Page 1
Oxygen Analyzer DF-310 E OPERATOR MANUAL...
Page 2
No part of this publication may be reproduced, stored in a retrieval system or transmitted in any form, or by any means including electronic, mechanical, photocopying, recording or otherwise without prior written permission of Servomex Corporation. Stablex, Bi-Strata and are trademarks of Servomex Corporation. VCR is a registered trademark of the Cajon Company.
Page 3
For more information or if you have questions, please do not hesitate to go to our website at Servomex.com, or contact your local Servomex Business Center as found on the back cover of this manual.
Page 4
(see Shipping in the Service section). Installation and Maintenance The DF-310E Process Oxygen Analyzer will provide years of accurate and dependable service if it is set up, operated and maintained properly. It is essential to make a careful and complete installation as outlined in the Installation and Setup section of this manual.
There are a number of warnings and cautions that must be observed to avoid damage to the analyzer as well to insure the safety of its users. The analyzer must be operated in a manner specified in this manual. Servomex cannot be responsible for direct or consequential damages that result from installing or operating the analyzer in a manner not described in this manual.
Page 12
For best performance at initial start or anytime the electrolyte is changed, it is important to allow the sensor to sit with electrolyte in it for 60 minutes before the gas is allowed to flow through the sensor. DF-310E Cautions...
0 - 1 X.XXX% 1 - 10 XX.XX H00P25 0-25% 0 - 2.5 X.XX% 2.5 - 25 XX.X *Scale A applies to High Resolution models only. Scale B extends down to 0 ppm or 0% on Standard Resolution models. Specifications DF-310E...
Page 14
Contacts failsafe to alarm condition upon loss of power. Not designed to switch AC power. POWER REQUIREMENTS 100 – 240 VAC (auto-switching), 1.3A, 50/60 Hz or 24 VDC (–2/+4VDC), 1A, 25 Watts; Optional Sample Pump 6W additional EMI SENSITIVITY DF-310E Specifications...
Page 15
S, Cl , NO , SO , HCl, etc. GAS SAMPLE MOISTURE CONTENT No limit (avoid condensation) OIL/SOLVENT MIST <0.03 mg/L Standard limit >0.03 mg/L Use filter SOLID PARTICLES <0.01 mg/L Standard limit, Use filter if >0.01 mg/L Specifications DF-310E...
The screens shown in this manual have values that may not match the actual values displayed during your setup. ROTOMETER VALVE (optional) POWER SWITCH ELECTROLYTE RESERVOIR WITH MIN-MAX LINES OXYGEN SENSOR SENSOR MOUNTING THUMB SCREWS Figure 2: Major Internal Components Installation and Setup DF-310E...
The sensor is shipped dry and must be charged with electrolyte before it is operated. NOTE Use only Hummingbird E-lectrolyte Blue for the DF-310E Oxygen Analyzer. Failure to do so will void warranty. Install one bottle. NOTE Do not apply power before adding electrolyte and thoroughly purging sample line.
Over-pressurizing the sensor can result in permanent damage to the sensor. Limit the backpressure to the analyzer to ±1 psig. NOTE Allow gas with very little oxygen to flow through the analyzer for approximately 60 minutes before powering up. Installation and Setup DF-310E...
AC Powered Units - Open the front door, locate the power switch and turn it on. See Figure 2. DC Powered Units - Turn on the remote 24 VDC power source, open the front door, locate the power switch and turn it on. See Figure 2. DF-310E Installation and Setup...
After power up, the analyzer will undergo a series of Diagnostic Procedures. After approximately 5 seconds, the Servomex Corporation logo is displayed. After 30 seconds, a WAIT message appears for 1.5 minutes. A display then appears that is similar to Figure 5 (values shown are only representative).
PUMP -, + connections on the rear panel connector J4. The contacts can be used to send a signal indicating the status of the internal pump or to control an external, Servomex supplied pump that is powered from a separate source.
6 hours Table 1: Battery Operation Time NOTE Use only Servomex P/N 16337070 when replacing the NiMH battery pack. In the event that the NiMH Battery Option is installed in an analyzer that also has the Case Purge Option, the NiMH Battery system must be disabled.
