Malfunctions

9 Malfunctions

9.1 Error Messages

If an error message is active, dispensing is canceled and locked automatically.
Error message
Fault
Error 0105
Measured values
are not being
displayed.
Error 0140
The UV lamp is
not recognized.
Check UV lamp,
replacement may be
needed.
Error 0150
The device has
a leak.
Check device for
leakage.
Please contact your
service technician.
Error 0160
Operation of the
device is not
Front cover is open.
possible.
Please put the front
cover back on.
Error 0163
The pretreatment
cartridge is not
Check proper fit of the
recognized.
pretreatment cartridge
(R).
Error 0166
The ultrapure
water cartridge
Check proper fit of the
is not detected.
ultrapure water
cartridge (L).
66
Operating Instructions arium
Cause
There is an internal
communications
error.
The UV lamp is not
connected properly
or defective.
Water has leaked
inside the device.
The front cover is
not properly
attached.
The pretreatment
cartridge has not
been properly
inserted.
The ultrapure water
cartridge has not
been properly
inserted.
®
mini plus | arium
®
mini | arium
Correction
Disconnect the device from the AC power
and wait a minute.
Reconnect the device to the AC power.
If the error persists: Contact Sartorius Service.
Check whether the black plug of the
UV lamp is connected correctly.
Remove the UV lamp and check for damage.
If required: Change the UV lamp.
If the error persists: Contact Sartorius Service.
Remove the left side cover and check that
the three bag hoses are properly connected
to the device connectors. If required:
Remove the bag and reconnect it.
Drain the tray at the bottom of the device
and check that the connections are tight
during operation.
If the error persists: Contact Sartorius Service.
Correctly attach the front cover to the device.
If the error persists: Contact Sartorius Service.
Press the pretreatment cartridge in firmly
until you hear a distinct clicking sound.
If the error persists:
− Remove the pretreatment cartridge.
− Insert the pretreatment cartridge.
If the error persists: Contact Sartorius Service.
Press the ultrapure water cartridge in firmly
until you hear a distinct clicking sound.
If the error persists:
− Remove the ultrapure water cartridge.
− Insert the ultrapure water cartridge.
If the error persists: Contact Sartorius Service.
®
mini essential
Chapter, page
7.1, page 45
8.5.5, page 60
8.5.5, page 60
8.5.2, page 56
8.5.3, page 58
8.5.4, page 59