National Instruments SCC-68 User Manual

National Instruments SCC-68 User Manual

I/o connector block with 4 scc signal conditioning slots for daq devices
Hide thumbs Also See for SCC-68:

Advertisement

USER GUIDE
SCC-68
I/O Connector Block with 4 SCC Signal Conditioning Slots for DAQ Devices
Contents
The SCC-68 is an I/O connector block for easy signal connection to a
National Instruments M Series or E Series DAQ device. The SCC-68
features screw terminals and a general breadboard area for I/O signal
connection, and bus terminals for external power and grounding. The
SCC-68 has four SCC slots for use in integrating thermocouple, strain
gauge, RTD input, frequency input, current input, voltage attenuator,
lowpass filter, load cell input, accelerometer, feedthrough, SPDT relay,
isolated digital I/O, and isolated voltage input SCC modules into the
measurement system.
Conventions ............................................................................................ 3
Using the SCC-68 Enclosure .................................................................. 4
Opening the Enclosure..................................................................... 4
Closing the Enclosure ...................................................................... 5
Stacking and Wall Mounting ........................................................... 6
Installing Rubber Feet...................................................................... 7
Connecting to the DAQ Device .............................................................. 8
SCC-68 Features ..................................................................................... 11
Screw Terminals .............................................................................. 12
Device Pinouts .......................................................................... 12
Wiring to Screw Terminals....................................................... 12
Integrated Strain Relief .................................................................... 12
SCC Modules................................................................................... 13
Breadboard Area .............................................................................. 13
Compensation (CJC) ..................................................................... 14

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SCC-68 and is the answer not in the manual?

Questions and answers

Summary of Contents for National Instruments SCC-68

  • Page 1: Table Of Contents

    SCC-68 I/O Connector Block with 4 SCC Signal Conditioning Slots for DAQ Devices The SCC-68 is an I/O connector block for easy signal connection to a National Instruments M Series or E Series DAQ device. The SCC-68 features screw terminals and a general breadboard area for I/O signal connection, and bus terminals for external power and grounding.
  • Page 2 Bus Terminals...................15 Power Supply..................16 Using SCC Modules with the SCC-68 ............16 Supported SCC Modules ..............17 Connecting Signals to SCC Modules in the SCC-68 .......17 Power Supply Considerations............19 LED ...................20 Installing the SCC Modules in the SCC-68........21 SCC Connector Pinouts.............21 Direct Access to SCC signals............23 Analog Input Measurement Considerations ..........23...
  • Page 3: Conventions

    This font is also used for the proper names of disk drives, paths, directories, programs, subprograms, subroutines, device names, functions, operations, variables, filenames, and extensions. SCC-XX This name refers to any SCC module. © National Instruments Corporation SCC-68 User Guide...
  • Page 4: Using The Scc-68 Enclosure

    To open the SCC-68, use a flathead screwdriver to loosen the captive screw at the front of the device. After the screw is loosened, pull up on the front of the enclosure cover to open. The top half of the SCC-68 enclosure can be removed completely.
  • Page 5: Closing The Enclosure

    Closing the Enclosure To close the SCC-68, align the two back tabs into the corresponding holes at the back of the base, as shown in the detail in Figure 2. With the cover and base of the enclosure aligned, snap the SCC-68 closed. To secure the SCC-68 enclosure, use a flathead screwdriver to push the screw down against the spring resistance and tighten the screw.
  • Page 6: Stacking And Wall Mounting

    SCC-68 connector blocks. Note The SCC-68 cannot be securely stacked onto another SCC-68 if the rear rubber feet are attached. The SCC-68 also has mounting holes on the bottom of the device for secure wall-mounting.
  • Page 7: Installing Rubber Feet

    Rubber non-skid feet are shipped with the device that can be attached for secure desktop use. To install the rubber feet, remove adhesive backing and affix to the SCC-68 where indicated in Figure 4. 1 Rubber Feet 3 Stacking Grooves 2 Wall-Mounting Holes Figure 4.
  • Page 8: Connecting To The Daq Device

    Connecting to the DAQ Device The following sections describe how to connect the SCC-68 to the DAQ device. The SCC-68 is designed to be used with M Series and E Series DAQ devices only. Other 68-pin devices may not function properly with the SCC-68.
  • Page 9: Using The Scc-68 With A 68-Pin E Series Device

    Attach the SCC-68 to Connector 0 of the DAQ device using one of the following cables: • SHC68-68-EPM • SHC68-68 • RC68-68 Proceed to the Taking Measurements with NI-DAQmx section for information about using the SCC-68 in measurement applications. To use the SCC-68 with a 32 AI channel M Series device using...
  • Page 10: Using The Scc-68 With The Ni 6025E, Ni 6033E, Or Ni 6071E

