Mindray Accutorr 7 Operator's Manual

Mindray Accutorr 7 Operator's Manual

Vital signs monitor
Hide thumbs Also See for Accutorr 7:
Table of Contents

Advertisement

Accutorr 7
Vital Signs Monitor
Operator's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Accutorr 7 and is the answer not in the manual?

Questions and answers

Summary of Contents for Mindray Accutorr 7

  • Page 1 Accutorr 7 Vital Signs Monitor Operator’s Manual...
  • Page 3 © Copyright 2013-2017 Shenzhen Mindray Bio-Medical Electronics Co., Ltd. All rights reserved. Release date: September 2017 Revision: 7.0 Accutorr 7 Operator’s Manual...
  • Page 4 , and are the registered trademarks or trademarks owned by Mindray in China and other countries. All other trademarks that appear in this manual are used only for editorial purposes without the intention of improperly using them. They are the property of their respective owners.
  • Page 5 This warranty does not cover consumable items such as, but not limited to, batteries, external cables, and sensors. Mindray shall not be liable for any incidental, special, or consequential loss, damage, or expense directly or indirectly arising from the use of its products. Liability under this warranty and the buyer’s exclusive remedy under this warranty is limited to servicing or...
  • Page 6 Exemptions Mindray's obligation or liability under this warranty does not include any transportation or other charges or liability for direct, indirect or consequential damages or delay resulting from the improper use or application of the product or the use of parts or accessories not approved by Mindray or repairs by people other than Mindray authorized personnel.
  • Page 7 Company Contact Manufacturer: Shenzhen Mindray Bio-Medical Electronics Co., Ltd. Address: Mindray Building, Keji 12th Road South, High-tech industrial park, Nanshan, Shenzhen 518057, P.R.China Website: www.mindray.com E-mail Address: service@mindray.com.cn Tel: +86 755 81888998 Fax: +86 755 26582680 Distributor: Mindray DS USA, Inc.
  • Page 8 This manual is based on the maximum configuration and therefore some contents may not apply to your product. If you have any questions, please contact Mindray. This manual is an integral part of the product. It should always be kept close to the equipment so that it can be obtained conveniently when needed.
  • Page 9: Table Of Contents

    3.4.1 Using Keys ............................3 - 5 3.4.2 Using the Knob ..........................3 - 6 3.4.3 Using the Touchscreen .......................3 - 6 3.5 Using the Timer .............................3 - 6 3.5.1 Displaying the Timer ........................3 - 6 3.5.2 Controlling the Timer ........................3 - 6 Accutorr 7 Operator’s Manual...
  • Page 10 5.6 Loading a Configuration ..........................5 - 3 5.7 Restoring the Latest Configuration Automatically ................5 - 3 6 Alarms ............................6 - 1 6.1 Alarm Categories ............................6 - 1 6.2 Alarm Levels ..............................6 - 2 6.3 Alarm Indicators ............................6 - 2 Accutorr 7 Operator’s Manual...
  • Page 11 7.8 Nellcor Information ............................7 - 7 8 Monitoring PR ..........................8 - 1 8.1 Overview ................................8 - 1 8.2 PR Source .................................8 - 1 8.3 Pulse Tone ...............................8 - 1 9 Monitoring NIBP .........................9 - 1 9.1 Overview ................................9 - 1 Accutorr 7 Operator’s Manual...
  • Page 12 Settings ......................... 11 - 4 11.7.1 Changing CO Alarm Settings ..................... 11 - 4 11.7.2 Setting the “No CO2 Changes Detected” Alarm Delay ..........11 - 4 11.7.3 Setting the CO Waveform ....................11 - 5 Accutorr 7 Operator’s Manual...
  • Page 13 13.7 Cleaning the Recorder Printhead ...................... 13 - 4 14 Network ...........................14 - 1 14.1 Network Connection ..........................14 - 1 14.1.1 Selecting a Network Type ..................... 14 - 2 14.1.2 Setting the Wired Network ....................14 - 2 Accutorr 7 Operator’s Manual...
  • Page 14 16.4 Checking Monitor Information ......................16 - 5 16.5 Visual Inspection ............................. 16 - 5 16.6 Power-on Test ............................16 - 6 16.7 Calibrating CO ............................16 - 6 16.8 Battery Check ............................16 - 7 Accutorr 7 Operator’s Manual...
  • Page 15 C.1.5 CO ..............................C - 4 C.2 General Configuration ........................C - 5 C.2.1 Alarm ..............................C - 5 C.2.2 Review .............................. C - 5 C.2.3 Record .............................. C - 6 C.2.4 Others ............................... C - 6 Accutorr 7 Operator’s Manual...
  • Page 16 D.2.7 System Alarm Messages ......................D - 12 E Symbols and Abbreviations ......................E - 1 E.1 Symbols ................................E - 1 E.2 Abbreviations ..............................E - 3 F Anomalies ........................... F - 1 F.1 NIBP Smart Inflation Anomaly ........................F - 1 Accutorr 7 Operator’s Manual...
  • Page 17: Safety

    Sudden power failure may lead to the loss of patient data. • The equipment is magnetic resonance (MR) unsafe. To avoid an explosion hazard, do not use the equipment in the presence of oxygen–rich Accutorr 7 Operator’s Manual 1 - 1...
  • Page 18: Cautions

    • Do not open the equipment housings. All servicing or future upgrades must be carried out by Mindray trained and authorized personnel. • Do not come into contact with patients during defibrillation. Otherwise serious injury or death could result.
  • Page 19: Notes

    At the end of its service life, the equipment, and its accessories, must be disposed of in compliance with the guidelines regulating the disposal of such products. If you have any questions concerning disposal of the equipment, please contact Mindray. Accutorr 7 Operator’s Manual 1 - 3...
  • Page 20: Equipment Symbols

    Admit patient key Insertion Direction DEFIBRILLATION –PROOF Input/Output TYPE CF APPLIED PART Network connector Equipotentiality USB connector MANUFACTURER Serial number DATE OF MANUAFACTURE CATALOGUE NUMBER Protection against fluid ingress Temperature limit Humidity limitation 1 - 4 Accutorr 7 Operator’s Manual...
  • Page 21 Conforms to AAMI Std ES 60601-1, IEC Std 60601-1-6, IEC Std 60601-1-8, IEC Std 60601-2-49, IEC Std 80601-2-30, ISO Std 80601-2-56, ISO Std 80601-2-61. Certified to CSA Std C22.2 NO. 60601-1, NO. 60601-1-6, NO. 60601-1-8, NO. 60601-2-49, NO. 80601-2-30, NO. 80601-2-56, NO. 80601-2-61. Accutorr 7 Operator’s Manual 1 - 5...
  • Page 22 This page intentionally left blank. 1 - 6 Accutorr 7 Operator’s Manual...
  • Page 23: The Basics

    Anyone unauthorized or untrained must not perform any operation on it. Applied Parts The applied parts of the monitor include followings: ■ sensor and cable ■ NIBP tubing and cuff ■ Temp probes and cable ■ sampling line Accutorr 7 Operator’s Manual 2 - 1...
  • Page 24: Main Unit

