95...
Page 149
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: V6-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.91 Acoustic min of [P α Parameter ta.α Z at max.I 3.43...
Page 150
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: V6-A Operating Mode: THI-B Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.44 Acoustic min of [P α Parameter ta.α...
Page 151
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: V6-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.89 Acoustic min of [P α Parameter ta.α 3.97 Z at max.I...
Page 152
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: V6-A Operating Mode: THI-B+M Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.89 Acoustic min of [P α Parameter ta.α...
Page 153
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: V6-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated Acoustic min of [P α Parameter ta.α Z at max.I 2.03...
Page 154
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: V6-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.84 Acoustic min of [P α Parameter ta.α Z at max.I 3.47...
Page 155
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7M-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated Acoustic min of [P α Parameter ta.α Z at max.I 1.95...
Page 156
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7M-A Operating Mode: THI-B Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.69 Acoustic min of [P α Parameter ta.α...
Page 157
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7M-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.83 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 158
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7M-A Operating Mode: THI-B+M Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.83 Acoustic min of [P α Parameter ta.α...
Page 159
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7M-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.09 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 160
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7M-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.98 Acoustic min of [P α Parameter ta.α Z at max.I 3.35...
Page 161
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC3-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.72 Acoustic 4.32 min of [P α Parameter ta.α...
Page 162
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC3-A Operating Mode: THI-B Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.69 Acoustic 4.32 min of [P α Parameter ta.α...
Page 163
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC3-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.69 Acoustic min of [P α Parameter 1.29 ta.α...
Page 164
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC3-A Operating Mode: THI-B+M Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.51 Acoustic min of [P α Parameter 1.29 ta.α...
Page 165
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC3-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.03 Acoustic min of [P α Parameter ta.α Z at max.I 3.66...
Page 166
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC3-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.81 Acoustic min of [P α Parameter ta.α 3.65 Z at max.I...
Page 167
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: R7-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.81 Acoustic min of [P α Parameter ta.α Z at max.I 1.95...
Page 168
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: R7-A Operating Mode: THI-B Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.79 Acoustic min of [P α Parameter ta.α...
Page 169
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: R7-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.83 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 170
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: R7-A Operating Mode: THI-B+M Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.83 Acoustic min of [P α Parameter ta.α...
Page 171
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: R7-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.09 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 172
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: R7-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.98 Acoustic min of [P α Parameter ta.α Z at max.I 3.35...
Page 173
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC6-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.94 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 174
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC6-A Operating Mode: THI-B Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.49 Acoustic min of [P α Parameter ta.α...
Page 175
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC6-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.61 Acoustic min of [P α Parameter ta.α 1.95 Z at max.I...
Page 176
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC6-A Operating Mode: THI-B+M Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.61 Acoustic min of [P α Parameter ta.α...
Page 177
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC6-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.11 Acoustic min of [P α Parameter ta.α Z at max.I 2.55...
Page 178
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: MC6-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.49 Acoustic min of [P α Parameter ta.α Z at max.I 0.01...
Page 179
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7S-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.92 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 180
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7S-A Operating Mode: THI-B Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 0.99 Acoustic min of [P α Parameter ta.α...
Page 181
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7S-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.92 Acoustic min of [P α Parameter ta.α 1.55 Z at max.I...
Page 182
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7S-A Operating Mode: THI-B+M Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 1.81 Acoustic min of [P α Parameter ta.α...
Page 183
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7S-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 2.06 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 184
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: L7S-A Operating Mode: Non-scan Index Label Aaprt≤1c Aaprt > Scan Non-scan Global Maximum Index Value Associated 2.26 Acoustic min of [P α Parameter ta.α Z at max.I pi.α...
Page 185
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: P3-A Operating Mode: B Index Label non-scan Scan non-scan ≤1 >1 aprt aprt Global Maximum Index Value (MPa) 1.84 r,α (mW) 269.95 min of min of [Pα(Zs),Ita,α(Z...
Page 186
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: P3-A Operating Mode: THI-B Index Label non-scan Scan non-scan ≤1 >1 aprt aprt Global Maximum Index Value (MPa) 1.76 r,α (mW) 269.95 min of min of (mW) [Pα(Zs),Ita,α(Zs)]...
