Table 3.1 Diagnostics/User Messages Chart (Continued)
Message
D605 - Drive
D606 - Drive
D607 - Drive
D608 - Drive
D701 - Drive
D804 - Door
D806 - Door
D808 - Door
Other Possible Problems
Operating problems that may not be indicated by diagnostic messages are described in
along with probable causes, listed in the probable order of occurrence, and corrective actions.
Perform the recommended corrective action in sequence, as listed. If you are unable to correct the
problem,
contact
PN B80286AC
Definition/Result
Drive Performance
Rotor coasts to a stop.
Drive Performance
Rotor coasts to a stop.
Drive Performance
Rotor brakes to a stop per
profile. Door remains locked
for 2 minutes.
Drive Performance
Rotor coasts to a stop. Door
remains locked for 45 minutes.
Imbalance Detector
Rotor brakes to a stop per
profile.
Door Latch
Rotor brakes to a stop at
maximum rate.
Door Latch
Rotor brakes to a stop at
maximum rate.
Door Latch
us.
Other Possible Problems
Recommended Action
Contact us
• Confirm the rotor is tied down.
• If the problem persists,
• Confirm the drive shaft rotates when
spun by hand.
• Confirm that a rotor is installed in the
chamber and the rotor is tied down.
• If the problem persists,
Contact us
Contact us
• Confirm that the door is fully closed.
• If the problem persists,
• Confirm that the door is fully closed.
• If the problem persists,
• Check for door obstructions.
• If the problem persists,