New laser model for patient centering. Spare Parts update and drawings added. New Owandy Radiology logo. (Ref. RDM 7086, RDM 7100, RDM 7121, RDM 7162, RDM 7192, RDM 7205, RDM 7249, RDM 7273) 15.01.13 14.01.14 Modification Owandy Radiology address and logo...
Page 4
SERVICE MANUAL Revision history The manufacturer OWANDY RADIOLOGY reserves the right to make modifications to its products or to their specifications in order to improve the performance, quality, or ease of production. Specifications of products or accessories may be modified without prior notice.
SERVICE MANUAL Contents Contents INTRODUCTION 1.1. Icons appearing in the manual ............1-2 1.2. Technical service ................. 1-2 SAFETY INFORMATION 2.1. Warnings .................... 2-2 2.2. Environmental risks and displacement ..........2-5 2.3. Symbols used ..................2-6 DESCRIPTION ...
Page 6
E280: Unexpected activation of zero position sensor on CEPH / E281: Timeout of zero position sensor on CEPH / E282: Zero position sensor for CEPH never active / E283: Zero position sensor for CEPH always active ......8-10 I-Max Touch...
Page 7
SERVICE MANUAL Contents 8.2.3. Errors with code from E300 to E399 ..........8-11 8.2.3.1. E300: Unexpected activation of zero position sensor on secondary CEPH collimator / E301: Timeout of zero position sensor on secondary CEPH collimator / E302: Zero position sensor for secondary CEPH collimator never active / E303: Zero position sensor fr secondary CEPH collimator always active ................
Page 8
Error due to a bad patient's positioning ........... 8-57 8.6.3. Images with artefacts ............... 8-64 SCHEMATICS AND DRAWINGS SPARE PARTS 10-1 APPENDIX 11.1. A ppendix A: Setup parameters table ............. 18 I-Max Touch...
The device must be used complying with the procedures described and never be used for purposes different from those herewith indicated. The I-Max Touch is an electro-medical device and it can be used only under the supervision of a physician or of highly qualified personnel, with the necessary knowledge on X-ray protection.
This icon indicates a WARNING message; the items marked by this icon refer to the safety aspects of the patient and/or of the operator. 1.2. Technical service For any technical queries please contact the technical service of your vendor. I-Max Touch...
X-rays generating units. Owandy Radiology cannot to be held responsible for: use of the I-Max Touch different than the intended use, damages to the unit, to the operator, to the patient, caused both by installation and maintenance procedures different than those described in this manual and/or by wrong operations, ...
Compatibility rules in force. Non-complying instruments, of which the poor immunity from electromagnetic fields is well known, must be installed at least 3 m (9.8 ft) from the I-Max Touch and supplied by a different electrical line. The I-Max Touch must be switched off while using devices such as electrical lancets or the like.
Page 13
1.5 m (4.9 ft) both from the X-ray source and from the patient, as shown in the picture below. Minimum distance from X-ray source 1.5mt Protected area Figure 2-1 - Panoramic version Minimum distance from X-ray source 1.5mt Protected area Figure 2-2 - Cephalometric version I-Max Touch...
Page 14
The system construction does not allow the repair of faulty/damaged parts that must be replaced using original spare parts supplied by Owandy Radiology. Only trained personnel are authorized to make service interventions on the unit, following the instructions contained in this manual.
SERVICE MANUAL Safety information 2.3. Symbols used In this manual and on the I-Max Touch itself, apart from the symbols indicated on the control panel, the following icons are also used: Symbols Description Device with type B applied parts The device contains in some of its parts, materials and liquids that at the end of the unit’s life, must be...
Description 3.2. Function, Models and Version The I-Max Touch, produced by Owandy Radiology, is a complete panoramic system, which enables to perform all radiographies commonly necessary in dental field (except for endoral X-rays). In some versions, certain examination modes are not available but the device (thanks to its computerised control system) can be expanded and updated with new releases, directly at the Dentist premises.
High Voltage from 60kV to 86kV, with 2kV steps, the anodic current from 6mA to 12mA with 1mA steps. The positioning of the sliding primary collimator, the secondary collimator and the Digital Sensor (inside the relative sensor holder) is automatic. I-Max Touch...
Block diagram This paragraph provides a brief description, at block diagram level, of the I-Max Touch. Aim of this paragraph is to provide a brief description of the system. More details about the electronic circuits which compose the system can be obtained by analyzing the schematics provided in Chapter 9.
The power supply assembly module also includes the Column CPU board (A1) that is dedicated to the control of the vertical column motion during all phases (slow speed, ramp up to high speed, ramp down from high to low speed, end run microswitches control, etc.). I-Max Touch...
The circuits of the CPU board are supplied starting from the +24Vdc (LED H12) provided by the Power supply assembly and generating on board the requested voltages (+5V, +3.3V and +1.5V). Three LED's on the board indicate the presence of these 3 voltages (+5V=LED H2, +3.3V=LED H3, +1.5V=LED H5). I-Max Touch...
Generator board and relevant Generator CPU. Possible anomalous conditions are then communicated to the main CPU board (A5) which in turn generates error codes to alert the operator. I-Max Touch...
(English, Italian, French, German, Spanish, Portuguese and Dutch). The language selection is only available for the messages dedicated to the user. The messages relative to the service programs (Password) are always in English. I-Max Touch 3-10...
3.4. Control panel - Descriptions and functions The I-Max Touch keyboard is divided into function areas, plus a display to view the operative messages and error signals. The next figure shows a general view of the keyboard, while details on each functional area are provided in the following pages.
Page 29
"Zoom in" button Acquired image Figure 3-3 WARNING: The USB port on the keyboard MUST NOT be used with an external Hard Disk with own mains connection. It has to be used only with USB Pen Drives. 3-13 I-Max Touch...