Note: If the analyzer is equipped with a VCR welded sample inlet connection, there is a .010 inch orifice at the inlet to the sensor which requires 15-20 psig of pressure. NOTE: For additional information on the proper purging of regulators after installation see page 37. Options DF-310E...
2 connected to the Analyzer's sample inlet fitting. For mist coalescing and collection for draining, plumb the filter with port 1 connected to the Analyzer's sample inlet fitting. NOTE: For additional information on the proper purging regulators after installation see page 37 Options DF-310E...
The lock is supplied with two keys. If the analyzer is operating, the key lock does not prevent adjustments from the front panel. Password Protection, described in the User Interface section DF-310E Options...
Use of this function provides information than could be interpreted remotely as an alarm or non-standard condition. See page 66 for additional information on the Sensor Off function. 5.11Relays Up to four optional form C (SPDT) relays (contact closures) are available to Options DF-310E...
The optional expanded range scale allows the analog output scaling to be automatically expanded to a larger value when the primary scaling range is exceeded. See page 72 for additional information. 5.14Panel Mount A panel mount option is available. See Figure 11 and Figure 12 below for details. DF-310E Options...
Following are the dimensions for the hole cutout and mounting screws. The connecting cable must be shielded with the ground attached only to the stud on the rear of the analyzer and wired as shown in Figure 16. DF-310E Options...
5.18Case Purge NOTE: The case purge option is available on AC powered analyzers only. The DF-310E analyzer can be equipped with an inert gas (nitrogen) purge system. The purge system provides improved protection against an explosion hazard by purging the enclosure to a concentration level below the lower explosive limit.
Table 2: Maximum Allowable Acid Gas Limits Contact the local Servomex Business Center for recommendations on using the STAB-EL sensor on acid gases other than those listed above. The STAB-EL limits shown in the table represent rough guidelines for continuous exposure.
Servomex offers a broad range of scrubbers for applications in severe environments. Standard scrubber columns are available in various sizes, and in single or dual bed configurations.
Table 3: Gas Scale Factors 6.2.1.1 Disclaimer The method used to correct the calibration of the DF-310E Oxygen Analyzer for measurement in non-nitrogen background gases is derived from a well- known theoretical mass transfer equation. This equation accounts for the change in oxygen diffusion rates through different gases.
(1.0 - 2.0 scfh) no regulator is required. A flow change of ±1.0 scfh may result in a small change to the oxygen reading. DF-310E Sample Gas Preparation and Delivery...
If the analyzer is not vented to atmosphere, the sensor pressure is influenced by the conditions downstream of the analyzer. A recalibration under your operating conditions may be desirable to remain within the stated accuracy Sample Gas Preparation and Delivery DF-310E...
6.4 Sample Gas Compatibility There are a wide range of considerations in determining the gas sample compatibility of the Process Oxygen Analyzer. Servomex attempts to identify all pertinent application details prior to quoting and order processing. All non-typical applications concerning gas sample compatibility must be reviewed by our in-house Application Engineers.
O), however, do not affect the electrode or sealing materials, or interfere with the O reduction/oxidation reactions. Contact the local Servomex Business Center for recommendations on a specific application. 6.4.3 Reactivity with KOH Electrolyte Many process sample streams contain various concentrations of acid gases.
33 for more detail. Contact the local Servomex Business Center for recommendations on using the STAB-EL sensor on acid gases other than those listed. 6.4.6 Sample Gas Temperature Gas temperature should not exceed 50 °C (122° F), nor should it fall below 0°...
For systems where the gas sample is not vented to atmosphere, the analyzer outlet should remain connected in the same manner during calibration, if possible. This ensures that downstream pressure effects on the sensor are the Sample Gas Preparation and Delivery DF-310E...
(factory calibration gas) versus the diffusion rate in the user’s calibration gas. The Sample Gas Preparation and Delivery section discusses the proper setting of the gas scale factor option during calibration as well as during process gas measurement. DF-310E Sample Gas Preparation and Delivery...
