500V r.m.s. a.c. test to earth. This must be taken into consideration at installation. About This Document This manual contains instructions for installation and operation of the Rosemount 396R and 396RVP pH/ORP Sensors. The following list provides concerning all revisions of this document.
Section 1: Description and Specifications Features and Applications Rosemount 396R and 396RVP Sensors are specifically designed for improved life in harsh, dirty applications where a separate sample stream is difficult to provide and greater insertion depths are required. These sensors are designed for use with a 1-1/4 in. or 1-1/2 in. ball valve for hot tap installation.
Range Hemi Bulb Flat Bulb 0-2 pH 2-12 pH 12-13 pH 13-14 pH Table 1-2: Rosemount 396R sensor specifications Measured Range pH range 0 to 14 pH ORP range -1500 mV to 1500 mV Maximum Pressure at Retraction or Insertion...
Instruction Manual Description and Specifications LIQ-MAN-396R-396RVP March 2017 Table 1-3: Rosemount 396RVP sensor specifications Measured Range pH range 0 to 14 pH ORP range -1500 mV to 1500 mV Maximum Pressure at Retraction or Insertion Code 21: 64 psig (542 kPa abs)
Description and Specifications Instruction Manual March 2017 LIQ-MAN-396R-396RVP Product Certifications Please see online certificates for further details. IECEx Sensors without preamp (pH and ORP) – Ex ia IIC T4 Ga (-20°C ≤ Ta ≤ +60°C) Sensors with SMART preamp (pH only) – Ex ia IIC T4 Ga (-20°C ≤ Ta ≤ +60°C)
For use with legacy transmitter model 1181. For use with Rosemount model 1056, 1057, 1066, 5081, and 56 transmitters. Must be selected with option 70. Also suitable for use with Rosemount model 1056, 1057, 1066, 5081, and 56 transmitters. Must be selected with option 55.
Page 12
Description and Specifications Instruction Manual March 2017 LIQ-MAN-396R-396RVP Description and Specifications...
March 2017 Section 2: Installation First Time Installation For first time Rosemount 396R installations, using the following guide is recommended: Retractable Mounting Choose one (required for all first time installations): PN 23166-00, 1 in. x 1 in. NPT process connector, 316 SST PN 23166-01, 1 in.
Mechanical Installation The Rosemount 396R Sensor may be installed through a weldalet or in a pipe tee or “Y”, as shown in Figure 2-1, when used with a ball valve. Insert the end of the sensor to a depth sufficient to ensure that the glass bulb is continuously wetted by the process fluid.
Page 15
2.3.2 Installation Without Ball Valve The Rosemount 396R Sensor may be installed through a weldalet or pipe tee or “Y” when used with a process connector (PN 23166-00 or 23166-01). The sensor should be installed within 80° of vertical, with the electrode facing down.
Page 16
Installation Instruction Manual March 2017 LIQ-MAN-396R-396RVP Figure 2-1: Typical Mounting Configurations for Rosemount 396R Figure 2-2: Typical Mounting Configurations for Rosemount 396RVP Calibration and Maintenance...
Page 17
Instruction Manual Installation LIQ-MAN-396R-396RVP March 2017 Figure 2-3: Dimensional Drawing — Rosemount 396R with Optional Ball Valve PN 23765-00 Note: Add five (5) inches to dimension A if mounting a sensor head junction box onto the sensor. Calibration and Maintenance...
Page 18
Installation Instruction Manual March 2017 LIQ-MAN-396R-396RVP Figure 2-4: Dimensional Drawing — Rosemount 396R with Optional Ball Valve PN 23240-00 Note: Add five (5) inches to dimension A if mounting a sensor head junction box onto the sensor. Installation...
Page 19
Instruction Manual Installation LIQ-MAN-396R-396RVP March 2017 Figure 2-5: Dimensional Warning Label for Rosemount 396R Hemi Bulb Sensors and Sensor Note: Pressure rating for flat glass sensors is 100-790 kPa (0-100 psig). Installation...
Page 20
Installation Instruction Manual March 2017 LIQ-MAN-396R-396RVP Figure 2-6: Dimensional Drawing — Rosemount 396RVP with Optional 1-1/2 inch Ball Valve PN 23240-00 Installation...
Page 21
Instruction Manual Installation LIQ-MAN-396R-396RVP March 2017 Figure 2-6: Dimensional Drawing — Rosemount 396RVP with Optional 1-1/4 inch Ball Valve PN 23765-00 Installation...
Page 22
Installation Instruction Manual March 2017 LIQ-MAN-396R-396RVP Installation...
Use Rosemount custom cable Part Number 9200273 for interconnect. The maximum distance from the sensor to the transmitter is 15 ft without an integral preamplifier.
Page 24
Wiring Rosemount 396R Instruction Manual March 2017 LIQ-MAN-396R-396RVP Figure 3-2: Wiring Rosemount 396R-54 to Models 54e, 81, 3081, 4081, and 5081 pH/ORP Figure 3-3: Wiring Rosemount 396R-54 with Remote J-Box and Preamp (PN 23555-00) Wiring Rosemount 396R...
Page 25
Instruction Manual Wiring Rosemount 396R LIQ-MAN-396R-396RVP March 2017 Figure 3-4: Wiring Rosemount 396R-50 for use with Remote Junction Box (PN 23309-03) Figure 3-5: Wiring Rosemount 396R-54 for use with Remote Junction Box (PN 23309-04) Wiring Rosemount 396R...
