Troubleshoot Flow Rate Error
The flow rate is the speed in which fluids pass through the fluidics system (µl/min). It is
measured before each run during the pre-run check. If the system is unable to measure the
flow rate, pump a volume of reagent (MiSeqDx SBS Solution (PR2)) through the system
before checking the flow rate again.
1
Use the drop-down list or on-screen keyboard to enter the following information:
•
Solution: PR2
•
Volume: 250 µl
•
Aspirate Rate: 2500 µl/min
•
Dispense Rate: 2500 µl/min
2
Select Pump.
Figure 47
3
When the pump step is complete, select Restart Check.
4
If the error persists, set the volume to pump 500 µl MiSeqDx SBS Solution (PR2) and
repeat the process 1 more time. Contact Illumina Technical Support if the second
attempt does not resolve the error.
MiSeqDx Instrument Reference Guide
Measure Flow Rate
67