Installing the SLSC-12251/12252 and
RTI-12306
Figure 1. Installing the RTI-12306 into the SLSC Chassis
Caution
system is energized.
Note
Install the SLSC-12251/12252 module and its corresponding RTI-12306 in
the chassis at the same time to ensure proper mating alignment. Complete the
following procedure to ensure proper installation of the module and RTI-12306.
1.
Power off the main DC power source or disconnect the power source from the chassis
before installing any modules or RTIs.
2.
Ensure that the chassis is powered off. The POWER LED should be off. If the POWER
LED is not off, do not proceed until it is off.
Notice
plug-in. The entire chassis must be powered off when a module is inserted or
removed.
3.
Loosen the screws on the upper rear panel of the chassis.
Do not touch the contacts or remove the I/O boards or cables while the
The SLSC chassis and the SLSC-12251/12252 do not support hot
SLSC-12251 and SLSC-12252 User Manual | © National Instruments | 5