When connecting the Process Oxygen Analyzer to a computer via an RS-232 or RS-485 communication cable, a Ferrite Sleeve is required around the cable in a single-turn configuration. It is recommended that the proper Servomex cable be used for this purpose. 7.2 Relay Ports Connections to four optional form C (SPDT) relays (contact closures) are provided on the rear of the analyzer at connector J1 and J2.
2. Label and remove all connections from the rear of the analyzer. 3. Open the door and disconnect the sensor and display cables. See page 13. 4. Remove the two screws from the rear of the unit. See page 18. Connecting to External Devices DF-310E...
12. With a DVM, confirm that the analog output voltage is proper. If it needs to be adjusted slightly, use the potentiometer located third from the top on the front of the circuit board, above the Servomex symbol. Full Scale Output Voltage Jumper Number 5.0 VDC...
If equipped, the oxygen sensor can be turned on and off remotely through the pins labeled EXT 1 or EXT 2 on the J6 connector. If equipped, the EXT Functions screen, see page 91, will indicate to what set of EXT contacts this option is assigned, either 1 or 2. Connecting to External Devices DF-310E...
The pump cable should be of sufficient size for the required run (see Table 10 below) and should not share the same conduit as the sensor cable. See the wiring diagram in Figure 30. DF-310E Connecting to External Devices...
If factory configured, a switch closure rated at 1A/30VDC can be supplied between the PUMP -, + connections on the rear panel connector J4. The contacts can be used to control a Servomex supplied pump that is powered from a separate 12 VDC, .3 A source. Control of the pump would be accomplished in the same manner as a standard pump.
A 0.010 inch orifice is included with 0-50ppm sensors and must be installed at the sensor inlet in order to control gas flow. Contact Servomex for additional information on remote sensor installations. 7.5.1 Sensor on Remote Bracket with Optional Pump...
In the event that the enclosure temperature must be adjusted, follow the steps below. 1) Obtain a temperature measurement device capable of measuring the desired operating temperature to an accuracy of +/- 2 degrees F. Connecting to External Devices DF-310E...
Page 61
(not in an alarm condition). If the purge switch activates, confirm that the enclosure is “tight” and increase the flow rate as necessary. 11. The analyzer may now be turned on. Connecting to External Devices DF-310E...
8 User Interface 8.1 The Data Display Screen When the DF-310E Process Oxygen Analyzer is powered up, it goes through a series of internal diagnostic tests which take about five seconds. After the tests, the Servomex logo appears for ten seconds. The display will then show the Data Display Screen as shown in Figure 32.
Page 64
The alarm number will clear only after the alarm condition is over. In the case of simultaneous alarms, each will alternately overwrite the display. Successive presses of ESC (as the overwrite is displayed) are necessary to clear the overwrite and DF-310E User Interface...
Page 65
(> 10 seconds), such as an internal electrical zero calibration. “Memory Error!” indicates that the instrument has failed the boot-up memory test. The letters “CHG”, “BAT” and “LOW” may appear vertically on the right side of the display on units equipped with the NiMH battery backup option. User Interface DF-310E...
Each level in the Main Menu allows the user to access options for setting and testing instrument parameters. Ellipsis (...) after an entry indicates that additional screens follow. User Interface DF-310E...
"ABANDON CHANGES?, FOR YES". All parameter changes will be lost if the key is pressed. Select the UPDATE & QUIT choice using to save the changes and automatically return to the previous menu. DF-310E User Interface...
(downstream) valve should be fully opened and the upstream flow control valve used for flow control. FAILURE TO FOLLOW THESE INSTRUCTIONS MAY CAUSE THE SENSOR TO EXPERIENCE OVER OR UNDER PRESSURE WHICH MAY CAUSE PERMANENT DAMAGE. User Interface DF-310E...
“Analyzer Trouble” relay such that an alarm condition will cause the relay to break the current loop. This method allows a computerized system to be configured to detect an analyzer fault condition whenever the 4-20 mADC signal is below 4 mA. DF-310E User Interface...
Figure 36: Setup Menu Each entry in Figure 36 leads to a sub-menu. To select the desired sub-menu, use the keys to place the asterisk next to it, then press . A new display will be shown as indicated below. User Interface DF-310E...