    Temperature probe well Probe cover pack holder AC power indicator ◆ On: indicates that the monitor is connected to the AC power. ◆ Off: indicates that the monitor is not connected to the AC power. 2 - 2 Accutorr 7 Operator’s Manual...
  • Page 25 Press this key to return to the main screen. Knob ◆ Rotate the knob clockwise or counterclockwise to move the cursor. ◆ Press the knob to select one item, such as accessing a menu or confirming the selection. Accutorr 7 Operator’s Manual 2 - 3...
  • Page 26: Side View

    2.3.2 Side View Handle Recorder indicator Start/stop recording key Paper outlet Recorder door Connector for NIBP cuff Connector for SpO cable gas outlet sampling line connector Recorder door latch 2 - 4 Accutorr 7 Operator’s Manual...
  • Page 27: Rear View

    USB connector: connects to barcode scanner or USB disk. AC power input Equipotential grounding terminal NOTE • When using the equipment with other devices, their equipotential grounding terminals should be connected together to eliminate a possible difference in ground potential. Accutorr 7 Operator’s Manual 2 - 5...
  • Page 28: Bottom View

    2.3.4 Bottom View Battery compartment door Quick release mount latch point 2 - 6 Accutorr 7 Operator’s Manual...
  • Page 29: Main Screen

    Main Screen The main screen has three display modes. They are all parameter screen, trend screen and NIBP list screen. ■ All Parameter Screen Accutorr 7 Operator’s Manual 2 - 7...
  • Page 30 ■ Trend Screen 2 - 8 Accutorr 7 Operator’s Manual...
  • Page 31 When there are multiple messages, they will scroll. indicates alarms are paused. Accutorr 7 Operator’s Manual 2 - 9...
  • Page 32 Battery status: indicates the status of the battery. For details, refer to 15 Battery. System time Tabular trend area. This area displays only in Trend screen mode. NIBP list area. This area displays only in NIBP List screen mode. 2 - 10 Accutorr 7 Operator’s Manual...
  • Page 33: Menu

    Heading: provide a title or description for the current menu. Main body: displays options, buttons, prompt messages, etc. A menu button followed by “>>” opens a secondary window to reveal more options or information. : select to exit the current menu. Accutorr 7 Operator’s Manual 2 - 11...
  • Page 34: Operating Modes

    The Monitor mode and Spot Check modes have all the features in common except the following: Functions Monitor Mode Spot Check Mode Configure and use Sat-Seconds (Nellcor) Access [Alarm Setup] tab Access [Continuous Trends] tab Access [Graphic Trends] tab Connect to the CMS 2 - 12 Accutorr 7 Operator’s Manual...
  • Page 35: Standby Mode

    Select [Main] → [Maintenance>>]. Select [Exit Demo]. WARNING • The Demo mode is for demonstration purpose only. To avoid the potential risk of the simulated data being mistaken for the monitored patient’s data, Accutorr 7 Operator’s Manual 2 - 13...
  • Page 36 Demo mode while monitoring a patient. Otherwise, improper patient monitoring and delayed treatment could result. 2 - 14 Accutorr 7 Operator’s Manual...
  • Page 37: Basic Operation

    • Use only screws and brackets specified by Mindray, or the screw could touch the internal battery, causing monitor damage. CAUTION •...
  • Page 38: Unpacking And Checking

    Mindray. If the packing case is intact, open the package and remove the equipment and accessories carefully. Check all materials against the packing list and check for any mechanical damage. Contact Mindray in case of any problem. CAUTION •...
  • Page 39: General Operation

    TEMP probe connector on the back of the monitor; if you have not already done so, insert the CO adapter into the CO receptacle on the side of the Accutorr 7 Operator’s Manual 3 - 3...
  • Page 40: Turning On/Off Power

    Contact your service personnel or Mindray. NOTE • Carefully check if the system performs the self-test as described above. Contact your service personnel or Mindray if the self-test is abnormal. 3.3.2 Turning off the Monitor Before turning off the monitor: Ensure that monitoring of the patient has been completed.
  • Page 41: Using Key, Knob, Touchscreen

    ■ Pop-Up Keys: Pop-up keys are task-related menu keys that appear automatically on the monitor screen when required. For example, the confirm pop-up key appears only when you need to confirm a change. Accutorr 7 Operator’s Manual 3 - 5...
  • Page 42: Using The Knob

    The available volume range is 0 to 10. 0 means off, and 10 the maximum volume. ■ Select [Start] or [Pause] to start or pause the timer. ■ Select [Clear] to clear current timing result. 3 - 6 Accutorr 7 Operator’s Manual...
  • Page 43: Changing General Settings

    Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required password → Select [Ok] and then access [User Settings] menu. Select [Patient Demographics]. Set [Display Patient Name] to [Yes] or [No]. ◆ [Yes]: the patient’s name display in the main screen. Accutorr 7 Operator’s Manual 3 - 7...
  • Page 44: Adjusting Alarm Volume

    Select [Main] → [General Setup>>]. Select [Display Setup] and set the main screen to different layout: All Parameters display, Trend display or NIBP List display. ◆ All Parameters display includes parameter area, waveform area, but no trend. 3 - 8 Accutorr 7 Operator’s Manual...
  • Page 45: Configuring The Timeout Of Measured Value

    Changing the Time Zone To set the time zone, follow this procedure: Select the [Main] → [Maintenance>>] → [User Settings>>]→ Enter the required password → Select [Ok]. Select the [Others] → [Time Setup>>]. Accutorr 7 Operator’s Manual 3 - 9...
  • Page 46: Enabling Auto Daylight Savings Time

    You can select the items to be printed on the strip. By default, all items are selected. To enter the [Print Setup] menu, select [Main] → [General Setup>>] → [Print Setup>>] and then the [Print Setup] menu pops up. 3 - 10 Accutorr 7 Operator’s Manual...
  • Page 47: Patient Data Management

    ◆ If a matching patient MRN is found in the ADT server, the monitor will automatically populate the [Patient Demographics] fields according to the patient data stored in the ADT server. Accutorr 7 Operator’s Manual 4 - 1...
  • Page 48: Admitting A Patient By Barcode Scanner

    To admit a patient with a 1D barcode scanner, follow this procedure: Connect the barcode scanner to the USB connector on the monitor. Aim the barcode scanner at the barcode. Select [Ok] on the [Patient Demographics] menu to admit the patient. 4 - 2 Accutorr 7 Operator’s Manual...
  • Page 49 [Patient Demographics]: the [Patient Demographics] menu displays. ◆ [ADT Database]: the [ADT Database] tab displays. NOTE • If there is existing open menu or window, the [Patient Demographics] menu or [ADT Database] tab will not pop up. Accutorr 7 Operator’s Manual 4 - 3...
  • Page 50: Admitting A Patient From [Patient List]

    When the monitor is connected to ADT database through the eGateway, the monitor can obtain patient information from the ADT database. Access the [ADT Database] screen by either way. ◆ Select [Patient List] QuickKey → [ADT Database] tab. 4 - 4 Accutorr 7 Operator’s Manual...
  • Page 51: Manually Input Patient Data

