7 Maintenance

7.1 Cleaning

1.
Unplug product before washing.
2.
Make sure all electrical parts (motors, control box and pendant)
are not damaged and that all connection ports are plugged. DO
NOT allow liquids to enter electrical components.
3.
Disinfect and wash all components. DO NOT submerge the
bed frame or electrical parts.
4.
DO NOT power wash or steam-clean. Use household water
pressure only.
5.
Rinse thoroughly with water (maximum temperature 110°F
or 43°C). DO NOT use solvents, bleach, alcohol, petroleum
products, caustic agents, or highly alkaline or acidic agents to
clean the bed.
6.
Retract each actuator to the minimum stroke length before
washing to avoid degreasing the piston rod.
7.
Dry all components before storing.

7.2 Annual Maintenance Check

Inspect the covering of the bed's control panel and the patient
control panel to assure that the covering is not cracked or
damaged.
Inspect for damaged or loose wiring. Have a qualified technician
replace any frayed or damaged cords.
Inspect for secure grounding.
58
Inspect rail latches. Ensure that all rails engage and lock as
specified.
Inspect rails for wear and damage and replace as required.
Lubricate rail pivot points as needed with white lithium grease.
Inspect bed, rails, Assist Rails or Assist Bars for the presence of
tubing end caps and replace as required.
Tighten, adjust or replace any parts or bolts etc. that are loose
or show signs of wear.
Lubricate all mechanical hinge points, bushings, and surface
contact points with white lithium grease.
Inspect for wear on clevis pins and hitch pins and replace if
worn or missing.
Inspect the rails for noticeable scratches and chips. Use a
matching enamel touch up paint to retouch if necessary. (Refer
to Options for paint specification).

7.3 Quarterly Maintenance Check

Unplug bed from wall outlet and verify battery function, if
equipped. Battery may be built-in or portable.