Page 30
">O<" (accept) button Navigational buttons Figure 3-4 WARNING: The USB port on the keyboard MUST NOT be used with an external Hard Disk with own mains connection. It has to be used only with USB Pen Drives. I-Max Touch 3-14...
Page 31
Protruded protruded, standard or retracted, as indicated within the button. Standard The arch selection does not influence the values of kV and mA but acts on the position of the focus layer. Retracted 3-15 I-Max Touch...
Page 32
ON/OFF the laser centring devices that allow the correct positioning of the medial-sagittal and Frankfurt planes, by adapting the I-Max Touch to the patient's anatomy. The key "Test" is used to avoid the Test OFF X-rays emission, in order to check the absence of collisions with the patient.
Page 33
(up, down, left, right). In the setup screen the following buttons are used to change parameters as indicated in the manual. 3-17 I-Max Touch...
> Repositioning the rotation group (to bring the Other group to the initial position after the examination 17 - Display acquired image key and to exit from the "making an exposure") 18 - USB Pen Drive key mode > Confirmation I-Max Touch 3-18...
Page 35
20 - Zoom out on the image Other 21 - Zoom in on the image 18 - USB Pen Drive key 19 - Return to control panel (main menu) USB port: to connect a pen drive 3-19 I-Max Touch...
Description 3.5. Service programs description The I-Max Touch allows the authorised technicians to access to the different functional parameters of the unit through a dedicated software composed by the following service programs. Each service program can be accessed through a dedicated password.
Page 38
SERVICE MANUAL Description THIS PAGE IS INTENTIONALLY LEFT BLANK I-Max Touch 3-22...
SERVICE MANUAL Technical data TECHNICAL DATA General features Type I-Max Touch Manufacturer OWANDY RADIOLOGY Croissy-Beaubourg, France Class Class II B for European Directive for Medical Devices 93/42 Class II for Canadian MDR Class I with type B applied parts according to IEC 60601-1...
Page 40
Sinus P/A projection Cephalometry and Carpus Normal resolution cephalometry in Latero-Lateral or Antero-Posterior projection (different formats) High resolution cephalometry Latero- Lateral or Antero-Posterior projection (different formats) High Resolution Carpus exam Motorized Soft Tissue Filter. I-Max Touch...
Page 41
> 3.1 mm Al eq. @ 80 kV Transformer insulation Oil bath Cooling By convection Leakage radiation at 1 m < 0.5 mGy/h @ 86 kV - 12 mA - 3 s duty cycle 1/16 Tube-head maximum thermic capacity 310 kJ I-Max Touch...
Page 42
Frankfurt planes (please refer to relevant paragraphs for detailed explanation). Wave length 650 nm ± 10 nm Divergence < 2.0 mRad Optical power < 1 mW Classification Class 1 laser product according to IEC Standard 60825-1:1993 + A1:1997 + A2:2001 I-Max Touch...
Page 43
Relative working humidity (RH) range 30% 75% Temperature range for transport and - 20° + 70° storing Humidity range for transport and < 95% without condense storing Minimum atmospheric pressure for 630 hPa transport and storing I-Max Touch...
IEC 60601-1-3:1994 IEC 60825-1:1993 + A1:1997 + A2:2001 Classification I-Max Touch is an electro-medical X-ray device belonging to Class 1 and Type B as per classification IEC 60601-1, foreseen for a continuous working at intermittent load. According to CE 93/42 directive for medical devices, the equipment belongs to class II B.
SERVICE MANUAL Technical data 4.2. Loading curve of the tube and cooling curve of the anode Tube "CEI - OPX / 105" (0.5 IEC 336) Load Anode cooling curve I-Max Touch...
Page 46
SERVICE MANUAL Technical data Tube-head heating and cooling curve I-Max Touch...
IEC 601-1. Particularly, in accordance with IEC 601-2-7, the maximum deviation (including the correction and instrument’s accuracy) is less than or equal to 10 for kV, while for tube current is less than or equal to 15%. I-Max Touch...
NOTE: This program allows you to carry out the measuring of the exposure parameters with the tube-head arm in a fixed position (not rotating) without variations due to spine compensation. 3. Place the measuring instrument. I-Max Touch 4-10...
Page 49
6. To end the control program, press any key other than the increase and decrease keys ; the display will indicate: A N O R A M I and the unit will return to standard mode. 4-11 I-Max Touch...
SERVICE MANUAL Technical data 4.5. Dimensions 1020 (40,16") Ø 1040 (40,95") Ø 1140 (44,88") with free standing base Figure 4-1 - Base version I-Max Touch 4-12...
Page 51
SERVICE MANUAL Technical data 1020 (40,16") Ø 1040 (40,95") Ø 1140 (44,88") with free standing base 1910 (75,20") 1970 (77,56") with free standing base Figure 4-2 - Version with cephalometric unit 4-13 I-Max Touch...
Page 52
SERVICE MANUAL Technical data THIS PAGE IS INTENTIONALLY LEFT BLANK I-Max Touch 4-14...
minimum width of 80 cm (31.50"). WARNING: In its standard versions, the I-Max Touch must be fixed to the wall with the two brackets supplied. Each dowel must support a max. extraction force of 120 kg (264 lbs). The fixing dowels to be employed, for any kind of wall, are the following: full or concrete bricks: screw anchors (provided with the ...
NOTE: The I-Max Touch, IS SET TO connect, at the entrance of the X-ray room, the following control and warning devices: REMOTE X-RAYS BUTTON: "Dead man switch" remote control, ...
Page 55
Power voltages for the relevant devices have to be provide from outside, making sure not to exceed the indicated ratings. To connect all control and warning devices it is advisable to set 6 wires with 0.5 mm² section minimum. I-Max Touch...