On (USED) or Off (NU). When an unused alarm (NU) is accessed, the display will appear as shown in Figure 39. (Oxygen ALARM 1 is used in the example shown in Figure 40.) To indicate that the alarm is DF-310E User Interface...
8.4.1.3 Low Flow Alarm The FLOW alarm is used to indicate a low flow condition in the sample stream. The optional low flow switch will trip if the gas flow rate drops below the value listed in Table 13. User Interface DF-310E...
This method allows a computerized system to be configured to detect an analyzer fault condition whenever the 4-20 mADC signal is below 4 mA. After accessing the Outputs on the Setup Menu, Figure 36, the display will be as shown in Figure 41. DF-310E User Interface...
0-100 ppm 100 % of Scale 50-85 ppm 35 % of Scale Table 14: Output Scaling on Standard Resolution Analyzer If an invalid Zero to FS window is entered the following error message will be briefly displayed. User Interface DF-310E...
The optional expanded range scale function allows the analog output scaling to be automatically expanded to a larger value when the primary scaling range is exceeded. For example, in the display shown in Figure 41, the analog outputs (0-10 VDC and 4-20 mA) are scaled over the 0 - DF-310E User Interface...
This port operates with an 8 bit, no parity, one stop bit setting. No hardware or software handshaking is used. See the Section on Connecting to External Devices on page 43 for more information. User Interface DF-310E...
Refer to the section on Calibration on page 79. 8.4.5 Display Setup Access to the controls related to the backlight, brightness and contrast of the display are gained through the display setup menu. See Figure 44 below. DF-310E User Interface...
Figure 45: Clock Setup Screen 8.5 The Password Menu The DF-310E Process Oxygen Analyzer may include optional password protection which can be used to limit access to the Control Menu, the Set-Up Menu, and the Diagnostics Menu. Note: When the Password entry is selected from the Main Menu, a DISABLING ALARMS message appears which notifies the user that the alarms have been temporarily disabled.
Analyzer will not display it again. If the master password is misplaced, contact the local Servomex Business Center. The master password and operator password can be changed as desired after the present master password has been entered.
– the reminder frequency is pre-set at the factory to eight weeks. 8.6.1.3 Assign the “Reminder” Relay The Reminder flag can be assigned to any of the available relays. See the section on relay assignment in the Oxygen Alarm section on page 68. DF-310E User Interface...
Alternately, the GSF factor can be entered manually. The software in the Analyzer supports the following gases in the GSF calculation: Ammonia Argon Butane Carbon Monoxide Ethane Ethylene Helium Hexane Hydrogen Methane Nitrogen Propylene Table 16: GSF Corrections User Interface DF-310E...
Note: Scrolling down the gas list from Figure 51 to Figure 52 will displace one line at a time. Because these figures are presented from the top and from the bottom of the gas list, H2 (Hydrogen) and NH3 (Ammonia) appear to be missing. DF-310E User Interface...
Page 85
The GSF for the gas used to calibrate the system may be different from that used during analysis. If the GSF is changed to reflect the composition of the calibrating gas, be sure to reset the GSF before analyzing samples. User Interface DF-310E...
Page 86
Disclaimer The method used to correct the calibration of the Servomex Oxygen Analyzer for measurement in non-nitrogen background gases is derived from a well known theoretical mass transfer equation. This equation accounts for the change in oxygen diffusion rates through different gases.
5.0 psig or less. The upstream flow control valve is used to set the flow at 1.0 (scfh). If the normal process sample is being supplied to the Analyzer under moderate vacuum User Interface DF-310E...
It may take several minutes before convergence occurs. During convergence, the Analyzer is verifying stability of the reading before accepting the data. After convergence two short beeps will be heard. The Analyzer's electronics can be updated to the new calibration information by selecting Update and Quit. DF-310E User Interface...
8.6.2.4 New Sensor The New Sensor entry is used after a new sensor is field installed. New sensors are supplied with calibration information. The procedure for installing a new sensor is described in instructions supplied with it. User Interface DF-310E...
See Section 8.1.1 for details on normal operating conditions. If operating outside normal operating conditions contact Servomex for an application specific recommendation on checking the zero of the instrument in the field. Should Servomex recommend checking the zero baseline calibration of the Analyzer, the DF-310E User Interface...
