Preparing the Unit and Setting the Acquisition Parameters
for the Pediatric Patient
To acquire a frontal AP or PA image, follow these steps:
1. Position the head clamps manually for the frontal AP or PA exam.
2. In the Cephalometric Acquisition interface, click the Program button to access the
Program pane. In the Program pane:
The
for a frontal AP exam is active.
Click
for a frontal PA exam.
Select the appropriate acquisition format option.
3. Click the Patient button to access the Patient pane. Select the pediatric patient
type.
4. If the default parameter setting is not adapted to your pediatric patient type, click the
Parameter button and select the appropriate parameters. To save the new parameter
settings as the default settings, click
IMPORTANT
You must position the head clamps manually because they
are not positioned automatically from the Program pane exam
type selection. In this case, the relevant exam type selection
icon becomes active.
and select Memorize settings.
Acquiring a Frontal AP or PA Image
CS 9300C User Guide (SM750)_Ed 01
5–9