Description of programs - Implantology - Endodontics - Surgery 9 List of errors / Troubleshooting 10 Default values Implantology 48+129 Endodontics 48+130 Surgery 48+130 11 Maintenance Servicing Information Cleaning-disinfection Important 12 Generalities and guarantee General information Terms of guarantee *only CHIROPRO L...
1 Meaning of symbols 0120 CE Marking with number of the notified body. Main switch ON - The instrument is switched on. OFF - The instrument is switched off. Fuse Ø 5 x 20 mm. Alternating current. Device of type B. CAUTION ! Dangerous voltage.
Electrical and electronic equipment may contain dangerous substances which constitute health and environmental hazards. The user must return the device to its dealer or establish direct contact with an approved body for treatment and recovery of this type of equipment (European Directive 2002/96/EC). *only CHIROPRO L...
3 Set supplied Set CHIROPRO L 1x CHIROPRO L control REF 1600613-001 REF 1700298-001 1x Micromotor MX-i LED REF 1600755-001 1x Cable for MX-i LED micromotor REF 1600606-001 1x Pack of 10 disposable sterile lines REF 1500984-010 1x 10 attachment collars for fastening the sterile irrigation line to a cable REF 1303711-010...
Contra-angle CA 20:1, without light REF 2100209 Contra-angle CA 20:1 L, with light REF 2100209 Contra-angle CA 20:1 L Micro-Series, with light REF 2100209 Contra-angle CA 20:1 L KM, light REF 2100209 Contra-angle CA 20:1 L KM Micro-Series, light REF 2100209 *only CHIROPRO L...
Page 10
Guidance and manu- CHIROPRO L/CHIROPRO is intended for use in the electromagnetic environment specified below. facturer’s declaration The customer or the user of CHIROPRO L/CHIROPRO should ensure that it is used in such an envi- - electromagnetic ronment. emissions...
Page 11
CHIROPRO L/CHIROPRO The CHIROPRO L/CHIROPRO is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The custo- mer or the user of the CHIROPRO L/CHIROPRO can help prevent electromagnetic interference by maintaining a minimum distance between por- table and mobile RF communications equipment (transmitters) and the CHIROPRO L/CHIROPRO as recommended below, according to the maxi- mum output power of the communications equipment.
6 Installation Installation Fuse A. CHIROPRO L/CHIROPRO may be positioned on a table, on a trolley or another surface, but in no circumstances on the floor. Power plug is the device for disconnection in case of problems, > ON/OFF and it must be easily accessible at all times.
Page 13
6 Installation H. Remove the single-use sterile irrigation line from its pouch. fig. 6 Fitting on the spray tube I. Connect the flexible hose of the irrigation line to the spray tube of the handpiece or contra-angle fig. 7. fig. 7 Installation on the peristaltic pump J.
MAINS CONNECTOR (100/115/230 VAC) adjusting Right key (+) Confirmation/selection key > In implantology/endodontics mode: next stage FUSE HOLDER PERISTALTIC PUMP LID MAIN SWITCH ON THE DEVICE MICROMOTOR CONNECTOR LABEL IRRIGATION ON/OFF CONTROL BUTTON BRACKET SUPPORT ON PEDAL MICROMOTOR *only CHIROPRO L...
The changes must always be confirmed with "SAVE", otherwise they will be lost. ORANGE Program key: - Short press next stage - Long press previous stage BLUE Activation/Deactivation of the peristaltic pump. GREY Activation/Deactivation of reversal of motor rotation. Motor speed control pedal. *only CHIROPRO L...
Select the system wanted Restore defaults System info and confirm with OK. Select the ratio to be changed OK, then ESC: change Restore defaults change of value with and with then save with SAVE. Continue with OK. ESC: back *only CHIROPRO L...
Page 19
Safety Avoid extended use. prevent motor overheating. Equipment initialisation error The following error may occur at start-up of CHIROPRO L/CHIROPRO 1. Check on the integrity of the CHIROPRO L/CHIROPRO memory INIT ERROR 1 The memory is corrupt! The memory data check failed.
In case of justified claim, Bien-Air Dental or its authorised Serial No. should be clearly indicated. *only CHIROPRO L...
Page 21
Default values Implantology: default values All the pre-programmed settings are indicative and MUST be validated by the user before starting the treatment/ operation. Always follow the implant manufacturer recommendations. The table shows the default operating values for the 8 implantology systems pre-programmed in the system, namely: •...
Page 23
Bien-Air Dental SA Bien-Air USA, Inc. Bien-Air UK Ltd Beijing Bien-Air Länggasse 60 Medical Technologies Arundel House Unit 1 - Ground Floor Medical Instrument Case postale 5 Corporate Park Amberley Court, Whitworth Road Technology Service Co. Ltd. 2500 Bienne 6, Switzerland Suite 160 Crawley, West Sussex, RH11 7XL, England Room 907, The Exchange Beijing,...