1 Shimadzu LC-10/20 System
Thermostats: CTO- 10A, CTO- 10AC, CTO- 10Avp, CTO- 10ACvp, CTO-
10ASvp, CTO-20A*, CTO-20AC*
The list continues to extend, for up to date list see the website
www.dataapex.com. Some devices also require a firmware of particular
minimal version for proper communication with Clarity (for more details
see the chapter Minimal version of device firmware required on pg 4).
* Devices marked with the asterisk will show in the
dialog only after the CBM-20A (Lite) controller has been selected.
Caution:
If any device from the LC- 20 (Prominence) line is to be used with the
LC- 10 line Controller, it is necessary to set it to the LC- 10 emulation
mode. This is done (for example on the LC-20AD pump) on the devices
display by pressing the VP key until the CALIBRATION screen appears.
Then press func key, enter the password and keep pressing the func
key until OP MODE screen shows. There, select the correct type of the
controller installed. Later in the
necessary to set the LC-20AD pump as LC-10AD when selecting the type
of the pump in the
° The autosamplers SIL-20AHT and SIL-20ACHT marked with the circle
do not possess their specific option in the
Autosampler dialog. In case of SIL- 20AHT autosampler is present it
should be configured through SIL- 20A option. If the SIL- 20ACHT
autosampler is present it is necessary to configure it through SIL-20AC
option.
** RID-20A, RF-20A and RF-20AXS detectors must be always switched to
the LC- 10 emulation mode in order to be controlled (using MIMIC
libraries) in Clarity.
*** SPD-M10Avp PDA detector is not supported on 64-bit Windows.
RF-10AXL and RF-20AXS (RF-20AXS with minimum firmware version
2.02) fluorescence detectors can be configured as a stand-alone control
module developed through Knauer company. This control module for RF-
10AXL
communication via special RS232 cable (p/n SK 11) . If the RF- 20A
detector was purchased through Knauer company (with special oem
firmware version) it can be connected to Clarity with above mentioned
special RS232 serial cable and controlled with above mentioned specific
control module.
Clarity does not control following controllers: CBM-10A, SCL-10A.
Shimadzu LC-10/20 System
and
RF- 20AXS
System Configuration
setup.
Shimadzu LC-10/20 System
fluorescence
detectors
- 2 -
Clarity Controls
System Configuration
dialog it is
supports
serial