B
PAP S
I
YNCHRONY
Error
Description of Error
Code
E74
Reserved for Future Use
E75
Incorrect Display Tape
E76
Failure in Loss of Power
Battery
E77
Invalid Error Code
E78
Low Pressure Condition
E79
Unable to Maintain
Pressure Support
E80
Barometric Pressure
Sensor Failure
E81
BarometricSensor Failure
E82
Invalid Error Code
T
& D
ROUBLESHOOTING
S
& T
ERVICE
ECHNICAL
Defective RAM chip
Incorrect LCD setup on Main PCA
Defective PCA
Reserved
Defective Motor Drive Chip
Defective Valve
Defective PCA
Excessive leak in patient circuit
Defective Valve
Defective PCA
Excessive leak in patient circuit
The raw value from the barometric
pressure sensor A/D converter is
outside its valid range of values
The barometric pressure reading
is outside its acceptable range.
Reserved
- P
24
IAGNOSTICS
AGE
I
NFORMATION
Probable Cause
Corrective Action
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Main PCA
Replace Valve
Replace Main PCA
Correct leak in
patient circuit
Replace Valve
Replace Main PCA
Correct leak in
patient circuit
Replace Main PCA
Replace Main PCA
Replace Main PCA
1032443, R
. 00
EV
Need help?
Do you have a question about the BiPAP Synchrony and is the answer not in the manual?
Questions and answers