    Select [Main] → [Maintenance >>] → [User Settings >>] → enter the required password → Select [Ok] to access [User Settings] menu. Select [Manual Inputs>>]. The following screen displays. Accutorr 7 Operator’s Manual 4 - 5...
  • Page 52: Setting The Monitor Location

    Select [Main] → [Maintenance >>] → [User Settings >>] → enter the required password → Select [Ok] to access [User Settings] menu. Input the following location of the monitor: ◆ [Facility]: your facility name. ◆ [Department]: your department name. ◆ [Bed No.]: bed number. 4 - 6 Accutorr 7 Operator’s Manual...
  • Page 53: Manually Save Patient Data

    ■ Scoring review In Spot Check mode, you can view Spot Check Trends and Score Review. 4.5.1 Spot Check Trends Select the [Spot Check Trends] tab to access the Spot Check Trends screen. Accutorr 7 Operator’s Manual 4 - 7...
  • Page 54 Select [Send] to transmit the selected patient’s trend data to the EMR through an eGateway or HL7. NOTE • The trend data can only be sent out when the monitor is connected to Electronic Medical Record system (EMR). In the spot check tabular trends: 4 - 8 Accutorr 7 Operator’s Manual...
  • Page 55: Continuous Trends

    ), and then use up or down arrow beside the key to turn the page. ■ Select [Print] to print the trend data of selected patient. ■ Select [Delete All] to delete the trend data of selected patient. Accutorr 7 Operator’s Manual 4 - 9...
  • Page 56: Graphic Trends

    Events are marked with colors in the event mark area. Red represents a high level alarm event. Yellow represents a medium/low level alarm event. In the Graphic Trends screen, you can: ■ Select [Zoom] to set the time length of the trend displayed on the screen. 4 - 10 Accutorr 7 Operator’s Manual...
  • Page 57: Scoring Review

    4.5.4 Scoring Review Select [Scoring Review] tab to access the Scoring Review screen. The Scoring Review screen displays the patient's scoring history. Scoring list Button area In this tab, you can: Accutorr 7 Operator’s Manual 4 - 11...
  • Page 58: Transferring Data From The Monitor To Usb Drive

    Do not remove the storage medium during data transfer process. Otherwise, data files may be damaged. • The normal monitoring function will be affected during data export. Do not perform any monitoring activity during data export. 4 - 12 Accutorr 7 Operator’s Manual...
  • Page 59: Managing Configurations

    For the important configuration items and their default values and user maintenance items, refer to C Default Configurations. Accessing [Manage Configuration] Menu Select [Main] → [Maintenance >>] → [User Settings >>]. Enter the required password and then select [Ok]. Select [Manage Configuration >>]. Accutorr 7 Operator’s Manual 5 - 1...
  • Page 60: Setting Default Configuration

    Selecting [Config on USB drive >>] will show the existing user configurations on the USB drive. Select the user configurations you want to delete and then select [Delete]. Select [Yes] in the popup. 5 - 2 Accutorr 7 Operator’s Manual...
  • Page 61: Transferring A Configuration

    The monitor restores the latest configuration if it restarts within 60 seconds after a power failure. And it restores the default configuration, rather than the latest configuration, if it restarts 120 seconds after a power failure. The monitor loads either Accutorr 7 Operator’s Manual 5 - 3...
  • Page 62 60-120 seconds after a power failure. 5 - 4 Accutorr 7 Operator’s Manual...
  • Page 63: Alarms

    Apart from the physiological and technical alarm messages, the monitor shows some messages telling the system status or patient status. System related messages are displayed in system message area; parameter related messages are displayed in the respective parameter message area. Accutorr 7 Operator’s Manual 6 - 1...
  • Page 64: Alarm Levels

    • When multiple alarms of different levels occur simultaneously, the alarm lamp flashes and the alarm tone sounds according to the alarm of the highest level, and the alarm messages are displayed circularly. 6 - 2 Accutorr 7 Operator’s Manual...
  • Page 65: Audible Alarm Tones

    For physiological alarms, the asterisk symbols (*) before the alarm message match the alarm level as follows: ■ High level alarms:*** ■ Medium level alarms:** ■ Low level alarms:* When there are multiple messages, the messages scroll. Accutorr 7 Operator’s Manual 6 - 3...
  • Page 66: Flashing Numerics

    Refer to 5.2.1 Setting Default Configuration. 6 - 4 Accutorr 7 Operator’s Manual...
  • Page 67: Selecting Alarm Properties

    → Select [Ok] to access [User Settings] menu. Select [Alarm Setup >>] to access [Alarm Setup] menu. Select [High Alarm Interval (s)], [Med Alarm Interval (s)] and [Low Alarm Interval (s)] in turn and then select the appropriate settings. Accutorr 7 Operator’s Manual 6 - 5...
  • Page 68: Adjusting Alarm Limits Automatically

    (PR-30) or 90 (PR×1.25) or (PR+40) or Adult/Pediatric: bpm, bpm, 240 bpm, 200 bpm, 35 to 240 bpm whichever is whichever is whichever is whichever is Neonate: greater greater smaller smaller 55 to 225 bpm 6 - 6 Accutorr 7 Operator’s Manual...
  • Page 69 30 to 245 whichever is whichever is mmHg greater smaller Pediatric: 30 to 180 mmHg Neonate: 25 to 105 mmHg Temp (T-0.5) °C (T-0.5) °C (T+0.5) °C (T+0.5) °C Same as the measurement range Accutorr 7 Operator’s Manual 6 - 7...
  • Page 70: Pausing Alarms

    2 seconds. When alarms are paused, ■ The alarms pause symbol and the remaining alarm pause time are displayed in Alarm Information area. ■ For physiological alarms, 6 - 8 Accutorr 7 Operator’s Manual...
  • Page 71: Switching Off Alarm Sound

    Select [Main] →[Alarm Setup >>]. Set [No CO2 Change Detected] to [Off]. WARNING • Switching off [No CO2 Change Detected] alarm may result in a hazard to the patient. Always keep the patient under close surveillance. NOTE Accutorr 7 Operator’s Manual 6 - 9...
  • Page 72: Resetting Alarms

    Some technical alarms are changed to the prompt messages. ■ Some technical alarms are cleared. The monitor gives no alarm indications. For details about the indications of technical alarms when the alarm system is reset, refer to D.2 Technical Alarm Messages 6 - 10 Accutorr 7 Operator’s Manual...
  • Page 73: Setting The Reminder Tone

    [All]: all alarms are latched; and ◆ [Off ]: no alarm will be latched. Only the unacknowledged physiological alarm can be latched. The latched alarms will be cleared when the monitor enters the alarm reset state. WARNING Accutorr 7 Operator’s Manual 6 - 11...
  • Page 74: Actions For Alarm Occurrence

