Fig. 79: Calibration Complete; Fig. 80: Fiducial Settings - LPKF ProtoMat S62 Operating Manual

Hide thumbs Also See for ProtoMat S62:
Table of Contents

Advertisement

Fig. 79: Calibration com-
plete

Fig. 80: Fiducial settings

Operating Manual 7.1
The message „Calibration complete!" should appear.
11. Click on OK.
12. In the Setting window, click on OK to save the settings.
13. In the Camera Image window, click on Setting, to open the Set-
ting window again.
14. Enter the actual value of the diameter in the Fiducial diameter:
input box (here: 1.5 mm)
15. In the Minimal diameter [mm] input put, enter the value 1.4
(1.3 might also be necessary).
16. Click on Accept to confirm the entries.
17. Click on OK to close the Setting window.
Care and maintenance
Rev.: 18.06.2008
139