PIE PIECAL 211 Operating Instructions Manual

PIE PIECAL 211 Operating Instructions Manual

Automated universal rtd calibrator

Advertisement

• Easy to use
With the PIECAL 211 you can check & calibrate all
your RTD instruments and measure RTD Sensors.
Automatic indication of connections on the display for
simple hookups.
• Take it into the shop, plant or field
Carry it without worry - it comes protected with a
rubber boot and rugged, low profile switch. Easy to
operate even in the dark areas of the plant with the
backlit display.
• Calibrate directly in temperature (°C & °F)
Stop carrying around a decade box and RTD
resistance tables. The PIECAL 211 works with the
RTDs you use including Platinum 100 (alpha = 3850,
3902, 3926) & 1000 (alpha = 3850, 3750) Ohm,
Copper 10 & 50 Ohm, Nickel 100 and 120 Ohm.
Easily set any value quickly to within 0.1° with the
adjustable digital potentiometer "DIAL" plus store any
three temperatures for instant recall with the
EZ-CHECK™ switch.
• Calibrate quickly with automatic output stepping
Choose between 2, 3, 5, 11 and 21 steps to
automatically increment the output in 100%, 50%,
25%, 10% or 5% of span. Select the step time to
match your system from 5, 6, 7, 8, 10, 15, 20, 25, 30
and 60 seconds.
• Compatible with all process instruments
Connect directly to the RTD inputs of smart
transmitters, PLCs, DCS and multichannel recorders and verify their outputs or
displays. Works with older instruments with fixed excitation currents and newer
multichannel instruments that switch the excitation current between input channels.
• Measure RTD Sensors
Trouble shoot sensor connections and find broken wires with patented technology. Connect your
two, three or four wire RTDs and the PIECAL 211 automatically detects the connections and
measures the RTD in degrees C or F.
82 East Main Street Suite 3.14 • Webster, NY 14580 Tel: 585.872.9350 • Fax: 585.872.2638 • sales@piecal.com • www.piecal.com
uality Products Online at:
Automated Universal RTD Calibrator
Automated Universal RTD Calibrator
Product Description
Practical Instrument Electronics
Copyright © 2011 All rights reserved • 211-9002 - Rev D 13 July 2011
GlobalTestSupply
www.
PIECAL 211
PIECAL 211
Operating Instructions
Operating Instructions
.com
sales@GlobalTestSupp
Page 1

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PIECAL 211 and is the answer not in the manual?

Questions and answers

Summary of Contents for PIE PIECAL 211

  • Page 1 • Measure RTD Sensors Trouble shoot sensor connections and find broken wires with patented technology. Connect your two, three or four wire RTDs and the PIECAL 211 automatically detects the connections and measures the RTD in degrees C or F.
  • Page 2: Basic Operation

    EZ-CHECK™ SWITCH SOURCE: Instantly output two preset RTD temperatures by moving the EZ-CHECK™ switch to Double click the knob to get into the PIECAL 211 the “LO” position or “HI” position. For fast three Configuration Mode. Use configuration to select point checks select the “DIAL”...
  • Page 3 Configuration EXIT MENU - exits this menu immediately and Configure the Calibrator saves any changes. Menu will automatically exit after 15 seconds of inactivity (countdown timer is Move POWER SWITCH to “SOURCE” or displayed). “READ”. TEMP UNITS - pressing the knob will toggle between °C and °F.
  • Page 4 2) Select “READ” with slide switch 3) Place the EZ-CHECK switch into the READ position. 4) Connect the PIECAL 211 to the device using 2, 3 or 4 wires. The PIECAL 211 measures the temperature signal and constantly updates the display with the current temperature reading.
  • Page 5: Automatic Stepping

    Storing EZ-CHECK Outputs STORING HI and LO EZ-CHECK Outputs Choose this function to provide a simulated RTD signal into controllers, temperature transmitters, indicators or any other input device that measure thermocouple sensors. 1) Store your high (SPAN) output temperature by moving the EZ-CHECK switch to the HI position and turn the EZ-Dial knob until the desired temperature is on the display.
  • Page 6 Ranges & Accuracies Alpha Degrees C Accuracy Degrees F Accuracy Type Range °C Range °F Pt 100 Ohm (DIN/IEC/JIS 1989) 1.3850 -200.0 to 200.0 ±0.3° -328.0 to 392.0 ±0.5° Based on ITS-90 (0.00385) 200.0 to 600.0 ±0.4° 392.0 to 1112.0 ±0.7°...
  • Page 7 PIECAL 211 Specifications (Unless otherwise indicated all specifications are rated from a nominal 23 °C, 70 % RH for 1 year from calibration) General Accuracy ±(0.025% of Reading + 0.05 Ohms) Temperature Drift ± 0.01% of span outside of 23°C ±10 °C (73°C ±18 °F) Operating Temperature Range -25 to 60 °C (-10 to 140 °F)
  • Page 8: Additional Information

    The equipment will be repaired, replaced or adjusted at our option. The liability of Practical Instrument Electronics (PIE) is restricted to that given under our warranty. No responsibility is accepted for damage, loss or other expense incurred through sale or use of our equipment.