Terminal Blocks......................19 Overview This document contains information about the SLSC-12251 (16 channel, 8 amperes) and SLSC-12252 (8 channel, 30 amperes) Fault Insertion Units (FIU). The SLSC-12251 and SLSC-12252 are switch matrices with current monitoring, which allows for connections from each channel to any of the following: Bus A, Bus B, a load, or a paired channel.
920 V peak Caution If you are using the SLSC-12251/12252 with voltages greater than 60 , the SLSC chassis must be made touch safe by panel mounting it in a closed rack to prevent user access to the rear of the device.
The chassis internal ambient temperature may reach 85 °C with all slots at the maximum allowed power dissipation. In the SLSC-12001 chassis this corresponds to an external ambient of 40 °C. 4 | ni.com | SLSC-12251 and SLSC-12252 User Manual...
Position the RTI-12306 at the desired slot and insert the securing screws, but do not fully tighten them. Insert an SLSC-12251/12252 module into the same slot as its corresponding RTI-12306 while firmly holding the RTI-12306 in place until the RTI-12306 is firmly connected to the module.
Bus High Current Limit for longer than its delay time. NI.FIU.BoardFault.OvercurrentOnBusB Boolean Device True when Bus B exceeds the Bus High Current Limit for longer than its delay time. 8 | ni.com | SLSC-12251 and SLSC-12252 User Manual...
Page 10
Update command is required for the change to take place. Enumeration: 0 : "IndependentChannels" : Channel n is not connected to n + 1 1 : "ConnectedChannels" : Channel n is connected to n + 1 10 | ni.com | SLSC-12251 and SLSC-12252 User Manual...
For temperature measurement, read the property NI.FIU.TemperatureSensorN where N = 0,1,2,3. The physical locations of the temperature sensors are as follows: Figure 3. Temperature Sensor Locations on SLSC-12251 and SLSC-12252 Boards 12 | ni.com | SLSC-12251 and SLSC-12252 User Manual...
Averaging data in multiples of 0.1 seconds helps remove 50 Hz or 60 Hz noise. Repeat the previous step at five or more DC current points from ±8 A on the SLSC-12251 and ±30 A on the SLSC-12252. Repeat Step 3 and Step 4 for all channels.
Ch6_LOAD Ch0_LOAD Ch5_LOAD Ch4_LOAD Ch1_LOAD Ch3_LOAD Ch2_LOAD Note The input to each channel is Chx_DUT. The output to each channel is Chx_LOAD. NC indicates not connected and should not be used. 16 | ni.com | SLSC-12251 and SLSC-12252 User Manual...
Accessories RTI-12306 The RTI-12306 is the required rear transition interface for use with SLSC-12251 and SLSC-12252 devices. The RTI-12306 provides connections to Bus A, Bus B, and all current sensor outputs through the MDR connectors. RTI-12306 Front View Figure 6. RTI-12306 Front View...
Ground reference for all signals. Terminal Blocks The following terminal blocks connect to the LOAD connectors on the SLSC-12251/12252 through SPxM-SPxF cables and provide easy spring terminal connections to the LOAD channels. The low voltage (<33 V) terminal block options also provide an HD44 connection for routing low current sense LOAD signals to an SLSC-12201 module.
Page 20
CH10 DGND TB-12397 The TB-12397 is a terminal block for use with the SLSC-12251 when all signals are <33Vdc. The HD44 connector J4 allows for sense connections (low current only) of the channels to another device such as the SLSC-12201.
Page 22
The HD44 connector J4 allows for sense connections (low current only) of the channels to another device such as the SLSC-12201. Figure 12. TB-12399 Front View CH0 CH1 CH5 CH6 DGND 22 | ni.com | SLSC-12251 and SLSC-12252 User Manual...
Page 24
Figure 15. Terminal Block Rear Hole Pattern 76.2 mm (3.00 in) 38.1 mm (1.50 in) 4x M4 X 0.7 Threads 7.0 mm [0.276 in] Max Insertion 38.1 mm (1.50 in) 76.2 mm (3.00 in) 24 | ni.com | SLSC-12251 and SLSC-12252 User Manual...
NI trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering NI products/technology, refer to the appropriate location: Help»Patents in your software, the file on your media, or the National Instruments Patent Notice at . You can find patents.txt ni.com/patents...