Page 26
Wiring Rosemount 396R Instruction Manual March 2017 LIQ-MAN-396R-396RVP Figure 3-6: Wiring Rosemount 396R-50/54 to Model 1181 pH/ORP Figure 3-7: Wiring Rosemount 396R-54 to Models 1054A/B pH/ORP, 2054 pH, and 2081 pH/ORP Wiring Rosemount 396R...
Page 27
Instruction Manual Wiring Rosemount 396R LIQ-MAN-396R-396RVP March 2017 Figure 3-8: Wiring Rosemount 396R-54 to Model SCL-(P/Q) Figure 3-9: Wiring Rosemount 396R-54-61 to Model Xmt-P-XX-10 Wiring Rosemount 396R...
Page 28
Figure 3-10: Wiring Rosemount 396R-50/54-60 for use with Sensor Head J-Boxes to Models 1181, 1054 Series, 2054, 2081 Figure 3-11: Wiring Rosemount 396R-54-61 for use with Sensor Head J-Box to Models 54e, 81, 3081, 4081, and 5081 Wiring Rosemount 396R...
Page 29
Instruction Manual Wiring Rosemount 396R LIQ-MAN-396R-396RVP March 2017 Figure 3-12: Wiring Rosemount 396R-50 for use with J-Box (PN 23707-00) to Models 1181, 1050, 1060, 1030, and 1023 pH Transmitters Wiring Rosemount 396R...
Page 30
Wiring Rosemount 396R Instruction Manual March 2017 LIQ-MAN-396R-396RVP Figure 3-13: Wiring Rosemount 396R-54 for use with J-Box (PN 23708-01) to Models 1054 Series, 2054, and 2081 pH Transmitters Wiring Rosemount 396R...
Page 31
Instruction Manual Wiring Rosemount 396R LIQ-MAN-396R-396RVP March 2017 Figure 3-14: Wiring Rosemount 396R-50/54-60 for use with Sensor Head J-Boxes to Models 1181, 1054 Series, 2054, 2081 Figure 3-15: Wiring Rosemount 396R-54-61 to Model 1055-10-22-32 Wiring Rosemount 396R...
Instruction Manual Wiring Rosemount 396RVP LIQ-MAN-396R-396RVP March 2017 Section 5: Startup and Calibration Sartup To obtain best accuracy, the sensor must be calibrated as a loop with the transmitter. Please refer to the respective transmitter instruction manual for proper calibration procedures.
Startup and Calibration Instruction Manual March 2017 LIQ-MAN-396R-396RVP 5.2.2 Recommended pH Sensor Standardization For maximum accuracy, the sensor can be standardized on-line or with a process grab sample after a buffer calibration has been performed and the sensor has been conditioned to the process.
Section 6: Maintenance Maintenance Rosemount 396R and 396RVP sensors are disposable type sensors and require minimum maintenance. These sensors should be removed from the process periodically and checked in buffer solutions. If the sensor will not calibrate, refer to your transmitter instruction manual for proper test procedures.
Maintenance Instruction Manual March 2017 LIQ-MAN-396R-396RVP WARNING Before removing the sensor from the ball valve, be absolutely certain that the ball valve is fully closed. Leakage from the male connector threads may indicate that the male connector is still under pressure.
Instruction Manual Maintenance LIQ-MAN-396R-396RVP March 2017 Automatic Temperature Compensator The temperature compensator element is temperature sensitive and can be checked with an ohmmeter. Resistance increases with temperature. The 3K element will read 3000 ohms ±1% at 25°C (77°F) and a Pt-100 will read 110 ohms. Resistance...
Maintenance Instruction Manual March 2017 LIQ-MAN-396R-396RVP Sensor Tube Replacement When Used With A Sensor Head Junction Box Replace ment of the retraction versions sensor tube assembly involves the removal and installation of two sets of male connectors: One at the process end of the sensor, and the other at the junction box end (See Figures 6-1, 6-2).
Page 55
Instruction Manual Maintenance LIQ-MAN-396R-396RVP March 2017 Connect the sensor wires to the terminals on the printed circuit board in the junction box in the manner recommended on the junction box cover,and reattach the BNC connector to the preamp. Screw on the cover of the junction box aside. Discard sensor tube.
Page 56
Maintenance Instruction Manual March 2017 LIQ-MAN-396R-396RVP Figure 6-1: Example of Sensor Tube Replacement Figure 6-2: Male Connector Tightening Diagram Maintenance...
Instruction Manual Diagnostics and Troubleshooting LIQ-MAN-396R-396RVP March 2017 Section 7: Diagnostics and Troubleshooting Diagnostics and Troubleshooting with Model 54/81/3081 pH/ORP The Model 54 and 54e transmitters and Models 81 and 3081 pH Transmitters automatically search for fault conditions that would cause an error in the measured pH value, as does the Model 1054A/B pH/ORP transmitter to a lesser degree.
Diagnostics and Troubleshooting Instruction Manual March 2017 LIQ-MAN-396R-396RVP Table 7-2: Troubleshooting without Diagnostics Trouble Probable Cause Remedy Meter reads off scale. (Display Defective preamplifier. Replace preamplifier (for code 02 sensors). For reads overrange.) code 01, replace sensor. T.C. element shorted.
Instruction Manual EC Declaration of Conformity LIQ-MAN-396R-396RVP March 2017 EC Declaration of Conformity Note: Please see website for most recent Declaration. EC Declaration of Conformity...
Page 62
Diagnostics and Troubleshooting Instruction Manual March 2017 LIQ-MAN-396R-396RVP EC Declaration of Conformity...
Need help?
Do you have a question about the 396R and is the answer not in the manual?
Questions and answers