    [Normally Closed]: Select if your hospital’s nurse call relay contact is normally closed. Select [Alarm Level] and set the alarm level for nurse call-triggering alarms. Select [Alarm Category] and then select the category to which the nurse call- triggering alarms belong. 6 - 12 Accutorr 7 Operator’s Manual...
  • Page 75 • Do not rely exclusively on the nurse call system for alarm notification. Remember that the most reliable alarm notification combines audible and visual alarm indications with the patient’s clinical condition. Accutorr 7 Operator’s Manual 6 - 13...
  • Page 76 This page intentionally left blank. 6 - 14 Accutorr 7 Operator’s Manual...
  • Page 77: Monitoring Spo

    PI is an indicator of the pulsatile strength. You can also use it to assess the quality of SpO measurement. ◆ Above 1 is optimal ◆ Between 0.3 and 1 is acceptable. Accutorr 7 Operator’s Manual 7 - 1...
  • Page 78: Safety

    Change the application site every four hours. For neonates, or patients with poor peripheral blood circulation or sensitive skin, inspect the sensor site more frequently. 7 - 2 Accutorr 7 Operator’s Manual...
  • Page 79: Identifying Spo Module

    Clean the application site. For example, remove colored nail polish. Apply the sensor to the patient. Select an appropriate adapter cable according to the connector type and plug this cable into the SpO connector. Connect the sensor cable to the adapter cable. Accutorr 7 Operator’s Manual 7 - 3...
  • Page 80: Changing Spo Settings

    [NIBP Simultaneous] on in the [SpO2 Setup] menu to lock the SpO alarm status until the NIBP measurement ends. If you switch [NIBP Simultaneous] off, low perfusion caused by NIBP measurement may lead to inaccurate SpO readings and therefore cause false physiological alarms. 7 - 4 Accutorr 7 Operator’s Manual...
  • Page 81: Sat-Seconds Alarm Management

    During such fluctuation, the monitor integrates the number of %SpO points, both positive and negative, until either the Sat- Seconds limit is reached, or the patient %SpO re-enters the non-alarm range and remains there. Accutorr 7 Operator’s Manual 7 - 5...
  • Page 82: Changing The Speed Of Pleth Wave

    Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized sensors or cables which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device. 7 - 6 Accutorr 7 Operator’s Manual...
  • Page 83: Nellcor Information

    Possession or purchase of this device does not convey any express or implied license to use the device with unauthorized replacement parts which would, alone, or in combination with this device, fall within the scope of one or more of the patents relating to this device. Accutorr 7 Operator’s Manual 7 - 7...
  • Page 84 This page intentionally left blank. 7 - 8 Accutorr 7 Operator’s Manual...
  • Page 85: Monitoring Pr

    You can change the pulse tone volume by adjusting [Beat Volume] in the [SpO2 Setup] menu. When a valid SpO value exists, the system will adjust the pulse tone according to the SpO value. Accutorr 7 Operator’s Manual 8 - 1...
  • Page 86 This page intentionally left blank. 8 - 2 Accutorr 7 Operator’s Manual...
  • Page 87: Monitoring Nibp

    American National Standard, Manual, electronic, automated sphygmomanometers, or standards of IEC80601-2-30, EN1060-1, EN1060-3, EN1060-4 and SP10. Accutorr 7 Operator’s Manual 9 - 1...
  • Page 88: Safety

    ■ If arterial pressure pulses are hard to detect ■ In the presence of excessive and continuous patient movement such as shivering or convulsions ■ During certain cardiac arrhythmias 9 - 2 Accutorr 7 Operator’s Manual...
  • Page 89: Nibp Measurement Mode

    • It is recommended that the patient sit still for 5 min before the first measurement is taken. • The operator should not touch the cuff and tubing during the NIBP measurement. Accutorr 7 Operator’s Manual 9 - 3...
  • Page 90: Preparing To Measure Nibp

    Access [NIBP Setup] menu, and then select [Start NIBP] key to start a manual, programmed or automatic NIBP measurement with preset interval. ■ Access [NIBP Setup] menu, and then select [NIBP STAT] to start a continuous NIBP measurement. 9 - 4 Accutorr 7 Operator’s Manual...
  • Page 91: Stopping Nibp Measurement

    The NIBP display generally shows numerics as shown below. Your display may be configured to look slightly different. Unit of pressure: mmHg or kPa Systolic pressure Diastolic pressure Mean pressure obtained after the measurement and cuff pressure obtained during the measurement Accutorr 7 Operator’s Manual 9 - 5...
  • Page 92: Setting Nibp

    Setting NIBP End Tone The monitor can issue a reminder tone at the completion of NIBP measurement. The NIBP End Tone is off by default. You can switch it on by accessing the [NIBP Setup] menu. 9 - 6 Accutorr 7 Operator’s Manual...
  • Page 93: Switching On/Off Measurement On Clock

    Time remaining to next measurement Program name Cycle name NIBP measurement Interval 9.7.6 Setting NIBP Alarm Properties Select [Alarm Setup>>] from [NIBP Setup] menu. You can set the alarm properties for this parameter in the popup menu. Accutorr 7 Operator’s Manual 9 - 7...
  • Page 94: Setting The Pressure Unit

    Hand or machine washes the cuff in warm water or with mild detergent. Clean the bladder with a damp cloth. Air dry the cuff thoroughly after washing. NOTE • Machine washing may shorten the service life of the cuff. 9 - 8 Accutorr 7 Operator’s Manual...
  • Page 95: Disinfection

    Antimicrobial Definition Bladderless cuffs are treated with an antimicrobial coating. Antimicrobial technology effectively controls a broad spectrum of bacteria, fungi, algae and yeasts on a wide variety of treated substrates. Accutorr 7 Operator’s Manual 9 - 9...
  • Page 96 This page intentionally left blank. 9 - 10 Accutorr 7 Operator’s Manual...
  • Page 97: Monitoring Temp

    Ensure that probe covers are disposed of according to local regulations or hospital's requirements. • Accuracy verification of the temperature module is required every two years or according to your hospital’s policy. Please contact Mindray Technical Support department if accuracy verification is needed. Accutorr 7 Operator’s Manual 10 - 1...
  • Page 98: Setting Temp

    ■ rectal probe (red) Use the blue oral/axillary probe with blue probe well, and use the red rectal probe with red well. Be sure to select correct probe according to the measurement site. 10 - 2 Accutorr 7 Operator’s Manual...
  • Page 99: Taking A Temperature In Predictive Mode

    The monitor sounds a beep when the temperature measurement is complete. Withdraw the probe. Firmly press the ejection button on the top of the probe to eject the probe cover. Replace the probe into the probe well. Accutorr 7 Operator’s Manual 10 - 3...
  • Page 100: Taking A Temperature In Monitor Mode

    In Monitor mode, record the measured value prior to taking the probe away from the measurement site. The monitor will automatically stop measuring temperature after 10 minutes from the start of the measurement. 10 - 4 Accutorr 7 Operator’s Manual...
  • Page 101: Disinfecting Temperature Probe