SERVICE MANUAL Pre-installation 5.2. Packaging The I-Max Touch in base version is delivered in two carton-board boxes; the device equipped with Cephalometric unit will be delivered in three boxes. Weight Contents Dimensions Gross - Unit head with rotation arm including the tubehead and sensor...
SERVICE MANUAL Pre-installation 5.3. Space requirements 5.3.1. Version without CEPH (*) A = minimum 600 mm (23.6"), recommended 800 mm (31.5") for service purpose Figure 5-1 I-Max Touch...
Pre-installation 5.3.2. Version with CEPH A = 600 mm (23.6") minimum, 800 mm (31.5") recommended for service purpose (**) B = 1460 mm (57.52") minimum front space, 1610 mm (63.39") recommended front space for service purpose Figure 5-2 I-Max Touch...
Installation INSTALLATION NOTE: The I-Max Touch is delivered pre-mounted in groups; it is contained in 2 boxes. The mechanical mounting consists exclusively in assembling the above mentioned groups. Most of the adjustments are carried out in factory. Two technicians will be necessary to perform some procedures; the phases requiring the intervention of two men are identified in the related chapters.
NOTE: The I-Max Touch has been designed for a wall fixing by two brackets, each of which requires to be fixed by two dowels. In order to find the right position of the brackets, it is necessary to use the quotes indicated in the Figure 6-1.
(this operation requires the presence of two persons). 2. With the column in vertical position, remove only the two fixing brackets "A". NOTE: Leave the retaining bracket "B" in place, because it will be necessary for the operations of support arm mounting. I-Max Touch...
Page 62
Rotate the column. NOTE: The I-Max Touch is equipped with a switch (S2) which allows to move the column vertically for service use. It is located in the back lower side of the column, protected by a metallic cover.
(Figure 6-4 - This operation requires the presence of two persons). Handling zones Handling zones Figure 6-4 3. Move the rotating arm near the column from lateral side in order to avoid contacts between the Digital Sensor holder and the chin support. I-Max Touch...
Page 65
6. Insert the reference pins "B" (Figure 6-6) between the rotating arm and column using a hammer. Insert the remaining two screws "C" (Figure 6-6) and tighten all the fixing screws. Retaining bracket Figure 6-6 I-Max Touch...
Page 66
9. Open the sensor holder and mount the rotating arm lower cover (Figure 6-8). NOTE: In case of Ceph up-gradable version or single sensor version, in order to mount the rotating arm lower cover, it is necessary to remove the sensor holder internal covers. I-Max Touch...
Page 67
SERVICE MANUAL Installation Figure 6-8 10. Mount the temple support group inserting and rotating it in the relevant shaft (Figure 6-9) until the group is locked. Temple support shaft Figure 6-9 (Rev. Preliminary) I-Max Touch...
Page 68
NOTE: For 110-120V version, to access to the CPU board connectors, it is necessary to remove the metallic cover. After the intervention, position the cables in the relevant seating and remount the cover. I-Max Touch 6-10...
Mounting of CEPH-arm (Optional) The CEPH kit can be installed on the machine both during the first installation and later as updating of the device. In case the I-Max Touch is sold already equipped with the Ceph arm, centering between the X-ray beam and the Sensor will be very quick as the arm is already adjusted and pinned in the factory.
Page 70
Ceph arm to the column (Figure 6-11). 5. Insert the cables in the hole between column and rotating arm and connect them to the relevant connectors on the CPU and Digital boards. 6. Remove the red handle. Reference pin Reference pin Figure 6-11 I-Max Touch 6-12...
"F" (Figure 6-3) from the upper bracket "C". Rotate the column in clockwise direction in order to reach the column back side. Figure 6-12 (Rev. Preliminary) 6-13 I-Max Touch...
The digital sensor can still function correctly also when the colour changes, displaying a fall that might also not have damaged the sensor. NOTE: The fallen sensor colour change interrupts the warranty on the sensor. I-Max Touch 6-14...
verification of the correct function and status of cleanness of the laser centering devices verification of the correct centering of the secondary collimator and of the Ceph sensor. I-Max Touch...
3. Place the centering tool (P/N 6107900200 - Figure 7-1) on the chin rest and the sensor calibration tool (P/N 5607900800 - Figure 7-2) in front of the sensor fixing it with tape. Figure 7-1 Centering tool P/N 6107900200 Figure 7-2 Sensor Calibration tool P/N 5607900800 I-Max Touch...
Page 76
7. Using the dedicated cursors, position the image to see the balls of the left side. 8. Select the instrument which measures distance and position the cursor on the center of the ball on the left of the image. I-Max Touch...
Page 77
12. Make a measurement also of the distance between the most right and most left ball: it must be inside the range 188 mm 2 mm. 13. If distance is outside the tolerance range, adjust the Y axis (see paragraph 8.3.4.2) accordingly and repeat the exposure. I-Max Touch...
This adjustment needs a personal computer directly connected to the I-Max Touch where the "QuickVision" program is installed and the use of the CD (P/N 5807304100) supplied with the centering tools kit. To verify the centering of the Cephalometric function it is necessary to: 1.
Page 79
The digital ceph centering complete procedure is composed by the steps reported on the following table, perform them in the correct order: Step Action Reference paragraph Rotation arm alignment 7.1.2.1 Ear centering circles setting 7.1.2.2 Secondary collimator centering 7.1.2.3 Ceph sensor centering 7.1.2.4 Soft Tissue Filter adjustment 7.1.2.5 I-Max Touch...
Page 80
WARNING: X-rays will be emitted during the performance of the following operations. Authorised Technicians are therefore recommended to use the greatest caution and to comply with the safety regulations and laws of their country. I-Max Touch...
12. Press the X-ray button and keep it depressed until the end of the exposure. NOTE: If the following message will be displayed: T A L E N S N O T R E A D Y check the connection with the PC. Press key (6/25) to reset the message. I-Max Touch...
