Calibration Guide

PREPARATION

The Navigator Balance has two calibration processes: Span and Linearity.
Follow the prompts for calibration mass values.
Note: This menu may be locked, either for security or for Legal for Trade applications. To regain
access, see Instruction Manual, Section 4.6. (For more detail, see Service Manual, Section 1.7.)
 Ensure that the balance is level, stable, and undisturbed during calibration.
(Otherwise the process could fail.)
 Allow balance to warm up for five minutes after stabilizing to room temperature.
 Use suitable calibration masses (available from Ohaus).
 Before beginning Calibration, be sure the "lock switch" is off.
After calibration, the balance returns to weigh mode.
Note: To cancel calibration, long-press:

SPAN CALIBRATION

5
Y
1
5
= Long Press
1
= Short Press
Ohaus Corporation www.ohaus.com
(Release button)
(Acquiring zero value)
(Place mass on platform)
(example for 200g capacity model)
Navigator
(See note above.)
Y
1
Y
1
i
Navigator
Balance
Calibration Guide

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the NAvigator and is the answer not in the manual?

Questions and answers

Summary of Contents for OHAUS NAvigator

  • Page 1: Span Calibration

    Navigator Balance PREPARATION The Navigator Balance has two calibration processes: Span and Linearity. Follow the prompts for calibration mass values. Note: This menu may be locked, either for security or for Legal for Trade applications. To regain access, see Instruction Manual, Section 4.6. (For more detail, see Service Manual, Section 1.7.) ...
  • Page 2: Linearity Calibration

    Be careful not to touch the scale or the table while calibration is in progress, as it will cause the process to fail. (Release button) (On some models, no need to press Yes) (Place mass on platform) (Place mass on platform) (Remove mass) *83033982* P/N 83033982  Navigator Calibration Guide Ohaus Corporation www.ohaus.com...

Table of Contents