Page 187
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: P3-A Operating Mode: B+C Index Label non-scan Scan non-scan ≤1 >1 aprt aprt Global Maximum Index Value (MPa) 1.35 r,α (mW) 92.18 min of min of (mW) [Pα(Zs),Ita,α(Zs)]...
Page 188
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: P3-A Operating Mode: B+C+PW Index Label non-scan Scan non-scan ≤1 >1 aprt aprt Global Maximum Index Value (MPa) 1.25 r,α (mW) 122.88 min of [W...
Page 189
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: P3-A Operating Mode: M Index Label non-scan Scan non-scan ≤1 >1 aprt aprt Global Maximum Index Value (MPa) 1.64 r,α (mW) 89.20 min of [W...
Page 190
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: P3-A Operating Mode: B+C+CW Index Label non-scan Scan non-scan ≤1 >1 aprt aprt Global Maximum Index Value (MPa) 0.42 r,α (mW) 178.00 172.00 min of min of [Pα(Zs),Ita,α(Z...
Page 191
ECO 6 Digital Color Doppler Ultrasound System Acoustic Output Table System: ECO 6 Transducer Model: P3-A Operating Mode: CFM-M Index Label non-scan Scan non-scan ≤1 >1 aprt aprt Global Maximum Index Value (MPa) 1.51 r,α (mW) 254.00 min of [W...
IEC 61000-3-3 2. Guidance and manufacturer’s declaration – electromagnetic immunity The ECO 6 is intended for use in the electromagnetic environment the ECO 6 should assure that it is used in such an environment. Immunity test IEC 60601...
Page 193
3 Guidance and manufacturer’s declaration – electromagnetic immunity The ECO 6 is intended for use in the electromagnetic environment specified below. The customer or the user of the ECO 6 should assure that it is used in such an environment.
Page 194
If the measured field strength in the location in which the ECO 6 is used exceeds the applicable RF compliance level above, the ECO 6 should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the ECO 6.
Page 195
ECO 6 Digital Color Doppler Ultrasound System Rated maximum Separation distance according to frequency of transmitter output power of transmitter 0,01 0,12 0,12 0,23 0,38 0,38 0,73 For transmitters rated at a maximum output power not listed above, the recommended separation...
ECO 6 Digital Color Doppler Ultrasound System Appendix E: Display Accuracy and Acoustic Measurement Uncertainties According to IEC60601-2-37 and NEMA UD-3 2004, the display accuracy and acoustic measurement uncertainties are summarized in the table below. Display accuracy of MI is ± 20%, and TI is ± 40% or <0.1, if MI, TI below 0.5.
ISO Guide tout ye Expression of uncertainty in measurement. Three transducer samples have been tested and the confidence coefficient is at 95%,the value of t.975 is 4.30. The measurement data were obtained under the test conditions employed at CHISON. Transducer...
ECO 6 Digital Color Doppler Ultrasound System Appendix G: Procedures of set network sharing in ECO series For Windows set up, set up a shared document 1. Choose the file you want to share, as the “test” file. 2. Right click this file, choose “properties”, and click “share”.
Page 200
ECO 6 Digital Color Doppler Ultrasound System 5. If the windows have not set the code, then turn off the password protected sharing is necessary. instructions as the image shows. a) Click “network and sharing center” in the sharing properties.
Page 201
ECO 6 Digital Color Doppler Ultrasound System Set up in ECO IP set up 1. First, confirm the service address of shared files, you can get the IP address in windows interface. In windows “start”-“run” type in “cmd” and enter, then type in “ ipconfig” and enter, you can see the IP address of local service.
Page 202
ECO 6 Digital Color Doppler Ultrasound System CAUTION: Ping: test if the IP is connected or not. Clear: clear all the IP address, user name, password and names of shared file Update: update the content to the chosen item. Delete: delete the chosen service item.
Need help?
Do you have a question about the ECO 6 and is the answer not in the manual?
Questions and answers