Page 82
(suggested value 2) and with keys turn the tubehead arm, increasing the displayed offset value if the left distance is lower than the right one or viceversa; repeat the test from point 11. Figure 7-3 I-Max Touch 7-10...
Page 83
C H A N G E S > < 16. Press key (6/25) to store the changes. While key (26) to cancel them. The sensor will move to the parking position. Write down the new value in the relevant box in Appendix A. 7-11 I-Max Touch...
18. Open the "Ear centering" program on the CD (P/N 5807304100) and wait until the message "Board is connected" is displayed on the bottom bar of the program. Check that in the menu "Image processing" all the items are not selected except for DNSU and PRNU. I-Max Touch 7-12...
Page 85
– i.e. as shown in Figure 7-9); if it is necessary to adjust the Ear Centering Circles position, perform the setting actions described in paragraphs 7.1.2.2.1 and 7.1.2.2.2. 23. Repeat steps 21 and 22, take a new images until the Ear Centering Circles are concentric. 7-13 I-Max Touch...
Page 86
Ceph group. Repeat the exposure until the vertical alignment is reached. Tighten the loosened screws "A". Figure 7-4 25. When the right setting is reached press key (6/25) to exit the menu item. I-Max Touch 7-14...
In the opposite case, unscrew adjustment screw "D" and the group will shift automatically. Once the correct centering is obtained (a test exposure is required), fix screws "C" and block adjustment screw "D" using security nut "E". Figure 7-5 (Top view) 7-15 I-Max Touch...
"G" (Figure 7-6). Once the aligned position has been reached tighten bolt "H" (Figure 7-6) and tighten screws "F". Upon completion of the adjustment a test exposure is required, to verify it. Figure 7-6 (Side view) I-Max Touch 7-16...
If the following message will be displayed: T A L E N S N O T R E A D Y check the connection with the PC. Press key (6/25) to reset the message and press key (24) again. 7-17 I-Max Touch...
Page 90
(Figure 7-7); to take the measure easily, you can use the QuickVision pseudo color function. Pseudo color 33. The right and left distances must not differ more than 3mm. Figure 7-7 I-Max Touch 7-18...
Page 91
D A T E C H A N G E S > < 36. Press key (6/25) to store the changes. While key (26) to cancel them. Write down the new value in the relevant box in Appendix A. 7-19 I-Max Touch...
This adjustment needs a personal computer directly connected to the I-Max Touch where the "QuickVision" program is installed and the use of the CD (P/N 5807304100) supplied with the centering tools kit. To perform the ceph sensor alignment start from point 36 of paragraph 7.1.2.3 and:...
Page 93
(suggested values: 60kV, 6mA and 0.2s). 42. Take X-ray pressing the X-ray button and check if in the obtained narrow image the projection of the slit of the centering tool is vertical and in the middle of the impressed area. 7-21 I-Max Touch...
Page 94
D A T E C H A N G E S > < 44. Press key (6/25) to store the changes. While key (26) to cancel them. Write down the new value in the relevant box in Appendix A. I-Max Touch 7-22...
NOTE: This adjustment needs a personal computer directly connected to the I-Max Touch where the "QuickVision" program is installed. This adjustment is accessed by activating password 124 as described in paragraph 7.1.2. 1. When in password 124 scroll the menu items pressing keys "Arrow right"...
Page 96
The system will calculate automatically the new Soft Tissue Filter offset. If distance "A" is 50 2 mm no correction is needed; press key "T" (26) to exit. In both cases the following message will be displayed: Z E R O ± I-Max Touch 7-24...
Page 97
Write down the new value in the relevant box in Appendix A. 10. Press key (26) to exit password 124, the following message will be displayed C O N F E X I > < > Press key (6/25) to exit the program and store the changes. 7-25 I-Max Touch...
After 3 seconds, the setup screen with the following message will be displayed E R V I M E N U C O N F G U R A T I I-Max Touch 7-26...
Page 99
To move to mA and exposure time, press "P" key once or twice again. In this situation, the keys increase and decrease will change the parameter. Pressing the X-ray button will start the emission. 7-27 I-Max Touch...
Page 100
(not rotating). WARNING: X-ray will be emitted during the performance of the following operations. Authorised technicians are therefore recommended to use the greatest caution and to comply with the safety regulations and laws of their country. I-Max Touch 7-28...
The high voltage value must be within 8% of the set value, while the time value must be within 10% of the set value. If the voltage is not within the expected range, replace the tubehead; otherwise replace Generator CPU board. 7-29 I-Max Touch...
Generator board (A10). The exposure time calculated at about 75% of the maximum kV value, must correspond to the set one +10%. In case the time is outside the prescribed values, replace the CPU. I-Max Touch 7-30...
4. After pressing the key, the display will show the following message: T O R E > < Press the key "Patient Entrance" (6) to confirm or the key "Test" (5) to cancel the setting. 7-31 I-Max Touch...
Table of pre-set anatomic parameters PANORAMIC TMJ open/close mouth Child Child Adult Adult Small Small Medium Medium Large Large SINUS Adult Child Small Medium Large CEPHALOMETRY (L.L.) CEPHALOMETRY (A/P - P/A) Adult Child Adult Child Small Small Medium Medium Large Large I-Max Touch 7-32...
Check the correct insertion of X57; if it is correct, replace the tubehead, otherwise insert it correctly and make another exposure. Replace the tubehead if it will glow again. As per H5 above. I-Max Touch...
SERVICE MANUAL Troubleshooting 8.2. Displayed messages The I-Max Touch is fully driven by a microprocessor which controls the programming of the emission parameters and indicates via displayed messages the different conditions of the unit and any defects and errors that occur.