    Using the SCC-68 with the NI 6025E, NI 6033E, or NI 6071E To use the SCC-68 with the NI 6025E, NI 6033E, or NI 6071E, complete the following steps: Attach the SCC-68 to the DAQ device using an SH1006868 cable. This cable has two 68-pin connectors labeled MIO 16 and Extended I/O.
  • Page 11: Scc-68 Features

    SCC-68 Features The following sections describe the hardware features of the SCC-68. FOR PATENTS: NI.COM/PATENTS FRONT 4.75-5.25V 2A MAX Mod 1 Mod 2 Mod 3 Mod 4 USE SCC CONNECTOR BACKSHELL FOR HAZARDOUS LIVE SIGNALS (>42.4 Vpk / 60VDC) SCC-68...
  • Page 12: Screw Terminals

    By lacing the connector wires through these posts, movement is limited. Note Route all SCC signals through the left wire-entry side of the SCC-68. Route all other signals through the right wire-entry side. SCC-68 User Guide...
  • Page 13: Scc Modules

    Breadboard Area The SCC-68 features a general breadboard area for custom circuitry such as filtering or attenuation. For soldering components to the underside of the breadboard, remove the printed circuit board by carefully snapping it loose from the SCC-68 enclosure.
  • Page 14: Using The Temperature Sensor For Cold-Junction Compensation (Cjc)

    Figure 8. SCC-68 Temperature Sensor Connection To use the temperature sensor for cold-junction compensation of thermocouple measurements, connect CJC+ (terminal 70) to AI channel 7 of the SCC-68. Figure 9 shows the temperature sensor connected to AI 7. CJC+ AI GND...
  • Page 15: Cold Junction Temperature Measurement Accuracy

    The bus terminals provide an easy interface for supplying external power, extra ground terminals, or grounding to any sensors or custom circuits used in the SCC-68. There are two buses—Bus A and Bus B—where all four screw terminals on each side are connected together internally. To use these...
  • Page 16: Power Supply

    I/O SCC modules. The SCC-68 is not compatible with analog output or counter SCC modules. Note If you use analog input SCC modules with the SCC-68, you must connect AI SENSE to AI GND, as shown in Figure 11. AI SENSE AI GND Figure 11.
  • Page 17: Supported Scc Modules

    Supported SCC Modules The SCC-68 supports all analog input and digital I/O SCC modules. Table 1 shows all SCC modules that are compatible with the SCC-68. Table 1. SCC-68-Compatible SCC Modules SCC Module Type SCC-AI0x Isolated Voltage Input SCC-TC0x Thermocouple Input...
  • Page 18 Figure 12. SCC Connector Backshell Refer to the SCC-XX user documentation for additional safety information. To connect an SCC module to the SCC-68, complete the following steps: Remove power from the signal lines. Strip insulation from the ends of the signal wires. Refer to the SCC-XX user documentation for the exact strip length for the module.
  • Page 19: Power Supply Considerations

    You are using +5V to power custom circuitry on the breadboard area If the +5V power from the DAQ device drops below about 4.5 V, SCC modules installed in the SCC-68 will not operate properly. The LED in the SCC-68 will flash seven times per second to indicate this condition. There is no error or notification in software.
  • Page 20: Led

    SCC-68 Figure 14. SCC-68 Power Distribution Block Diagram The LED in the SCC-68 shows the status of the power supply. The SCC-68 monitors the voltage of the +5V signal from the DAQ device, as shown in Figure 14. Table 2 shows the correlation between power levels in the SCC-68 and the LED status.
  • Page 21: Installing The Scc Modules In The Scc-68

    Installing the SCC Modules in the SCC-68 To install the SCC modules on the SCC slots of the SCC-68, plug the SCC modules onto the appropriate connector block sockets. When properly oriented, SCC modules plug easily onto the connector block socket. Never force an SCC module onto the socket.
  • Page 22 SCC-68 User Guide ni.com...
  • Page 23: Direct Access To Scc Signals

    If you are using SCC modules for some of the AI signals, you can also measure other AI signals using the AI screw terminals in the SCC-68. Refer to Figure 6 for the location of the AI screw terminals.
  • Page 24: Differential

    A single-ended connection is a connection in which the device AI signal is referenced to a ground that it can share with other input signals. National Instruments M Series and E Series devices provide two single-ended connection configurations. SCC-68 User Guide...
  • Page 25: Non-Referenced Single Ended (Nrse)

