This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
Invacare® Carroll CS Series™
CS3, CS5, CS7, CS9FX600
en Beds
User Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the Carroll CS Series and is the answer not in the manual?

Questions and answers

Summary of Contents for Invacare Carroll CS Series

  • Page 1 Invacare® Carroll CS Series™ CS3, CS5, CS7, CS9FX600 en Beds User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
  • Page 2 All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™ and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
  • Page 3: Table Of Contents

    6.3 Optional Touch-up Paint ......48 Contents 6.4 Optional Attendant Control Panel Operation ..48 6.5 Optional Pendant HI/LO Lock Out .
  • Page 4: General

    Invacare® Carroll CS Series™ 1.2 Symbols 1 General Signal words are used in this manual and apply to hazards 1.1 Intended Use or unsafe practices which could result in personal injury or property damage. See the information below for definitions The bed is intended to be used as a mattress support of the signal words.
  • Page 5 General Symbols Used on the CS Bed Intertek’s ETL Listed Mark for US and Canada. A product bearing the ETL Listed Mark Protective Conductor Terminal is determined to have met the minimum requirements of prescribed product safety Equipotentiality standards. Battery Charging (Yellow) Warning –...
  • Page 6 Invacare® Carroll CS Series™ Up Control – Raise the corresponding section of the mattress support platform. Crush hazard symbol (based on ANSI Z535) Down Control – Lower the corresponding section of the mattress support platform. Drop/Fall Hazard symbol (based on IEC 61310)
  • Page 7: Overview

    Overview 2 Overview 2.1 Product Labeling On Both Sides of CS9FX600 Bed Head En d Foot End Serial Number Label ETL Label - 120V (CS3, CS5, CS7, CS9FX600) ETL Label - (CS5, CS7) On Both Sides of Bed All labels applied to bed frame except where noted. 1153410-Q...
  • Page 8 Invacare® Carroll CS Series™ This hand cont rol pe nda nt ca n be connecte d to either the lef t or right s ide connect ors loca te d unde rne a th the bed fr ame When s upplied, the pe ndant holster...
  • Page 9: Specifications

    Overview 2.2 Specifications Certifications All models (120 V~ models - CS3, CS5, CS7, CS9FX600 and 230 V~ models - CS5, CS7) conform to UL60601-1 and IEC60601-1. Certified to IEC60601-2-38 and IEC60601-1-2 and CAN/CSA C22.2 No. 601.1. 230 V~ models - CS5 and CS7 are CE marked for Australia and New Zealand.
  • Page 10 Invacare® Carroll CS Series™ Electrical Specifications CS9FX600 120V~ 120V~ or 230V~ 120V~ or 230V~ 120V~ Control box mains voltage 3.5A max 3.5A max Control box input 3.5A max (1.5A max 3.5A max (1.5A max current for 200V~ models) for 230V~ models)
  • Page 11 Overview Safety Features CS9FX600 Protective earth ground Class I Type B Electric shock protection IPX4 Enclosure protection Floor lock not available Optional standard Built-in battery back-up not available optional (standard with Trendelenburg 24V DC 9.2 Amps) Security - Pendant Lock-Out CS9FX600 Pendant connection standard...
  • Page 12 Invacare® Carroll CS Series™ Dimensions and Weight CS9FX600 Length (overall) 76 in and 80 in CS Beds 84¼ in (2140 mm) 80 in CS Bed with 4 in 88¼ in (2242 mm) Extension Length between head and foot ends 76 in CS Bed 78½...
  • Page 13 Overview Dimensions and Weight CS9FX600 Width with assist rails - 39¼ in (997 mm) 36 in standard deck Width with Thinksoft® positioning devices 36 in Standard 41¼ in (1048 mm) 39 in Wide 44¼ in (1124 mm) 42 in Wide 47¼...
  • Page 14: Mattress Specifications

    Invacare® Carroll CS Series™ 2.3 Mattress Specifications Model Number Weight SPS1084 24.0 lbs (10.9 kg) The mattresses in this table have been tested to comply with FDA Entrapment Guidelines when sold new and unused, and microAir Series* 17 – 56 lbs (7.7 – 25.4 kg)
  • Page 15: Safety