Page 109
Sensor ceph micro always active 8.2.2.5 Secondary collimator on Digital CEPH Code Error description Reference chapter Unexpected activation of optical sensor secondary 8.2.3.1 collimator Timeout on ceph secondary collimator 8.2.3.1 Secondary collimator micro never active 8.2.3.1 Secondary collimator micro always active 8.2.3.1 I-Max Touch...
Page 110
Soft Tissue filter Code Error description Reference chapter Timeout of zero position optical sensor of STF 8.2.4.1 STF zero position sensor always active 8.2.4.1 Code Error description Reference chapter Wrong setting in PAN + CEPH system configuration 8.3.3.8 I-Max Touch...
Page 111
ON/OFF command for vertical motor not changed on 8.2.6.2 planned time Keyboard Code Error description Reference chapter More than one key pressed on Power on 8.2.7.1 Column up and/or Column down pressed on power on 8.2.7.2 Key RESET (patient centering) pressed during 8.2.7.3 movement I-Max Touch...
8.2.1.1. E108: Hardware key fault This error is shown when the firmware of the I-Max Touch does not sense the presence of the U.I.C. (Unique Identification Code). The unique code is read, with its check byte, from the control system at the start-up;...
CPU. 2. If the arm moves, verify the cable up to the CPU board (A5). If it still continues to have no variation of the signals when there is an activation of the switches, change the CPU board. I-Max Touch...
CPU board (A5), connector X25/X25. 2. If the motor is running, check the optical sensor B2 and its cables to X60 of A7 up to connector X25/X26 of main CPU board (A5). Replace the sensor or the faulty cables. I-Max Touch...
In case of no motion, it is possible that the problem is related to the motor: in order to check this, connect the driver of the Y-axis motor to the rotation motor, putting cable X28 into connector X26 on CPU board (A5). I-Max Touch...
X1-2 of the board up to X23, X24 and X34 of the main CPU; check the continuity of the interested wires and replace the faulty one (if any). If all cables are OK, the fault is on the main CPU, so replace it. I-Max Touch 8-10...
X1-2 of the board up to X23, X24 and X34 of the main CPU; check continuity of the interested wires and replace the faulty one (if any). If all cables are OK, the fault is on the main CPU, so replace it. 8-11 I-Max Touch...
X64 to X29 and X30 of the main CPU Check continuity of the above cables, from M5 to X64 and from X64 up to X29 and X30 of the main CPU A5. Repeat the power on sequence, verifying the motion. I-Max Touch 8-12...
If this does not reset the error message: Check the proper function of optical sensor B6. Check continuity of cables from optical sensor to board A8, connector X48, and from that board to main CPU board (A5), connector X33. 8-13 I-Max Touch...
230 VAC and 400 VDC. Before accessing the Generator CPU and power board (A9 and A10), it is mandatory to switch off the mains and wait up to 2 minutes in order to allow the discharge of the capacitor. I-Max Touch 8-14...
In both cases, the error can be generated by a bad connection of the CANBus or a fault on the local Generator CPU board, so the steps above described for E380 are applied. 8-15 I-Max Touch...
13.3k ± 2% while is 14.3k ± 2% between pins 4 and 3 of the same connector; if correct values are measured, replace before the tubehead and, eventually, the Generator board. If incorrect values are measured, replace the tubehead. 8-17 I-Max Touch...
If the error is present in stand-by mode, replace the Generator board (A10). If the error appears during an exposure, reset the error and repeat the exposure; if the error still appears, replace the tubehead. I-Max Touch 8-18...
If the LED H4 continues to be ON, replace the Generator board. If the LED, after the power ON is not lighted, repeat an emission and, if the message is appearing again, replace the Generator board. 8-19 I-Max Touch...
CPU A5 (cable X20-X20). Check the correct insertions of X56 and X57 between the Generator board to the tubehead. If all connections are OK and the fault is repetitive, replace the Generator CPU board (A9). I-Max Touch 8-20...
This message has the same meaning as the corresponding error E362, but it is generated by the Generator CPU board (A9), that is signalling a possible broken connection with the main CPU (A5). Repeat the test as per error E774 above. 8-21 I-Max Touch...
Check that the column movement power supply fuse F4 is not blown. Check the safety column movement microswitches work correctly, otherwise replace the microswitches assembly (microswitch + cable). Check CANBus cable X11-X11 and replace it if faulty, otherwise replace the Column CPU board. I-Max Touch 8-22...
This message is signalling that only one of the two buttons is pressed. Release the pressed button and power off the system and power on it again. If the message is displayed again, replace first the keyboard membrane and after the control board. 8-23 I-Max Touch...
Check if one button has been involuntary been pressed and restart the operation with a new power on procedure. In case there is a stuck key, one of the previous message E850 and E851 must be displayed, so acts accordingly. I-Max Touch 8-24...
NOTE: SYSTEM and BURN-IN operations are used only during factory set-up; there are special passwords not accessible to Service Engineers. For this reason, they are not explained on the manual. 8-25 I-Max Touch...
This operation is signalled by a specific message on the display. Left/Right scroll: are used to scroll into different parameters or menus. Parameter increase and decrease: are used to change the value of the selected parameter. I-Max Touch 8-26...
In this menu, it is possible to select the language of displayed message of the user’s panel (language of service menus is always English) in one of the following: ENGLISH ITALIANO FRANCAIS DEUTSCH ESPANOL PORTUGUES NETHERLAND TURKCE ENGLISH … 8-27 I-Max Touch...
Values are displayed and changed using a couple of digits, with a variation of one unit for each pressure of the parameter increase/decrease keys; the auto repeat function for fast changes is enabled. 8-29 I-Max Touch...
2. By means of the keys Arrow right/Arrow left VALUE1 and using the increase/decrease keys , set the second couple of digits equal to the relevant value. 3. Repeat point 2 until VALUE7. I-Max Touch 8-30...