New Sensor zero calibration if the sensor has been field replaced. If the Reset Orig Zero entry is selected, the display will ask Erase?... FOR YES. Press to use the factory set calibration. The Reset Orig Zero line of Figure 57 will disappear. It is User Interface DF-310E...
Figure 59: Zero Cal Screen Pressing the ESC key at any time aborts the process and returns the user to the SENSOR ZERO DF-310E User Interface...
10% percent steps of full scale from 0% to 100%. After setting the % FS Level, press . The analog output response should match the %FS Level value that was entered. For example, if 80% is entered for the %FS value on a 0-10 User Interface DF-310E...
Next, the microprocessor internal memory (IRAM) is tested, followed by the system “external” random access memory (XRAM). As each portion of the memory is successfully tested an OK will appear at the end of the line. If any memory test fails, DF-310E User Interface...
After the test has been completed, the display will return to the Diagnostics Menu, Figure 56. Pressing ESC will abort the screen test. If an error message appears, or a pixel is inactive, contact the local Servomex Business Center.
For process applications containing more significant quantities of acid gases or particulate, or where liquids may be encountered, contact Servomex for a recommendation on calibration verification for your specific case. For Analyzers used in clean gas applications, and operated under Normal Operating Conditions there is no need for zero calibration checks in the field.
Operation at elevated temperatures and/or with sample gases at very low dew points will increase the frequency of replenishing the electrolyte. The Oxygen Analyzer is equipped with an Electrolyte Condition alarm to indicate that the DF-310E Troubleshooting and Calibration...
Be sure to cap the bottle immediately after use. In an emergency, distilled water can be used as an alternative, however this is not recommend over an extended period. Do not add electrolyte solution to restore the electrolyte level. Do not overfill. Troubleshooting and Calibration DF-310E...
Analyzer from the sample delivery system, and to allow the system to pressurize. Apply Snoop® or another type of liquid leak detector to all of the fittings on the system. Any fitting that shows bubbles should be tightened or replaced. Troubleshooting and Calibration DF-310E...
- Over Range or TEMP OVER RANGE L M N Z - NOVRAM Failure - Uncalibrated Span reading is unacceptably high (>50% high) R C J Z Span reading is unacceptably low (>50% low) R J E Z DF-310E Troubleshooting and Calibration...
Page 103
Voltage levels between any other combination of wires should be less than 0.10 VDC. If there is any deviation from these values, contact the local Servomex Business Center Change the electrolyte. Use only electrolyte supplied by Servomex. Other types of electrolyte can damage the sensor and will void the warranty. Always rinse and drain the cell with distilled water at least three times before refilling the sensor with fresh electrolyte.
100 ppm, and certainly standards below 10 ppm, should be prepared in aluminum cylinders. Contact the local Servomex Business Center For faster service, have the instrument serial number and model number in hand before calling. Always be certain to drain the sensor of electrolyte before returning it to the factory for repair.
Page 105
III, Type T fuse is rated at 2.5A. There are two fuses that are located in the AC input connector located behind the cover on the rear of the cabinet. Refer to the spare parts list on page 96 for Servomex replacement part numbers. 9.4.3.2 DC Power Fuse If configured for 24 VDC operation, the 1.0A type TE-5 fuse is located on the under side of the...
Note: If you are returning the analyzer to the factory, first call Servomex to obtain a Return Material Authorization number (see complete details below), then proceed as follows: 1. Turn off and disconnect the power source from the analyzer.
10 Theory of Operation 10.1The Oxygen Sensor The Servomex Coulometric Sensor uses an ambient temperature oxygen reaction that is non-depleting. The cell produces a current flow that is determined by the number of oxygen molecules that are reduced at the cathode. The sensor reaction is driven by 1.3 Volts applied across the electrodes.
10.2 The Electrolyte Conditioning System The Process Oxygen Analyzer is equipped with Servomex's patented electrolyte conditioning system and is composed of two specialized electrode pairs. The patented secondary electrode pair protects the sensing electrodes from the deleterious effects of trace impurities inevitably found in the electrolyte.