    Bed accessories designed by other manufacturers professional, dealer or technical personnel have NOT been tested by Invacare. Use of before attempting to use this equipment. Non-Invacare bed accessories may result in injury – Do not use this product or any available or death.
  • Page 16 Invacare® Carroll CS Series™ DANGER! DANGER! Risk of Injury or Death Risk of Death, Injury or Damage To reduce the risk of electrical shock: To avoid injury or damage when operating – Always unplug the product from the electrical product: outlet before cleaning.
  • Page 17 When transferring Refer to the Bed Rail Entrapment Risk Notification into or out of the bed, ALWAYS lock the casters. Guide at www.invacare.com for additional safety Even with casters properly locked, some information. flooring surfaces (such as tile or wood) will allow the bed to move under some conditions.
  • Page 18 – Always leave bed in lowest height when entrapment: unattended. – The Invacare mattress MUST fit firmly against – For the CS3 bed only: Place the bed in the the bed frame AND bed side rails to prevent lowest position possible without disengaging patient entrapment.
  • Page 19 – Read and understand the instructions or label incompatible with Invacare beds. provided with mattress. – To avoid injury, ensure that only Invacare bed – If you do not understand the instructions or if rails are used with Invacare beds at all times.
  • Page 20 The may become entangled, adjust the height Invacare mattress MUST fit firmly against the of the bed rail (if applicable), or provide bed frame AND bed side rails to prevent alternative means of patient protection.
  • Page 21 Risk of Injury or Damage cut or remove the round grounding prong from To avoid product damage and personal injury: any plug used with or for Invacare products. – Disconnect pendant when not in use and store out of reach of children.
  • Page 22 Invacare® Carroll CS Series™ Weight Limitations Operating Information DANGER! DANGER! Risk of Death, Injury or Damage Risk of Death, Injury or Damage To avoid bed collapse and resulting injury or To avoid serious injury or product damage from damage: tripping, strangulation, entanglement and/or –...
  • Page 23 (the internal pendant or junction boxes. No user serviceable battery is optional for the CS7 and CS9FX600 parts are inside. Contact dealer or Invacare for models and standard on the CS9FX600 with repair/replacement. Trendelenburg). To avoid injury, damage or death –...
  • Page 24: Emi Information

    Invacare® Carroll CS Series™ 3.2 EMI Information The sources of radiated EMI can be broadly classified into three types: CAUTION! 1. Handheld Portable transceivers (transmittersreceivers Risk of Damage with the antenna mounted directly on the transmitting To avoid product damage from electromagnetic unit.
  • Page 25: Electromagnetic Compatibility Information

    Safety 3.3 Electromagnetic Compatibility Information 3.4 Disposal Follow local governing ordinances and recycling plans Portable and mobile RF communications equipment regarding disposal of the product or components normally can affect medical electrical equipment. used in operation. The product does not generate waste or residue in operation.
  • Page 26: Set-Up

    Make sure to have – DO NOT attempt to operate product controls the truck driver sign the copy acknowledging the count. prior to completion of assembly. 4. If shortage occurs, IMMEDIATELY contact Invacare Customer Service department at 1–800–668–2337. CAUTION! Risk of Damage To avoid product damage from misuse: –...
  • Page 27 Set-up 1. Release/cut the straps A that secure the packaged bed B to the shipping pallet C. 2. Remove the straps from the assembly. The bed is packaged so that the wheels (not shown) of the bed are always located on the same side as the latch mechanisms D.
  • Page 28 Invacare® Carroll CS Series™ 5. Cut the untrimmed or green tie holding the head deck 7. Cut the untrimmed or green tie holding the mattress to the frame. keepers on either side of the bed. Fig. 4-2 Fig. 4-4 6. Cut the untrimmed plastic tie holding the knee-foot 8.
  • Page 29: Inspecting

    DANGER! Risk Of Injury Or Damage To avoid injury or damage, inspect the following and contact a qualified technician or Invacare if any of the following issues are present: – Pendant covering is cracked or damaged. – Power cord or plug have cuts or damage.
  • Page 30: Installing The Mattress Keeper Rod

    Invacare® Carroll CS Series™ 1. With the mattress keeper upright, insert the rod ends DANGER! into the appropriate deck holes. Risk Of Injury Or Damage 2. There are two choices for mattress keeper rod insertion. To avoid injury or damage: The 76 inch bed option should use the insertion holes –...
  • Page 31 Set-up 3. Fold the mattress keeper around the bed in the rod. 4. 76 inch Bed Only - With a 7/16 inch socket wrench The mattress keepers should rest flat on the foot deck and 7/16 inch wrench loosen and remove the bed end and the head deck.
  • Page 32 Invacare® Carroll CS Series™ 6. 76 inch Bed Only - Slide the bed end bracket in the 8. 76 inch Bed Only - Tighten the fasteners with two 7/16 mounting channel. Align the hole in the bracket with inch wrenches. Reposition the other three brackets and the channel hole.
  • Page 33: Installing The Head And Foot Ends