In this menu, it is possible to set the system for digital or film base; the selection is performed using the parameter increase/decrease keys and selection is confirmed in the usual way. WARNING: Changing the set mode will affect the complete system functionality. 8-31 I-Max Touch...
"PAN ONLY IS OFF" for PAN ready for Digital CEPH or PAN + CEPH systems. WARNING: A wrong setting provides the message "Close cassette to Panoramic" at the start-up for a PAN only system, or error 420 for a PAN+CEPH system when the Cephalometric exam is selected. I-Max Touch 8-32...
It is possible to change the set value using the parameter increase/decrease keys; select "SENSOR IS MOBILE" for 1 sensor configuration or "SENSOR IS FIXED" for 2 fixed sensors configuration. WARNING: A wrong setting provides operative messages that are not consistent with the equipment configuration. 8-33 I-Max Touch...
Those values are used only by the DSPU, so only on the Digital units; do not update those in case of an analog (film) unit, and/or change those only to adhere at the specified values of the network manager in case of IP address conflict. I-Max Touch 8-34...
Changes to one or more offset values affect the system’s functionality, so take care to not alter those values if not needed. Normally only the replacement of faulty parts (motors, belts, zero optical sensors, etc.) will require to act on those data. 8-35 I-Max Touch...
"a" is the status of the zero optical barrier; 1 means engaged (i.e. barrier interrupted), 0 otherwise. "xxxx" is the current value of the offset, that can be changed with the left/right keys. I-Max Touch 8-36...
ENABLE SENSOR Y OFFSET ROTATION OFFSET LINING UP TEST CEPH S. COL. ZERO CEPH SENSOR ZERO STF ZERO OFFSET ENABLE SENSOR … The selection is confirmed pressing the patient centering (>0<) keys. I-Max Touch 8-38...
"a" is the status of the zero optical barrier; 1 means engaged (i.e. barrier interrupted), 0 otherwise. "xxxx" is the current value of the offset, that can be changed with the left/right keys. I-Max Touch 8-40...
0 otherwise. "xxxx" is the current value of the offset, that can be changed with the left/right keys. 8.4.1.4. Lining up test This menu is used to align the ear centering device; refer to paragraph 7.1.2.2. 8-41 I-Max Touch...
"xxxx" is the zero offset actual value, that can be changed with the left/right keys. "abcd" represents the status of various optical barrier sensors; "a" is the one of the secondary collimator, that is 0 when not engaged and 1 otherwise. I-Max Touch 8-42...
"abcd" represents the status of various optical barrier sensors; "b" is the one of the zero of digital sensor, that is 0 when not engaged and 1 otherwise. 8.4.1.7. STF zero This program is used to adjust the offset of the Soft Tissue motorized filter. 8-43 I-Max Touch...
I2C BUS TEST TEST COLUMN … NOTE: Of these submenus, actually only the Test Column, Test Input Ports, PAN RX emission, CEPH RX emission, Pre-heating time and Preheating level are active. The selection is confirmed pressing the Patient centering (>0<) keys. I-Max Touch 8-44...
The digits are numbered from 7 to 0. The 7 digit corresponds to the character furthest to the left on the display, while digit 0 corresponds to the character furthest to the right. 8-45 I-Max Touch...
This function allows the set of the Tubehead pre-heating time. After the selection, the display will show the following message: R E - H E A T I V A L U E It is possible to change the set value using the parameter increase/decrease keys. I-Max Touch 8-46...
L E V E L Y Y Y where: "XX" is the Tubehead filament current value, that can be changed using the left/right keys. "YYY" is the Tubehead pre-heating value, that can be changed with the parameter increase/decrease keys. 8-47 I-Max Touch...
Page 154
(*) Functionality only on Analog version (**) Functionality only on Digital version NOTE: The logic status of these signals depends on the physical position of the relevant optical sensor / microswitch with respect to the "0" position. I-Max Touch 8-48...
Where the symbol ">" is showing the parameter to be changes; in this case the kV. To move to mA and exposure time, press key once or twice again. In this situation, the increase and decrease keys will change the parameter. Pressing the X-ray button will start the emission. 8-49 I-Max Touch...
The screen is cleared and the user is prompted to touch 9 points on the screen successively. Please use a stylus to touch each point as precisely as possible. After the last point has been touched, the screen reverts back to the Setup display. I-Max Touch 8-50...
(In CEPH examination) Panoramic chin-rest Repeat the examination removing white area on the mounted. the PAN chin-rest. lower part of the film. 8-51 I-Max Touch...
If the value is correct but the effect on the image is poor, retake the exposure setting a lower value for the STF. A symmetric format was Select an asymmetrical format selected. (which will enable the STF filter). I-Max Touch 8-52...
Troubleshooting 8.5.3. Defects due to the unit The experience from field problems observed on the I-Max Touch has allowed to list under this chapter some possible typical problems that can affect the radiological quality. 1. The Panoramic or Cephalometric image results in a light one also if...
Normally, the panoramic equipment has a narrower focussed layer in the front area, and therefore a proper positioning of the anterior teeth is of the utmost importance. Figure 8-2 8-55 I-Max Touch...
Page 162
The occlusal plane is shown slightly tilted upward, the palatal plane does not overlap the apex of the upper arch and therefore allows a good view of the same. I-Max Touch 8-56...
If, after the correct patient’s positioning, the problem still remain, check the alignment of rotating arm. A more accurate check can be performed using the service tool and verifying the distance between the two lateral metal parts. 8-57 I-Max Touch...
Page 164
If, after the correct patient’s positioning, the problem still remains, check the alignment of rotating arm. A more accurate check can be performed using the service tool and verifying the distance between the two lateral metal parts. I-Max Touch 8-58...
