CALIBRATION PROCEDURE
SCXI -1303 T

Introduction

What Is Calibration?

Why Should You Verify?

How Often Should You Verify?

National Instruments
, ni.com
trademarks or trade names of their respective companies.
370253B-01
ERMINAL
This document contains information and step-by-step instructions for
verifying the performance of the temperature sensor on the
National Instruments SCXI-1303 terminal block. This temperature sensor
is for cold-junction compensation of thermocouples on the terminal block.
Calibration consists of verifying the measurement accuracy of a device and
correcting for any measurement error. For SCXI-1303 terminal blocks,
calibration is simply verifying the measurement accuracy of the
components on the terminal block. Because these components are not
user-adjustable, calibration consists of verification only, without correcting
for any error. Verification is measuring the performance of a device and
comparing the results to the factory specifications of the device.
The accuracy of electronic components drifts with time and temperature,
which can affect measurement accuracy as the device ages. Verification
ensures that your SCXI-1303 terminal block still meets
National Instruments standards. If the results of the procedure indicate that
the temperature sensor on your terminal block is out of specification, return
the sensor to National Instruments for repair or replacement.
The measurement accuracy requirements of your application determine
how often you should verify the performance of your SCXI-1303 terminal
block. National Instruments recommends you verify your terminal block
at least once every year. You can shorten this interval to six months or
90 days, based on the demands of your application.
, and SCXI
are trademarks of National Instruments Corporation. Product and company names mentioned herein are
© Copyright 2000 National Instruments Corp. All rights reserved.
B
LOCK
November 2000

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SCXI-1303 and is the answer not in the manual?

Questions and answers

Summary of Contents for National Instruments SCXI-1303

  • Page 1 ™ National Instruments , ni.com , and SCXI are trademarks of National Instruments Corporation. Product and company names mentioned herein are trademarks or trade names of their respective companies. 370253B-01 © Copyright 2000 National Instruments Corp. All rights reserved. November 2000...
  • Page 2 Software and Documentation You can find all the necessary information to verify the performance of the SCXI-1303 in this verification procedure. No other software or documentation is required. If you would like more information on the SCXI-1303, refer to the SCXI-1303 Terminal Block Installation Guide, which you can download from the National Instruments Web site at ni.com/manuals...
  • Page 3 Verification Procedure This section contains step-by-step instructions for verifying the performance of the temperature sensor on your SCXI-1303 terminal block. Verifying Temperature Sensor Performance Complete the following steps to verify the performance of the temperature sensor on your terminal block: Connect a +5 VDC power source to the terminal block.
  • Page 4 ------------ -   Calculate --------------------------------- - – 273.15 meas is in °C where T meas –3 a = 1.295361 × 10 –4 b = 2.343159 × 10 –7 c = 1.018703 × 10 SCXI-1303 Terminal Block Calibration Procedure ni.com...
  • Page 5 • If T meas nonfunctional. Do not substitute parts or modify equipment. Return the terminal block to National Instruments to ensure that the safety features are not compromised. > (T + 0.5 °C), the terminal block temperature sensor is •...
  • Page 6 A3 (Temp Sensor) A2 (GND) A1 (+5 VDC) Figure 1. SCXI-1303 Front Connector Pin Assignments SCXI-1303 Terminal Block Calibration Procedure ni.com...
  • Page 7 Jumper W1 Figure 2. Location of Jumper W1 on the SCXI-1303 Terminal Block © National Instruments Corporation SCXI-1303 Terminal Block Calibration Procedure...

Table of Contents