Limit the backpressure to the analyzer to ±1 psig. Be sure the downstream isolation valve (if so equipped) is toggled open before gas flow is started. CAUTION DO NOT SHIP THE ANALYZER WITH ELECTROLYTE – THOROUGHLY DRAIN AND RINSE SENSOR BEFORE SHIPPING Safety DF-310E...
Page 112
If the interference disappears, try one or more of the following methods to correct the problem: Reorient the receiving antenna. Move the instrument with respect to the receiver. Place the analyzer and receiver on different AC circuits. DF-310E Safety...
1. IDENTIFICATION OF THE SUBSTANCE Trade Name Electrolyte Solution, E-lectrolyte Gold, E-lectrolyte Blue, E- lectrolyte Black, DF-E05, DF-E06, DF-E07, DF-E09 Manufacturer Servomex Corp., 4 Constitution Way, Woburn, MA 01801-1087, USA, Tel + 1-781-935-4600 USA: 1-800-424-9300 Emergency Contact International: 1-813-979-0626 (collect)
Page 114
Airborne Exposure This material is supplied as an aqueous solution and will not be present in the atmosphere in normal use. Potassium Hydroxide Exposure Limits DF-310E Safety...
Page 115
Will degrade by reaction with carbon dioxide from the atmosphere to produce a Degradability non-hazardous product. Accumulation Information not available. No long-term effects expected due to degradation. The Ecotoxicity preparation is already in dilute solution and adverse aquatic effects are not expected Safety DF-310E...
Page 116
Does not contain any Class 1 Ozone depletors Does not contain any Class 2 Ozone depletors Clean Water Act CAS# 1310-58-3 Listed as a Hazardous Substance under the CWA None of the chemicals are listed as Priority Pollutants under the CWA DF-310E Safety...
Page 117
Liability is expressly disclaimed for loss or injury arising out of use of this information or the use of any materials designated. Users should make their own investigation to determine the suitability of the information for their particular purpose. Safety DF-310E...
11.2 Replenishment Solution MSDS MATERIAL SAFETY DATA SHEET 1. IDENTIFICATION OF THE SUBSTANCE Trade Name Replenishment Solution, RS-A Manufacturer Servomex Corp., 4 Constitution Way, Woburn, MA 01801-1087, USA, Tel + 1-781-935-4600 USA: 1-800-424-9300 Emergency Contact International: 1-813-979-0626 (collect) Supplier and contact in UK (for use in the UK only) 2.
Page 120
17.5 mm Hg @ 20 Vapor Pressure 10. Stability & Reactivity Stable Chemical Stability Strong reducing agents, acid chlorides, phosphorus trichloride, Conditions/Materials to Avoid phosphorus pentachloride, phosphorus oxychloride. Hazardous Decomposition Products Not applicable. Hazardous Polymerization Has not been reported DF-310E Safety...
Page 121
Liability is expressly disclaimed for loss or injury arising out of use of this information or the use of any materials designated. Users should make their own investigation to determine the suitability of the information for their particular purpose. Safety DF-310E...
12 Warranty Servomex Corporation warrants each instrument manufactured by them to be free from defects in material and workmanship at the F.O.B. point specified in the order, its liability under this warranty being limited to repairing or replacing, at the Seller's option, items which are returned to it prepaid within one year from delivery to the carrier and found, to the Seller's satisfaction, to have been so defective.
Page 127
Storage Temperature, 10 Sample Gas, 79 Sample Gas Preparation and Delivery, 33 Background Gas Effects, 36 Backpressure Effects, 38 TEMP OVER RANGE, 61, 82 Flow Rate Effects Temperature, 10, 69, 100 Leakage Checks, 36 Temperature Alarm, 69 Sensor Performance, 36 Index DF-310E...
Page 128
Weight, 11 UNCALIBRATED, 61, 98 ZERO CAL, 88 Zero Calibration Frequency UNDER RANGE, 61 Update And Quit, 74 Clean Gas Applications, 93 Upstream Valve, 65 ZERO REF, 87 Z-Purge, 56 UR, 61 User Interface, 59 Data Display Screen, 59 DF-310E Index...
Need help?
Do you have a question about the DF-310E and is the answer not in the manual?
Questions and answers