    Set-up 4.5 Installing the Head and Foot Ends 9. Position the mattress on the deck. Head and foot ends are normally packaged together All Models Except CS9FX600 either in a cardboard box or on a wooden skid separate from the bed. Bed end model IHBEADJ must be used with bed model CS9FX600.
  • Page 34 Invacare® Carroll CS Series™ 1. Remove the key from the keyhole. 3. Slide the boards down while plate/post screws are in the bracket slots. Bed end not shown. 2. Position Head End next to the head board bracket. Foot end shown.
  • Page 35 Set-up 5. Guide the key through the small bracket hole. The key 6. Turn and lock the key once it is fully inserted. Repeat is attached to the bed bracket with a lanyard. on the other side, then the opposite bed end, securing both mounting points on each board.
  • Page 36: Storing The Bed

    Invacare® Carroll CS Series™ 4.6 Storing the Bed WARNING! Risk of Injury or Damage To avoid product damage and personal injury: – DO NOT place other objects on top of the cartons. WARNING! Risk of Injury or Damage To avoid product damage and personal injury: –...
  • Page 37: Operation

    – After the thermal protection activation period product is not working properly, contact a is complete, the bed may be prone to sudden qualified technician or Invacare for repair. movement. Use caution when using the product after the protection function time...
  • Page 38: Pendant Storage And Location

    Pendant malfunction may cause injury or damage. 2. To lower the entire bed, press the Bed Down button B. – Unplug bed from electrical outlet and contact qualified technician or Invacare for repair if any Head and Foot Sections malfunction is detected.
  • Page 39 Operation 1. Place the pendant holster between the top rail bars. 3. Alternatively, use the pendant clip to secure the pendant to bed linen. 2. Slide the pendant into the holster. 4. The pendant can also be attached to the back of either side of the Head end.
  • Page 40 Invacare® Carroll CS Series™ 5. To move the pendant to the other side of the bed, 7. On the other side of the bed, remove the cap from the disconnect the main power then unscrew the coupling Y cable housing. Replace the cap on the other side of cap on the pendant cable.
  • Page 41: Power Cord Storage

    Operation 5.3 Power Cord Storage 8. Align arrows then insert the pendant cable end into the Y cable housing. Disconnect the main power cable from the wall outlet and store the power cord when the bed is not in use. 9.
  • Page 42: Raise And Lower The Heel Lift Ratchet

    Invacare® Carroll CS Series™ 5.4 Raise and Lower the Heel Lift Ratchet 3. Slowly lower the platform to engage the nearest stop. 1. With knee-foot platform adjusted to the highest position, lift the foot panel. 4. The platform cannot lower past the nearest stop while being adjusted.
  • Page 43: Foot Pad Lock Mechanism

    Operation 5.5 Foot Pad Lock Mechanism 5. To reset the ratchet mechanism, raise the platform to the highest level then lower the platform in one motion. For CS3 Model Only CAUTION! Risk of Damage Moving the bed while the floor lock, caster lock or foot pad is engaged may damage the bed.
  • Page 44: Bed Steer Mechanism

    Invacare® Carroll CS Series™ 5.6 Bed Steer Mechanism 3. When engaged the bed will steer easily. 1. The steering mechanism is on either side of the head end of the bed. 2. With the locking wheel tab facing outward, flip the steer mechanism.
  • Page 45: Optional Accessories

    Patient entrapment from the use of bed side should be replaced. rails may cause injury or death. To avoid patient entrapment: – The Invacare mattress MUST fit firmly against DANGER! the bed frame AND bed side rails to prevent Risk of Death, Injury or Damage Proper patient assessment and monitoring, and patient entrapment.
  • Page 46 For the CS3 bed only: Place the bed in the use a trapeze. lowest position possible without disengaging – Only use the Invacare Heavy Duty Trapeze the foot pad lock mechanism. (1153386) with the CS Bed. – Invacare recommends removing the trapeze –...
  • Page 47 Optional Accessories The listed bed rails and positioning devices are available: WARNING! Risk of Injury or Damage Model Description Risk of FALL when Assist Bar is UNLATCHED. Assist Bar may rotate quickly and drop to the IHCSRLAS CS Assist Rails - Set next locked position causing the patient to lose IHCSRLAS-NG CS No Guard Assist Rail -...
  • Page 48: Other Accessories

