Canister Full Therapy Interrupted Alarm

Canister Full
Therapy Interrupted
WARNING: If Canister is full, replace and press 'Reset'.
If not full, press 'Cancel'. Press '?' for more information
Audio Pause
To resolve this alarm:
1. Check fluid level of canister by holding the therapy unit so that the Graduated Marks on the
canister are level and parallel to the floor.
Graduated Marks
A full canister is approximately 300 ml.
2. If canister is not full, press "Cancel".
3. If canister is full, change canister and press '
Screen'. See the
4. Restart therapy by pressing
WARNING: Under clinician supervision, replace V.A.C. Dressing with alternate
dressing if therapy is interrupted for more than two hours.
®
Alerts and Alarms (cont.)
This alarm screen appears when the ActiV.A.C. Therapy Unit
Cancel
detects that the canister is full and should be replaced.
This alarm will be accompanied by a repeating audible tone.
Reset
This alarm may be silenced for two (2) minutes during
troubleshooting by pressing
Canister Change
section of this manual for additional information.
On/Off
'Audio Pause'
Reset
' on this screen to return to the 'Home
.
®
.
®

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the ActiV.A.C. and is the answer not in the manual?

Questions and answers

william Barrett
April 12, 2025

how long does the battery last without needing to be recharged.

1 comments:
Mr. Anderson
April 12, 2025

The battery life of the KCI ActiV.A.C. is approximately 14 hours, depending on settings.

This answer is automatically generated

Table of Contents

Save PDF