    SCC-68, the following cases should also be noted. Grounded Signal Sources If you have SCC modules installed in the SCC-68, differential measurements are recommended over NRSE measurements. When you have SCC modules installed, connect AI GND to AI SENSE. If you connect the ground of the signal source to AI SENSE, you are also shorting it to AI GND.
  • Page 26: Floating Signal Sources

    AI GND causes an error when measuring the signal. The amount of this error depends on several factors including the length of the cable connecting the DAQ device to the SCC-68 and the amount of current being drawn by the SCC-68 from the DAQ device +5V power supply.
  • Page 27: Taking Measurements With Ni-Daqmx

    If RSE is used to measure floating signals, there will be a measurement error due to the current being drawn by the SCC-68 from the DAQ device +5V power supply. This potential will vary depending on the length of the cable connecting the DAQ device to the SCC-68 and the amount of current being drawn from the +5V power supply by the SCC-68.
  • Page 28: Configuring The Scc-68

    Configuring the SCC-68 The SCC-68 can be configured as an SCC carrier and as a screw terminal accessory. To configure the SCC-68 as both an SCC carrier and a screw terminal accessory, you need to follow the configuration instructions in the Configuring the SCC-68 as an SCC Carrier section and the Configuring the SCC-68 as a Screw Terminal Accessory section.
  • Page 29: Configuring Channels And Tasks

    Configure a Task When using NI-DAQmx, configure tasks with the DAQ Assistant. To create tasks using the SCC-68, you must have NI-DAQmx version 8.1 or later. • To launch the DAQ Assistant in MAX, right-click Data Neighborhood and select Create New.
  • Page 30 The DAQ Assistant prompts you to create a new task. Select an I/O type, such as analog input. For thermocouple measurements, you must first ensure the hardware is configured correctly and that the SCC-68 has been configured in MAX. For more information, refer to the Configuring the SCC-68 section.
  • Page 31: Configure Global Virtual Channels

    CJC sensor is wired to AI 7, then select Built-In for the CJC source. If this option is not available, check your configuration. For more information, refer to the Configuring the SCC-68 section. For CJC sensor wiring instructions, refer to the...
  • Page 32: Using Your Task In An Application

    Refer to and enter the information code ni.com/info rddq73 step-by-step instructions for using a task in LabWindows/CVI. Measurement Studio Refer to and enter the information code ni.com/info rddqms step-by-step instruction for using a task in Measurement Studio. SCC-68 User Guide ni.com...
  • Page 33: Specifications

    Typical ..........1 mA with no signal conditioning installed Note The maximum power consumption of the SCC-68 is a function of the SCC modules installed and any circuits constructed on the general-purpose breadboard area. Refer to the Power Supply Considerations section for more information.
  • Page 34 C = 9.4978060 × 10 –8 ------------------- - – = 4,870 Ω where is the CJC voltage measurement For more information about the temperature sensor, refer to the Using the Temperature Sensor for Cold-Junction Compensation (CJC) section. SCC-68 User Guide ni.com...
  • Page 35 17.78 cm (2.09 in.) (7.00 in.) 24.46 cm (9.63 in.) Figure 20. SCC-68 I/O Connector Block Dimensions Weight ............ 623.7 g (1 lb 6 oz) I/O connectors ........One 68-pin male SCSI connector SCC sockets Number of sockets ......4 SCC types supported.......
  • Page 36 Channel-to-earth ........11 VDC, Measurement Category I Caution Do not use the SCC-68 for connections to signals or for measurements within Categories II, III, or IV. Environmental Operating temperature ......0 to 55 °C Storage temperature ........–20 to 70 °C Humidity ..........5 to 90% RH, noncondensing...
  • Page 37 Nonoperating random vibration profiles for the SCC-68 exceeds the requirements of Note MIL-PRF-28800F, Class 3. Safety This product is designed to meet the requirements of the following standards of safety for electrical equipment for measurement, control, and laboratory use: •...
  • Page 38: Quick Reference Guides

    This section provides the quick reference guides for use with Connector 0 and Connector 1 of 68-pin M Series devices, and E Series devices. The quick reference guides provide information about pin placements, their corresponding signal assignments, and wire-stripping length for signal connections. SCC-68 User Guide ni.com...
  • Page 39 © National Instruments Corporation SCC-68 User Guide...
  • Page 40 SCC-68 User Guide ni.com...
  • Page 41 © National Instruments Corporation SCC-68 User Guide...
  • Page 42 Instruments trademarks. Other product and company names mentioned herein are trademarks or trade names of their respective companies. For patents covering National Instruments products, refer to the appropriate location: Help»Patents in your software, the patents.txt file on your CD, or ni.com/patents.