Check the positioning of the median-sagittal plane by using the relevant centring light beam. A more accurate check is performed using the special tool and making an image; the dimensions measured on the obtained image must match those reported in the paragraph 7.1.1. 8-59 I-Max Touch...
Page 166
If not, a possible cause can be due to a non perfect horizontality of the chin arm that must be adjusted using the relevant screws. I-Max Touch 8-60...
Page 167
A chin support plane too high during the patient positioning or when extending the spine may generate this mistake. In this condition, the rear side of the patient’s head may also interfere with the rotating arm of the panoramic equipment. 8-61 I-Max Touch...
Page 168
Figure 8-9 Possible cause: Patient’s head tilted downward, as on the aside schema. Remedy: Check the positioning of the patient by aligning it with the help of the corresponding light beam. I-Max Touch 8-62...
Page 169
In this case, a light tilting downward of the Frankfurt plane causes the palate to be projected over and far enough from the roots of the teeth of the maxillary arch, without distortion of the incisor teeth, as on Figure 8-11. Figure 8-11 8-63 I-Max Touch...
(the other side of the mandible). That therefore results as a clearer area overlapped to the real image. Very often the resulting darker area in the bottom corner is noticed and is considered as an artefact of the radiological image. I-Max Touch 8-64...
Page 171
This situation may occur especially if the patient has large prosthesis or metal cures and if it is associated to a positioning error that projects the shadow of the metal part on a wide area of the image. 8-65 I-Max Touch...
Page 172
This area is identified with reference "1" in Figure 8-12. Remedy: Ask the patient to position the tongue against the palate during the exposure. I-Max Touch 8-66...
8. Lay-out and Circuit diagram Rotation Group PCB A7 9. Lay-out Generator CPU PCB A9 10. Circuit diagram Generator CPU PCB A9 11. Lay-out Generator PCB A10 12. Circuit diagram Generator PCB A10 13. Lay-out and Circuit diagram CEPH arm connection PCB A12 I-Max Touch...
Page 174
SERVICE MANUAL Schematics and drawings THIS PAGE IS INTENTIONALLY LEFT BLANK I-Max Touch...
Page 175
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 1 of 9 I-Max Touch...
Page 177
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 2 of 9 I-Max Touch...
Page 179
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 3 of 9 I-Max Touch...
Page 181
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 4 of 9 I-Max Touch...
Page 183
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 5 of 9 9-11 I-Max Touch...
Page 185
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 6 of 9 9-13 I-Max Touch...
Page 187
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 7 of 9 9-15 I-Max Touch...
Page 189
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 8 of 9 9-17 I-Max Touch...
Page 191
SERVICE MANUAL Schematics and drawings I-MAX TOUCH General diagram Code 39069001 - Rev. 2 Page 9 of 9 9-19 I-Max Touch...
Page 193
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Lay-out Column CPU PCB A1 Code 58071010 - Rev. P Page 1 of 1 9-21 I-Max Touch...
Page 195
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit Diagram Column CPU PCB A1 Code 39079005 - Rev. P Page 1 of 3 9-23 I-Max Touch...
Page 197
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit Diagram Column CPU PCB A1 Code 39079005 - Rev. P Page 2 of 3 9-25 I-Max Touch...
Page 199
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit Diagram Column CPU PCB A1 Code 39079005 - Rev. P Page 3 of 3 9-27 I-Max Touch...
Page 201
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Lay-Out CPU PCB A5 Code 58073021 - Rev. 2 Page 1 of 1 (Rev. 0) 9-29 I-Max Touch...
Page 203
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 1 of 9 9-31 I-Max Touch...
Page 205
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 2 of 9 9-33 I-Max Touch...
Page 207
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 3 of 9 9-35 I-Max Touch...
Page 209
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 4 of 9 9-37 I-Max Touch...
Page 211
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 5 of 9 9-39 I-Max Touch...
Page 213
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 6 of 9 9-41 I-Max Touch...
Page 215
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 7 of 9 9-43 I-Max Touch...
Page 217
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 8 of 9 9-45 I-Max Touch...
Page 219
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram CPU PCB A5 Code 39079007 - Rev. 1 Page 9 of 9 9-47 I-Max Touch...
Page 221
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Lay-out Microprocessor PCB A6 Code 58073023 - Rev. 8 Page 1 of 1 9-49 I-Max Touch...
Page 223
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Microprocessor PCB A6 Code 39079009 - Rev. 1 Page 1 of 4 9-51 I-Max Touch...
Page 225
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Microprocessor PCB A6 Code 39079009 - Rev. 1 Page 2 of 4 9-53 I-Max Touch...
Page 227
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Microprocessor PCB A6 Code 39079009 - Rev. 1 Page 3 of 4 9-55 I-Max Touch...
Page 229
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Microprocessor PCB A6 Code 39079009 - Rev. 1 Page 4 of 4 9-57 I-Max Touch...
Page 231
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Lay-out and circuit diagram Rotation group PCB A7 Code 58073025 - Rev. P Page 1 of 1 9-59 I-Max Touch...
Page 233
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Lay-out Generator CPU PCB A9 Code 58074023 - Rev. 2 Page 1 of 1 9-61 I-Max Touch...
Page 235
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Generator CPU PCB A9 Code 39079011 - Rev. 1 Page 1 of 2 9-63 I-Max Touch...
Page 237
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Generator CPU PCB A9 Code 39079011 - Rev. 1 Page 2 of 2 9-65 I-Max Touch...
Page 239
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Lay-out Generator PCB A10 Code 58074021 - Rev. 3 Page 1 of 1 9-67 I-Max Touch...
Page 241
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Generator PCB A10 Code 39079013 - Rev. 2 Page 1 of 3 9-69...
Page 243
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Generator PCB A10 Code 39079013 - Rev. 2 Page 2 of 3 9-71...
