Using the HP 81101A
The Mode/Trigger Screen

Triggered Burst Mode

The following figures show typical timings for trigger mode TRIGGERED
and pulse mode BURST. The bursts are triggered by the rising edge of the
arming source.
For the first example, the synchronously triggerable internal oscillator is
used to source the period.
ARMING SOURCE
PERIOD SOURCE
Characteristics
• A burst of pulse periods is triggered by an active edge at the selected
• You can select the number of pulse periods per burst in the range of
• You can select single pulses or double pulses per pulse period.
50
Trg'd by:
(Trg'd by:) EXT INPUT
or MAN Key
or PLL
Period
internal Osc.
OUTPUT
Single Pulses
OUTPUT
Double Pulses
TRIGGER OUT
STROBE OUT
1st period
arming source
– MAN key on front panel, triggered by press or release or both.
– EXT INPUT triggered by rising or falling or both edges.
– PLL (internally triggered bursts)
When using the internal PLL as the arming source, you must specify
the triggering period.
2–65536.
Oscillator starts synchronously
Burst of 3
last period
1st period