    Wipe off all the remaining disinfectants from the probe with a soft cloth dampened with water. Dry the probe in a cool place. WARNING • Properly dispose of the used soft cloth. Accutorr 7 Operator’s Manual 10 - 5...
  • Page 102 This page intentionally left blank. 10 - 6 Accutorr 7 Operator’s Manual...
  • Page 103: Monitoring Co

    Inspect the airway for a tight connection and make proper settings before attaching it to the patient. • Squeezing or bending the sampling line during the CO measurement may cause inaccurate CO readings or no reading. Accutorr 7 Operator’s Manual 11 - 1...
  • Page 104: Measurement Limitations

    Fraction of inspired CO (FiCO Airway respiration rate (RR) waveform Scale 11.5 Automatic CO Module Zeroing The CO module performs zero calibration automatically when needed. NOTE • The CO module temporally stops measuring during zeroing. 11 - 2 Accutorr 7 Operator’s Manual...
  • Page 105: Measuring Co

    Connect the other end of the sampling line to the patient. ◆ For intubated patients requiring an airway adapter, install the airway adapter between the patient circuit and the ventilator Y-piece. Sampling line Connect to the ventilator Airway adapter Connect to the patient Accutorr 7 Operator’s Manual 11 - 3...
  • Page 106: Changing Co Settings

    “No CO2 Changes Detected” Delay. To set the delay time, follow this procedure: Select the CO numeric area or waveform area to enter the [CO2 Setup] menu. Select [Delay Time] and then select the appropriate setting. 11 - 4 Accutorr 7 Operator’s Manual...
  • Page 107: Setting The Co Waveform

    ] mode when not using the CO module. The Accutorr 7 monitor is preset to monitor when the it is turned on and the CO accessory is connected. The preset operating mode of the CO module is Measure. It is...
  • Page 108: Setting Humidity Compensation

    Calibrating the CO Module A calibration should be performed once a year or when the readings go far beyond the specified measurement range. For details on calibrating the CO module, refer to 16.7 Calibrating CO 11 - 6 Accutorr 7 Operator’s Manual...
  • Page 109: Clinical Scoring

    Obstructive Pulmonary Disease) patients and those under 16 years old. *Please also refer to Subbe CP, Kruger M, Rutherford P, Gemmel L. Validation of a modified Early Warning Score in medical admissions. Quarterly Journal of Medicine 2001; 94: 521–526. Accutorr 7 Operator’s Manual 12 - 1...
  • Page 110: Mews (Modified Early Warning Score)

    Individual parameter scoring (IPS): provide a clinical response whenever any individual parameter value is out of range. The available parameters in customizable scoring include: ■ Respiration Rate ■ ■ Supplemental Oxygen ■ Temperature ■ Systolic NIBP 12 - 2 Accutorr 7 Operator’s Manual...
  • Page 111: Entering The Calculation Screen

    Inspired O ■ Airway ■ Three customizable parameters The applicable patient category can be defined with the Mindray Clinical Scoring Config Tool. Refer to Clinical Scoring Config Instruction for Use (P/N: 046-007126-00) for customizable scorings. 12.2 Entering the Calculation Screen You can calculate a score in the scoring screen or in the score tile.
  • Page 112: Scoring Screen

    The MEWS, NEWS or multi-parameter score screen display as follows: Icon for manual input You need to manually input the parameter. Icon for real-time monitoring The parameter value is from the monitor Recommended clinical response 12 - 4 Accutorr 7 Operator’s Manual...
  • Page 113: Obtaining The Total Score

    When you are in the scoring screen, if an alarm for the monitoring parameter occurs, the alarm indicators still can be seen in the monitor display. The parameter input range is as follows: Parameter Range Pulse Rate 20 bpm -300 bpm Accutorr 7 Operator’s Manual 12 - 5...
  • Page 114: Unfreezing The Screen

    Clear, Obstruction Customizable The input range depends on the decimal point. The parameter decimal point is customizable in Mindray Clinical Scoring Config Tool. 0 – 9999 (decimal point as 1) 0.0 – 999.9 (decimal point as 0.1) 0.00 – 99.99 (decimal point as 0.01) 12.6...
  • Page 115: Managing Scorings

    Set the default scoring for the applicable patient category. After the default scoring is set, when a patient category is changed, the monitor will automatically use the default scoring. 12.7.4 Deleting a Scoring To delete the scoring: Accutorr 7 Operator’s Manual 12 - 7...
  • Page 116: Loading A Scoring

    The default scoring may not be appropriately for the new patient. You can load a scoring so as that the scoring is appropriate for your patient. To load a scoring: Select [Scoring] → [Setup] → [Load Scoring>>]. Select the desired scoring to be loaded. Select [Load Scoring]. 12 - 8 Accutorr 7 Operator’s Manual...
  • Page 117: Recording

    13.2 Loading Paper Press the latch in the upper right corner of the recorder door to open the door. Insert a new roll into the compartment as shown below. Close the recorder door. Accutorr 7 Operator’s Manual 13 - 1...
  • Page 118: Setting The Recorder

    [Automatically Record on Manual Save] is enabled from the [User Settings>>] menu. ■ To manually stop the current recording, you select hardkey. Recordings stop automatically when: ■ A recording is completed. 13 - 2 Accutorr 7 Operator’s Manual...
  • Page 119: Reports

    Select [Record] to start recording. In the continuous trends strip: ■ The mark ‘?’ before the time indicates that system time is changed. ■ The mark ‘*’ before the parameter value indicates that an alarm is triggered. Accutorr 7 Operator’s Manual 13 - 3...
  • Page 120: Spot Check Trends Recording

    After the alcohol has completely been dried, reload the paper and close the recorder door. Refer to 13.2 Loading Paper for additional information. CAUTION • Do not use anything that may destroy the thermal element. • Do not add unnecessary force to the thermal head. 13 - 4 Accutorr 7 Operator’s Manual...
  • Page 121: Network

    The monitor can be connected to the ADT system and the EMR system through the eGateway. ■ The Accutorr 7 transmits the query command message to the ADT system. When the ADT system receives the query command message, the ADT system will transmit the patients’ information to the Accutorr 7.
  • Page 122: Selecting A Network Type

    ◆ If [Address Type] is set to [Manual], you need to manually input the monitor IP address, subnet mask and gateway address. Select [WLAN Setup >>] to access the WLAN setup menu. 14 - 2 Accutorr 7 Operator’s Manual...
  • Page 123: Wlan Test

    The design, installation and maintenance of the wireless network’s distribution should be performed by authorized personnel. • In a wireless (optional) network, the existence of obstacles (such as walls) will affect data transmission or even cause wireless network interruption. Accutorr 7 Operator’s Manual 14 - 3...
  • Page 124: Selecting Wlan Band And Channels

    The total throughput of all the wireless devices connected to the wireless network should be less than the effective transmitting capability of the wireless network. The throughput capacity of a single Accutorr 7 is 700 kbps. • Do not connect non-approved devices to the wireless network.
  • Page 125: Setting Data Send Method

    [Query from Local Facility] to [No], you can query patient information from all networked facilities. Select [Ok] to exit the menu. NOTE • The domain name supports only English characters, symbols and numbers. Accutorr 7 Operator’s Manual 14 - 5...
  • Page 126: Emr Communication Setup

    Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required password → Select [Ok] to access User Settings menu. Select [Network>>] → [Alarm Server Setup]. Set the IP address and port. 14 - 6 Accutorr 7 Operator’s Manual...
  • Page 127: Setting The Data Encryption Switch

    Select [Main] → [Maintenance>>] → [User Settings >>] → enter the required password → Select [Ok] → [Network >>]. Set [Select CMS] to [On]. Select [Central Station Setup >>]. Set CMS names and corresponding IP addresses. Accutorr 7 Operator’s Manual 14 - 7...
  • Page 128: Clearing The Selected Cms At Startup

    Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required password → Select [Ok] → [Others >>] Select [DIAP Communication>>] to access [DIAP Communication Setup] menu. Set the baud rate. Select [Ok] to exit the menu. 14 - 8 Accutorr 7 Operator’s Manual...
  • Page 129: Mldap

    14.6 MLDAP MLDAP refers to Mindray LDAP (Lightweight Directory Access Protocol). It is an independent process which can be installed on eGateway or other application server (Windows). MLDAP provides user identity and authentication. The MLDAP server is connected to the hospital’s LDAP or AD server. All monitoring...
  • Page 130 Local Alarm Setup Password>>], enter the new password, and select [OK]. To change the monitor’s local password for accessing the [User Settings] menu, select [Modify Local User Maintenance Password>>], enter the new password, and select [OK]. 14 - 10 Accutorr 7 Operator’s Manual...
  • Page 131: Battery

    Charge the battery or replace the battery with a fully charged battery when the monitor gives the alarm [Low Battery] or [Battery Depleted]. The device will not allow the user to enter Standby mode when the battery power is low. Accutorr 7 Operator’s Manual 15 - 1...
  • Page 132: Charging A Battery

    Do not store the battery at a temperature beyond -20 °C - 60 °C (-4°F - 140°F). CAUTION • Keep the battery out of children’s reach. 15 - 2 Accutorr 7 Operator’s Manual...
  • Page 133: Battery Maintenance

    Again connect the monitor to the AC power. Fully charge the battery again for use or charge it to 40 – 60% for storage. Accutorr 7 Operator’s Manual 15 - 3...
  • Page 134: Checking A Battery

    Note the time again. Calculate the run time by subtracting the start time from the end time. The operating time of a battery directly reflects its performance. If the operating time of a battery is noticeably shorter than that stated in the specifications, contact Mindray or your service personnel. NOTE •...
  • Page 135 WARNING • Do not disassemble batteries, dispose of them in fire, or cause them to short circuit. They may ignite, explode, leak or heat up, causing personal injury. Accutorr 7 Operator’s Manual 15 - 5...
  • Page 136 This page intentionally left blank. 15 - 6 Accutorr 7 Operator’s Manual...
  • Page 137: Care, Cleaning And Maintenance

    Otherwise, undue equipment failure and possible health hazards could result. • If you discover a problem with any of the equipment, contact Mindray or your service personnel. • The responsible hospital or institution shall carry out all cleaning and disinfection procedure specified in this chapter.
  • Page 138: Cleaning

    For infection control methods, consult your hospital’s Infection Control Officer or Epidemiologist. CAUTION • If you spill liquid on the equipment or accessories, contact Mindray or your service personnel. 16.1.1 Cleaning Clean your equipment on a regular basis. Consult your hospital's regulations before cleaning the equipment.
  • Page 139: Sterilization

    Care should be taken to change the settings in [User Settings >>] and [Factory Maintenance >>] menus to avoid data loss. • Service personnel should acquaint themselves with the test tools and make sure that test tools and cables are applicable. Accutorr 7 Operator’s Manual 16 - 3...
  • Page 140 Following any repair or replacement of the Recorder check recorder. 1. When first installed. 2. Whenever a battery is replaced. Functionality test Battery check Performance test Once a year or if the battery run time reduced significantly. 16 - 4 Accutorr 7 Operator’s Manual...
  • Page 141: Checking Monitor Information

    Inspect all external connections for loose connectors, bent pins or frayed cables. Inspect all connectors on the equipment for loose connectors or bent pins. Make sure that safety labels and data plates on the equipment are clearly legible. Accutorr 7 Operator’s Manual 16 - 5...
  • Page 142: Power-On Test

    Select [Main]→ [Maintenance >>]→ [User Settings >>]→ enter the required password→[Module Maintenance >>]→ [Calibrate CO2 >>]→ [Zero]. After the zero calibration is finished successfully, connect the equipment as follows: 16 - 6 Accutorr 7 Operator’s Manual...
  • Page 143: Battery Check

    [Calibrate CO2] menu. If the calibration fails, the message [Calibration Failed!] is displayed. In this case, attempt to calibrate the CO module again. If the problem still exists, contact Mindray service personnel. 16.8 Battery Check Refer to 15.5.2 Checking a Battery for battery check instructions.
  • Page 144: 16.10 Formatting The Storage Card

    The monitor is configured with an SD card for saving data. To format the storage card: Select [Main] → [Maintenance >>] → [User Settings >>] → Enter the required password → Select [Ok] → [Others >>]. Select [Format Storage Card], and then select [Ok] in the pop-up dialog. 16 - 8 Accutorr 7 Operator’s Manual...
  • Page 145: Accessories

    Remarks Part No. Masimo SpO 8-pin Masimo LNCS Cable, 2.1 m 115-020768-00 Masimo SpO RD to LNC Adapter Cable 040-003381-00 Masimo SpO 8-pin Masimo RD Cable 040-003310-00 Nellcor SpO 8-pin, 2.5 m 0010-20-42712 Accutorr 7 Operator’s Manual 17 - 1...
  • Page 146 Adult (> 40 kg, finger Disposable or toe) 4003 040-003385-00 Neonate (< 3 kg, finger or toe) Neonate (< 1 kg, foot 040-003386-00 or hand) Neonate (< 1 kg, foot 4005 040-003387-00 or hand) 17 - 2 Accutorr 7 Operator’s Manual...
  • Page 147: Nibp Accessories

    115-027714-00 Adult 24 to 35 115-027715-00 Large Adult 33 to 47 115-027716-00 Adult Thigh 46 to 66 115-027717-00 Adult Long 24 to 35 115-027718-00 Large Adult 33 to 47 115-027719-00 Starter kit 115-031807-00 Accutorr 7 Operator’s Manual 17 - 3...
  • Page 148: Temp Accessories

    Temp Accessories Probe Well Type Description Part No. Blue, Oral/Axillary M09A-20-62062 Reusable Red, Rectal M09A-20-62062-51 Temp Probes Patient Type Measurement Site Part No. Category Adult, Pediatric, Oral/Axillary 6006-30-39598 Reusable Neonate Adult, Pediatric Rectal 6006-30-39599 17 - 4 Accutorr 7 Operator’s Manual...
  • Page 149: Co Accessories