Page 245
SERVICE MANUAL Schematics and drawings I-MAX TOUCH Circuit diagram Generator PCB A10 Code 39079013 - Rev. 2 Page 3 of 3 9-73...
Page 247
SERVICE MANUAL Schematics and drawings + FOTOCEL. (+PHOTOCELL) - FOTOCEL. (-PHOTOCELL) + FOTOCEL. (+PHOTOCELL) - FOTOCEL. (-PHOTOCELL) I-MAX TOUCH Lay-out and circuit diagram PCB A12 Code 58098150 - Rev. 0 Page 1 of 1 9-75...
SERVICE MANUAL Spare Parts SPARE PARTS 1 - COLUMN Cables Electrical and mechanical parts 2 - UPPER MOVEMENT ASSY Cables Electrical and mechanical parts 3 - ROTATION ARM 4 - CEPH DEVICE (valid for S/N ≤ 10020532) 5 - CEPH DEVICE (valid for S/N ≥ 10020533) 6 - COVERS 7 - ACCESSORIES AND SERVICE TOOLS 10-1...
Page 250
SERVICE MANUAL Spare Parts THIS PAGE IS INTENTIONALLY LEFT BLANK 10-2...
Page 251
SERVICE MANUAL Spare Parts 1 - COLUMN Cables Rif. Order code Description Note 6207091200 General supply cable 6207091300 Ground cable #1 Generator board A10 power 6207091900 supply cable X2-X70 6207092200 DSPU board – Touch screen connection cable J1-J2 / J3-J4 6207090600 CanBus cable X11 6207092000...
Page 252
220-240V version 110-120V 6607090300 Fuses kit version 2100440400 Column base grub screw cap diam. 12.7 mm (1x) 2100440200 Column base grub screw cap diam. 4.8 mm (1x) 5106094300 Column lateral I-Max Touch logo 5106094000 Column front Owandy Radiology logo 10-4...
Page 254
SERVICE MANUAL Spare Parts 2 - UPPER MOVEMENT ASSY Cables Ref. Order code Description Note 6207040100 Rotation group motor and signals cable X25 / X26 – X62 6207040200 Y-axis signal cable X27 – B1 6207040600 CanBus cable X20 6207040700 Ground cable # 2 6207040900 Collimator optical sensor cable X70 –...
Page 255
SERVICE MANUAL Spare Parts Electrical and mechanical parts 6 (B3) 6 (B1) 6 (B2) Ref. Order code Description Note 6607035000 M2 Y-Axis stepper motor assembly 6607025000 M3 rotation stepper motor assembly 4990802300 Carriage belt model HTD 843-3M-15 6607303000 CPU board assembly 5807302100 CPU board A5 6607302300...
Page 256
SERVICE MANUAL Spare Parts 3 - ROTATION ARM Ref. Order code Description Note 6607000000 HF Tubehead 5807402100 Generator board A10 5807402300 Generator CPU board A9 4192212300 Line filter Z2 5807302900 Optical sensor board B6 / B7 / B8 / B9 6607120200 Soft Tissue Filter motor M6 6607120100...
Page 257
SERVICE MANUAL Spare Parts 5 (B 6 / B 7 ) 5 (B 8 / B 9 ) 10-9...
Page 258
SERVICE MANUAL Spare Parts 4 – CEPH DEVICE (valid for S/N ≤ 10020532) Ref. Order code Description Note 5807302900 Optical sensor board B10 / B11 5207081800 Motor joint 6209815600 Secondary collimator motor M8 5809815000 Interconnection board A12 6209815500 Ceph sensor motor M9 6607403100 Sensor holder connector plate for S/N...
Page 259
SERVICE MANUAL Spare Parts 1 (B11) 1 (B10) 10-11...
Page 260
SERVICE MANUAL Spare Parts 5 – CEPH DEVICE (valid for S/N ≥ 10020533) Ref. Order code Description Note 5807302900 Optical sensor board B10 / B11 6607081000 Secondary collimator motor M8 6607081200 Ceph Sensor motor M9 4990804800 Secondary collimator / Ceph L = 861 mm Sensor dragging belt 5207084000...
Page 261
SERVICE MANUAL Spare Parts 1 (B11) 1 (B10) 10-13...
Page 262
Rotation arm lower cover 5407011905 Temple clamp cover 6606070005 PAN / CEPH Sensor holder internal cover 6606160600 Console covers (upper + lower) 5106094400 Console Owandy Radiology logo 6606160700 Console arm (DX + SX) 6606140205 Base cover 6606140105 Base cover (free standing base version)
Page 264
SERVICE MANUAL Spare Parts 7 - ACCESSORIES AND SERVICE TOOLS Ref. Order code Description Note 6607090100 PAN centring bite (50 pcs) 6107110200 Disposable bite protective sleeves (100x) 6607080200 Ceph ear pivot (50 pcs) 6606090108 Panoramic standard chin support 5407098108 Edentulous patients appendix 6606090208 SINUS/TMJ chin support 6609503600...
Page 266
APPENDIX 11.1. Appendix A: Setup parameters table The following table lists those adjustment parameters stored in the unit during factory testing and that must be re-entered into the non-volatile memory in case of replacement of the main CPU board (A5). This is due to the fact that the new CPU board, provided as a spare part, has been factory tested from the functional point of view, but contains only default parameters which are not related to the unit where it will be installed.
Page 267
SERVICE MANUAL Spare Parts I-Max Touch Unit code: Unit S/N: U.I.C.: Factory Parameter setting setting setting setting setting Date Y axis motor offset Rotation axis motor offset PAN only (ON/OFF) Sensor Handling Digital Ceph status (OFF=disabled; ON=enabled) Primary collimator motor offset...
Need help?
Do you have a question about the I-Max Touch and is the answer not in the manual?
Questions and answers