LCD Messages
A10 ERROR 7
ATTENTION
A10 ERROR 8
A10 ERROR 9
CHECK A10 COM
FLOW AUTO STOP
Alarm stops or
1. What it means.
does not stop
2. What to do.
Milli-Q System
Not Alarm Stop
1. The temperature inside the A10 during its Analysis Mode
2. If the message continues, then contact Millipore.
If you cancel any A10 ERROR (i.e. A10 ERROR 3) using the
Keypad (see Section 6-3.6), go to STANDBY Mode and then
back to READY Mode. Dispense water for at least 9 minutes.
A new TOC Analysis is done during this time. If the new TOC
value is satisfactory, then proceed to use the Milli-Q Water.
Contact Millipore if the A10 ERROR returns after 1 hour.
Not Alarm Stop
1. The TOC sample oxidation was not completed in the
2. If the message continues, then contact Millipore.
Not Alarm Stop
1. The A10 is not detecting a TOC value.
Not Alarm Stop
1. The communication between the A10 TOC Monitor and
2. Contact Millipore. When this is fixed, dispense water for
Alarm Stop
There are two different reasons for this Alarm. These are listed
below.
1. The Q-POD Dispenser is not available for dispensing. See
2. Push the Q-POD Dispenser Plunger all the way down and
Alert and Alarm messages
exceeded 55°C. Unplug the power cord, then plug it in to
power on the Milli-Q System. Dispense water for several
minutes.
allotted time.
caused by:
The A10 Solenoid Valve is not closing and could have a
particle stuck in it or
The A10 Lamp is not turning on.
Perform an A10 Cleaning Mode. This might dislodge a
stuck particle or replace the A10 Lamp.
the Milli-Q System PC Board is interrupted. The TOC value
is no longer reported.
at least 9 minutes. A new TOC Analysis is done during this
time. If the new TOC value is satisfactory, then proceed to
use the Milli-Q Water.
Section 4-6. Or
The Milli-Q System has automatically stopped dispensing
water. The POD FLOW STOP timer has reached 0 minutes.
release. This resets the dispenser timer and makes the Q-
POD Dispenser available for dispensing. See Section 3-
18.4 for information about changing this timer.
Section 4-6 for more information.
This can be
See
85