Quick User guide
• Fridge-tag
2
®
• Fridge-tag
2 L
®
2
Read history (record daily checks (2x) in memory and clear an X if alarmed)
Recording daily checks
Ext. Sensor
max.
55
0 00
0
°C
2
Press the READ button 3 times to record your check
• 1st press: see UP arrow blink = HI temp. so far today
• 2nd press: see DOWN arrow blink = LO temp. so far
today
• 3rd press: records in memory the time of your manual check
• A Fridge-tag
2 records an AM or PM for morning or
®
afternoon checks
• A Fridge-tag
2 L records the actual time it is checked
®
in the morning and in the afternoon.
3
Downloading the report
STEP 1
Download the report from the Fridge-tag
connecting it to any PC / MAC.
2 = 60 day report
• Fridge-tag
®
• Fridge-tag
2 L = 56 day report (with 8 weekly graphs)
®
1
Reading the display during measurement
OK display
The following information is visible on screen:
!
•
temperatures are OK
• Current date: 06.19.2015
• Current time in military format:13:47
• Current temperature: +5.6 °C
Clearing the X /
X
Example of a LO ALARM
Press READ
3 times
Press the READ button 4 or 5 times to clear the X and
!
record your check.
• There will be an UP
today in the upper right of the screen indicating a HI or
LO ALARM
If both arrows are present then it is typically an ALARM
for the external sensor being disconnected for more
than 15 minutes. Double arrows need 5 presses of
READ to clear the X.
STEP 2
Connect the
Fridge-tag
2 / 2L via USB
®
with your PC / MAC
device by
®
Click on the Fridge-tag file to open it
0
6
19
2
0
1
5
056
1 47
3
°C
2
0
9
19
2
0
1
5
1st
!
1 21
8
Press
2
or DOWN
arrow under
!
STEP 3
ALARM display
0
6
!
1 42
5
If an alarm occurs the following additional symbols appear on screen:
•
ALARM symbols, temperature alert
!
!
•
= Max higher (HI) temperature ALARM
!
•
= MIN lower (LO) temperature ALARM
!
•
= ALARM sensor disconnected for +15 minutes
The arrow appears under today and / or the day an
ALARM occurs
0
9
19
!
1 21
8
!
0 35
1
!
0 35
1
1st press: see UP arrow blink = HI temp. so far today
2nd press: see DOWN Arrow blink = date & time of the
alarm
3rd press: see DOWN arrow blink & dur = duration of
ALARM and the lowest temperature during that alarm
4th press: clears the X and records in memory
Double click on the PDF file to open
the report. Note: store reports for
3 years to be VFC compliant
Berlinger USA LLC
222 Turnpike Road
Suite 3
Westborough, MA 01581
USA
Headquarters: Berlinger & Co. AG, 9608 Ganterschwil, Switzerland
19
2
0
1
5
088
°C
2
2nd
2
0
1
5
press
2
Ext.
Sensor
3rd
-011
press
min.
°C
2
Ext.
Sensor
-011
min.
°C
4th
2
press
4
Support
To change date / time or
Celsius to Fahrenheit, view
videos at:
www.berlinger.com/videos
Tel. +1 508 366 0084
Fax +1 508 366 0087
info.us@berlinger.com
www.berlinger.com
20160413_V1
Need help?
Do you have a question about the Fridge-tag 2 and is the answer not in the manual?
Questions and answers