3.1.4
ScanStation height
setup
step-by-step
30
4.
Turn on the instrument by pressing the ON/OFF button (f). Go to
Status, Level and Laser Plummet, Plummet and activate the laser
plummet (g).
5.
Move the tripod legs (a) and use the tribrach footscrews (h) to cen-
tre the plummet (i) over the ground point.
6.
Adjust the tripod legs (j) to level the circular level (k).
7.
By using the electronic level (Status, Level and Laser plummet,
Level) turn the tribrach footscrews (h) to precisely level the instru-
ment.
8.
Centre the instrument precisely over the ground point (i) by shifting
the tribrach on the tripod plate.
9.
Repeat steps 7. and 8. until the required accuracy is achieved.
Instrument Height
007870_001
To get an accurate height measurement use the GHM008 instrument height
meter in conjunction with the GHT196 distance holder. Both are included with
the scanner.
1.
Place tripod centrally over the ground point, level instrument.
2.
Click GHT196 distance holder to tribrach. It must "snap" onto the
cover over an adjusting screw.
3.
Unfold measuring tongue, pull out tape measure a little.
4.
Insert GHM008 instrument height meter in the distance holder and
attach.
5.
Swivel measure in the direction of the ground point, pull out until the
tip of the measuring tongue touches the point on the ground, keep
under tension and do not allow to sag, clamp if necessary.
6.
Read height of the instrument (ground - tilt axis) in the reading win-
dow at the red marking (in the example 1.627 m).
2
6
5
Operation