Status Reporting System
STATus QUEStionable:FREQuency Register
This register comprises information about the reference and local oscillator.
It can be queried with commands STATus:QUEStionable:FREQuency:CONDition? and "STATus
:QUEStionable:FREQuency[:EVENt]?.
Table 5-6
Meaning of bits in STATus:QUEStionable:FREQuency register
Bit No.
Meaning
0
OVEN COLD
This bit is set if the reference oscillator has not yet attained its operating temperature. 'OCXO' will then be
displayed.
1
LO UNLocked (Screen A)
This bit is set if the local oscillator no longer locks. 'LO unl' will then be displayed.
2
LO LEVel (Screen A)
This bit is set if the level of the local oscillator is smaller than the nominal value. 'LO LVL' will then be displayed.
3
not used
4
not used
5
not used
6
not used
7
not used
8
not used
9
LO UNLocked (Screen B)
This bit is set if the local oscillator no longer locks.' LO unl' will then be displayed.
10
LO LEVel (Screen B)
This bit is set if the level of the local oscillator is smaller than the nominal value. 'LO LVL' will then be displayed.
11
not used
12
not used
13
not used
14
not used
15
This bit is always 0.
1065.6016.12
5.26
FSE
E-16