Configure advanced options for process measurement
Overview
The transmitter can write the current value of four totalizers or inventories to a log, at
user-specified intervals. You can generate a log file from this data for viewing and analysis.
Procedure
1.
2.
3.
4.
Related information
6.7

Configure Process Variable Fault Action

Display
ProLink III
Web browser
Overview
Process Variable Fault Action specifies the values that will be reported via the display and
digital communications if the device encounters a fault condition. The values are also sent
to the outputs for processing against their configured fault actions.
Procedure
Set Process Variable Fault Action as desired.
Default: None
Restriction
If you set Process Variable Fault Action to NAN, you cannot set mA Output Fault Action or Frequency
Output Fault Action to None. If you try to do this, the transmitter will not accept the configuration.
Important
If you want the mA output to continue reporting process data during fault conditions, you must
set both Process Variable Fault Action and mA Output Fault Action to None. If mA Output Fault
Action is set to None and Process Variable Fault Action is set to any other option, the mA output
will produce the signal associated with the selection.
94
Specify the date on which totalizer logging will begin.
You must specify a future date. If you try to specify the current date, the transmitter
will reject the setting.
Specify the time at which totalizer logging will begin.
Specify the number of hours between records.
Select up to four totalizers or inventories to be logged.
Totalizer history and log
Generate history log files
Menu > Configuration > Alert Setup > Output Fault Actions
Device Tools > Configuration > Fault Processing
Configuration > Alert Setup > Output Fault Actions