Performing a Volume Test

An obstruction in the fluidics lines might be the cause of poor reagent delivery or
sequencing results. If an obstruction in the fluidics lines is suspected, perform a volume
test.
A volume test checks the health of the fluidics system by estimating the volume between
two bubbles as they pass by the sensors. To perform a volume test, the wash tray and wash
bottle must be loaded with laboratory-grade water and a used flow cell must be in place.
Follow the onscreen prompts to perform the test.
1
Make sure that a used flow cell is loaded on the instrument.
2
From the Manage Instrument screen, select System Check.
3
Select Conduct Volume Test, and then select Next.
Figure 45
4
Fill each reservoir of the wash tray with 6 ml laboratory-grade water.
5
Fill the 500 ml wash bottle with 350 ml laboratory-grade water.
6
Load the wash tray and wash bottle onto the instrument.
a
Open the reagent compartment door and reagent chiller door, and slide the wash
tray into the reagent chiller until it stops. Close the reagent chiller door.
98
System Check Screen
Part # 15027617 Rev. H