    Invacare® Carroll CS Series™ 6.2 Other Accessories 6.4 Optional Attendant Control Panel Operation The listed items are available as optional accessories: The ACP panel is designed to be used alongside the pendant to adjust the height and position of the bed sections. It is...
  • Page 49: Optional Pendant Hi/Lo Lock Out

    Optional Accessories 3. To raise and lower the head and foot sections of the Optional Battery Indicator Light bed simultaneously, press the corresponding arrows C. For beds equipped with an optional battery, the ACP may be This function is optional depending on whether equipped with an indicator light H to show battery charge the pendant also has this function.
  • Page 50: Optional 39 Inch And 42 Inch Wide Decks

    Invacare® Carroll CS Series™ 6.7 Adjustable Width Bed Model CS9FX600 1. To operate the lockout function, either move the HI/LO switch A to OFF or press the lock button A. The following bed options are not compatible with 2. When the lockout function is not activated, the arrow the CS9FX600: 68 inch rail.
  • Page 51 Optional Accessories Opposite side is the same as labeled side. 1153410-Q...
  • Page 52 Invacare® Carroll CS Series™ 36 inch Width 39 inch Width 42 inch Width Side View Side View Side View Top Bed Deck View Top Bed Deck View Top Bed Deck View 1. Loosen the knob A securing the adjustable deck wing 3.
  • Page 53: Adjusting The Head Board Width For Cs9Fx600

    Optional Accessories 5. Tighten the knob to secure the adjustable deck wing. 1. Firmly grasp one of the board wings A. 6. Adjust the head board width. Refer to 6.7.2 Adjusting 2. While holding the board wing, remove the knobs B the Head Board Width for CS9FX600, page 53.
  • Page 54: Maintenance

    – All repair and maintenance on the bed should – Keep the product away from sparks, flame and be performed by a qualified technician. open heat sources. – Contact your dealer or Invacare for repair. DANGER! Risk of Death, Injury or Damage WARNING!
  • Page 55: Annual Maintenance Check

    Maintenance 7.2 Annual Maintenance Check • Inspect the covering of the bed’s control panel and the patient control panel to assure that the covering is not cracked or damaged. • Inspect for damaged or loose wiring. Have a qualified 1. Unplug product before washing. technician replace any frayed or damaged cords.
  • Page 56: Troubleshooting

    Invacare® Carroll CS Series™ 8 Troubleshooting 8.1 Troubleshooting Effect Possible cause Verification Corrective action Bed does not stay in place Floor lock not engaged. Green plunger on floor lock Activate floor lock - green post is visible. plunger should NOT be visible.
  • Page 57 Troubleshooting Effect Possible cause Verification Corrective action Actuator not working. For non-openbus (CS3, CS5) Rotate the lockout dial to HI/LO lock-out is on. the “unlocked” position - A Lockout box (a square black box) with a toggle (horizontal), or flip switch to switch.
  • Page 58 Invacare® Carroll CS Series™ Effect Possible cause Verification Corrective action Actuator not working. Faulty actuator. Disconnect power supply, Contact Invacare Continuing swap cable from unresponsive Care Service Department 1-800-668-2337. actuator with a functional actuator. Reconnect power and test functions. Faulty...
  • Page 59: Warranty

    MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the original purchaser who purchases this product when new and unused from Invacare Continuing Care, Inc (ICCI) or a dealer. This warranty is not extended to any other person or entity and is not transferable or assignable to any subsequent purchaser or owner. Coverage under this warranty will end upon any such subsequent sale or other transfer of title to any other person.
  • Page 60 Invacare Continuing Care, Inc. A subsidiary of Invacare Corporation. Designed by Invacare Corporation Invacare Continuing Care, Inc. www.invacare-cc.com Manufactured in Mexico for Invacare Corporation Canada Invacare Australia Pty. Ltd. One Invacare Way 570 Matheson Blvd E Unit 8 1 Lenton Place...

This manual is also suitable for:

Carroll cs3Carroll cs5Carroll cs7Carroll cs9fx600