    Rolling Stand with Quick Release Mount 045-002960-00 Bin for Rolling Stand with Cable Management Accessories 045-002840-00 Bin for Rolling Stand 115-037732-00 Casters for Rolling Stand 034-000302-00 Casters for Rolling Stand 0401-00-0055 Wall Mount Bracket 045-001059-00 Accutorr 7 Operator’s Manual 17 - 5...
  • Page 150 Barcode Scanner Kit 1D Bar Code Scanners 023-000254-00 (115-008393-00) USB Cable of ASYMBOL 009-001397-00 Scanner Accutorr 7 Welch Allyn Accutorr 7 Welch Allyn Temp 115-022908-00 Temp Support Kit Support Assembly (115-025041-00) Accutorr 7 Welch Allyn Temp 046-006016-00 Support Installation Guide...
  • Page 151: A Product Specifications

    0 to 40 -20 to 60 5 to 40 (with Temp module) 10 to 40 (with CO module) Relative humidity 15% to 95% 10% to 95% (noncondensing) Barometric (kPa) 57.0 to 107.4 16.0 to 107.4 Accutorr 7 Operator’s Manual A - 1...
  • Page 152: Power Supply Specifications

    Monitor power off: less than 5.5 hours to 90%; less than 6.5 hours to 100%. Monitor power on: less than 10.5 hours to 90%; less than 11.5 hours to 100%. Shutdown delay At least 20 minutes (after a low battery alarm first occurs) A - 2 Accutorr 7 Operator’s Manual...
  • Page 153: Physical Specifications

    1 (green) A.5.2 Audio Indicator Speaker Give alarm tones (45 to 85 dB), key tones, pulse tone; power-on self check tone, support PITCH TONE and multi-level tone modulation; alarm tones comply with IEC 60601-1-8. Accutorr 7 Operator’s Manual A - 3...
  • Page 154: Monitor Interface Specifications

    For 2.4G frequency bands (FCC): 2.4GHz to 2.483GHz (only use Channels 1 to 11) For 5G frequency bands (FCC): 5.15GHz to 5.35GHz, 5.725GHz to5.825GHz Priority All communication data types shall have the same priority. A - 4 Accutorr 7 Operator’s Manual...
  • Page 155 Accutorr 7 moves from the coverage area of AP1 to that of AP2. System capacity Number of the Accutorr 7 supported by a single AP is ≤ 16. Each Accutorr 7 can communicate with the CMS. Data security Security standards: WPA-PSK, WPA2-PSK, WPA-...
  • Page 156: Measurement Specifications

    60% to 95%) averaging time 2-4 s, 4-6 s, 8 s, 10 s, 12 s, 14 s, 16 s Low perfusion conditions Pulse amplitude: >0.02% Light penetration: >5% Low perfusion SpO accuracy ±2% A - 6 Accutorr 7 Operator’s Manual...
  • Page 157 ±1%, to compensate for the theoretical effect on oximeter measurements of fetal hemoglobin in neonatal blood. Information of the Test Subjects of the Clinical Study Report Skin color Gender Number Age (years) Health Black Male 28.2±9.19 Healthy Female Yellow Male Female Accutorr 7 Operator’s Manual A - 7...
  • Page 158: A.7.2 Pr

    20 to 250 bpm: ±3 bpm 251 to 300 bpm, not specified Refreshing rate PR from NIBP Module Measurement range 30 to 300 bpm Resolution 1 bpm Accuracy ±3 bpm or ±3%, whichever is greater A - 8 Accutorr 7 Operator’s Manual...
  • Page 159: Nibp

    0 mmHg to 300 mmHg measurement range Static pressure ±3 mmHg measurement accuracy Resolution 1mmHg Initial cuff Adult: 80 to 280 inflation Pediatric: 80 to 210 pressure range Neonate: 60 to 140 (mmHg) Accutorr 7 Operator’s Manual A - 9...
  • Page 160 American National Standard for Electronic or Automated Sphymomanometers (ANSI/AAMI SP10) in terms of mean error and standard deviation by comparing with intra-arterial measurements (depending on the configuration) in a typical patient population. A - 10 Accutorr 7 Operator’s Manual...
  • Page 161: Temp

    Meets standard of ISO 80601-2-55 Technique Infrared absorption [No CO2 Changes 10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s Detected] delay time Measurement range 0 to 20% (0 to 150 mmHg) Accutorr 7 Operator’s Manual A - 11...
  • Page 162 1 to 99 mmHg awRR High Adult, pediatric: (low limit + 1 rpm 2) to 100 rpm Neonate: (low limit + 2) to 150 rpm awRR Low 0 to (high limit - 2) rpm A - 12 Accutorr 7 Operator’s Manual...
  • Page 163 Inaccuracy specifications are affected by the breath rate and I: E change. The EtCO accuracy is within specification for breath rate ≤ 60rpm and I/E ratio ≤1:1, or breath rate ≤30rpm and I/E ratio ≤ 2:1. Accutorr 7 Operator’s Manual A - 13...
  • Page 164 This page intentionally left blank. A - 14 Accutorr 7 Operator’s Manual...
  • Page 165: B Emc And Radio Regulatory Compliance

    Portable and mobile communication equipment may affect the performance of this monitor. • Other devices that have RF transmitter or source may affect this device (e.g. cell phones, PADs, PCs with wireless function). Accutorr 7 Operator’s Manual B - 1...
  • Page 166 This equipment/ system may cause radio interference or may disrupt the operation of nearby equipment. It may be necessary to take mitigation measures, such as re-orienting or relocating the [ME EQUIPMENT or ME SYSTEM] or shielding the location. B - 2 Accutorr 7 Operator’s Manual...
  • Page 167 Note: U is the A.C. mains voltage prior to application of the test level. Accutorr 7 Operator’s Manual B - 3...
  • Page 168 ME SYSTEM] should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the [ME EQUIPMENT or ME SYSTEM]. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than3V/m. B - 4 Accutorr 7 Operator’s Manual...
  • Page 169 Note 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies. Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and people. Accutorr 7 Operator’s Manual B - 5...
  • Page 170: Radio Regulatory Compliance

    Operation of this equipment requires the prior coordination with a frequency coordinator designated by the FCC for the Wireless Medical Telemetry Service. This Wi-Fi device complies with Canadian ICES-001. Cet appareil ISM est conforme a la norme NMB-001 du Canada. B - 6 Accutorr 7 Operator’s Manual...
  • Page 171 (2) this device must accept any interference received, including interference that may cause undesired operation. Any changes or modifications to this equipment not expressly approved by Mindray may cause harmful radio frequency interference and void your authority to operate this equipment.
  • Page 172 This page intentionally left blank. B - 8 Accutorr 7 Operator’s Manual...
  • Page 173: Parameter Configuration

    Manual Clock NIBP End Tone Program Program 1 Cuff Press. Adult (mmHg) in Pediatric VeniPuncture Neonate Initial Pressure Adult Pediatric Neonate NIBP-Sys High Adult (mmHg) Pediatric Neonate NIBP-Sys Low Adult (mmHg) Pediatric Neonate Accutorr 7 Operator’s Manual C - 1...
  • Page 174: Spo

    Alarm On/Off Alarm Priority High Adult/pediatric:100 Neonate: 95 Desat Low Sat-Seconds (Nellcor) NIBP Simultaneous Sweep 25 mm/sec Beat Volume Sensor Off Lev. Monitor mode: Low Spot Check mode: Off Sensitivity (Masimo) Normal Averaging (Masimo) C - 2 Accutorr 7 Operator’s Manual...
  • Page 175: Temp

    Affected by Power Failure Defaults Alarm On/Off Alarm Priority Temp High 38.0 Temp Low 35.0 Temp Type Predictive Temp Position Oral/Axillary probe: Oral for adult and pediatric Axillary for neonate Rectal probe: Rectal Accutorr 7 Operator’s Manual C - 3...
  • Page 176: C.1.5 Co

    Alarm On/Off Alarm Priority FiCO High No CO2 Changes Detected 20 s Operating Mode Measure BTPS Compensate Scale Sweep 6.25 mm/s Wave Type Draw Compensate O Compensate AA Compensate Auto Standby 60 min C - 4 Accutorr 7 Operator’s Manual...
  • Page 177: General Configuration

    Saved at Affected by Power Failure Defaults Spot Check Name/MRN Trends button Option for Name/MRN button Filter Continuous Name/MRN Trends button Option for Current patient Name/MRN button Interval 30 s Graphic Trends Zoom Accutorr 7 Operator’s Manual C - 5...
  • Page 178: Record

    Key Volume Display Setup All Parameters Parameter Time Out 15 min Height Weight Pressure mmHg Temp °C Glucose mg/dl I/O Fluid Date Current date Time Current time Date Format yyyy-mm-dd Time Format 24 h C - 6 Accutorr 7 Operator’s Manual...
  • Page 179: User Maintenance Items

    Default Config Saved at Affected by Power Failure Defaults Spot Check Tone Mode 1 Clinician ID Time Out 10 min Language English Network Type Clear CMS IP at Startup Data Send Method eGateway Accutorr 7 Operator’s Manual C - 7...
  • Page 180 This page intentionally left blank. C - 8 Accutorr 7 Operator’s Manual...
  • Page 181: D Alarm Messages

    Check High The patient stops Check the patient’s breathing. No breathing, or the condition, module and Changes respiration signal was so patient connections. Detected weak that the monitor cannot perform respiration analysis. Accutorr 7 Operator’s Manual D - 1...
  • Page 182: Technical Alarm Messages

    An error occurred to Restart the monitor. the NIBP module, or If the error remains, there is a problem contact Mindray or with the your service communications personnel. between the module and the monitor. D - 2 Accutorr 7 Operator’s Manual...
  • Page 183 NIBP Excessive Patient’s arm moves Check the patient’s Motion too much. condition and reduce the patient motion. NIBP Cuff The NIBP airway Check the airway Overpress. may be occluded. and measure again. Accutorr 7 Operator’s Manual D - 3...
  • Page 184 NIBP Illegally An illegal reset Check if the airway Reset occurred during is occluded. NIBP measurement. VeniPuncture System deflates the No operation is timeout cuff after a certain required. time. D - 4 Accutorr 7 Operator’s Manual...
  • Page 185: D.2.2 Spo2 Alarm Messages

    (Masimo, Nellcor) module and the Init Err High monitor. Board Fault There is a problem Do not use the with the SpO module and contact (Masimo) measurement Mindray or your board. service personnel. Accutorr 7 Operator’s Manual D - 5...
  • Page 186 (Masimo, Nellcor) inadvertently. personnel. PR Overrange The measured PR Contact Mindray or value exceeds the your service (Masimo, Nellcor, measurement personnel. or from NIBP) range. * means the alarm level is user-adjustable. D - 6 Accutorr 7 Operator’s Manual...
  • Page 187: Temperature Alarm Messages

    Temp Voltage Err The voltage is too Check the power high or too low. supply. Temp Prediction The measuring Retry the operation is measurement. improper. Accutorr 7 Operator’s Manual D - 7...
  • Page 188: Co Alarm Messages

    CO2 Comm High module, or there error remains, Abnormal is a problem with contact Mindray or your service CO2 Comm High communication personnel. Stop between the CO2 Init Err High module and the monitor. D - 8 Accutorr 7 Operator’s Manual...
  • Page 189 XX Overrange The measured Contact Mindray parameter or or your service measurement XX personnel. value is not within the specified range. XX represents a measurement or parameter label. Accutorr 7 Operator’s Manual D - 9...
  • Page 190: Recorder Alarm Messages

    Restore the control Pos. the recorder is in lever of the recorder wrong position. to its previous position. Recorder out of The recorder paper Replace with a new paper is used up. paper roll. D - 10 Accutorr 7 Operator’s Manual...
  • Page 191: Power Alarm Messages

    PWR interrupted. Power supply to the Check the Check meas. monitor was measurements State. interupted. when the monitor restarts. If the problem still exists, contact Mindray or your service personnel. Accutorr 7 Operator’s Manual D - 11...
  • Page 192: System Alarm Messages

    D.2.7 System Alarm Messages Alarm Message Clearable? Alarm Cause Solution Level (Yes/No) No CMS The monitor is Check network disconnected from connection. the CMS. D - 12 Accutorr 7 Operator’s Manual...
  • Page 193: E Symbols And Abbreviations

    Microsecond ampere ampere hour beat per minute bit per second ºC centigrade centimeter decibel dyne second ºF fahrenheit gram gigahertz hour hertz inch kilo kilogram kilopascal litre pound meter milliampere hour mega byte Accutorr 7 Operator’s Manual E - 1...
  • Page 194 ~ + plus equal to 〉 less than < greater than ≤ less than or equal to ≥ greater than or equal to ± plus or minus × multiply © copyright E - 2 Accutorr 7 Operator’s Manual...
  • Page 195: Abbreviations

    European Economic Community electromagnetic compatibility electromagnetic interference Electronic Medical Record enflurane end-tidal EtCO end-tidal carbon dioxide fraction of inspired FiCO fraction of inspired carbon oxygen Glasgow Comma Scale Halothane identification International Electrotechnical Commission Accutorr 7 Operator’s Manual E - 3...
  • Page 196 NIBP-M NIBP-mean pressure NIBP-S NIBP-systolic pressure nitrous oxide oxygen power photodetector pediatric Pleth plethysmogram pulse rate random access memory read-only memory sevoflurane SpO2 arterial oxygen saturation from pulse oximetry temperature difference TEMP temperature E - 4 Accutorr 7 Operator’s Manual...
  • Page 197: F Anomalies

    Although there is no distinct difference between smart inflation and normal inflation, it is recommended to keep the function [Off]. NOTE • The anomaly does not present a safety hazard to either the operator or the patient. Accutorr 7 Operator’s Manual F - 1...
  • Page 198 This page intentionally left blank. F - 2 Accutorr 7 Operator’s Manual...
  • Page 200 P/N: 046-005276-00 (7.0) SW Version: 03.10.XX...

Table of Contents

Save PDF