Instructions for use
0120
REF 2100219-0003/2014.05ENG

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CHIROPRO L and is the answer not in the manual?

Questions and answers

Summary of Contents for Bien Air CHIROPRO L

  • Page 1 CHIROPRO L / CHIROPRO Instructions for use 0120 REF 2100219-0003/2014.05ENG...
  • Page 2 Set CHIROPRO L REF 1700298-001 1x REF 1600613-001 1x REF 1600755-001 1x REF 1303711-010 1x REF 1600606-001 1x REF 1600631-001 1x REF 1303393-001 1x REF 1500984-010 Set CHIROPRO REF 1700387-001 1x REF 1600724-001 1x REF 1600825-001 1x REF 1303711-010 1x REF 1600606-001...
  • Page 3 CHIROPRO L / CHIROPRO...
  • Page 4 Summary Starting display CHIROPRO L Available values Irrigation MAIN MENU Steps Ratio Speed in rpm Torque in Ncm in ml/min Implantology Round bur 1 128:1 100 - 40’000 rpm 0.48 - 4.8 Ncm 30 ml/min 20% Endodontics Round bur 2...
  • Page 5: Table Of Contents

    Description of programs - Implantology - Endodontics - Surgery 9 List of errors / Troubleshooting 10 Default values Implantology 48+129 Endodontics 48+130 Surgery 48+130 11 Maintenance Servicing Information Cleaning-disinfection Important 12 Generalities and guarantee General information Terms of guarantee *only CHIROPRO L...
  • Page 6: Meaning Of Symbols

    1 Meaning of symbols 0120 CE Marking with number of the notified body. Main switch ON - The instrument is switched on. OFF - The instrument is switched off. Fuse Ø 5 x 20 mm. Alternating current. Device of type B. CAUTION ! Dangerous voltage.
  • Page 7: Description

    Electrical and electronic equipment may contain dangerous substances which constitute health and environmental hazards. The user must return the device to its dealer or establish direct contact with an approved body for treatment and recovery of this type of equipment (European Directive 2002/96/EC). *only CHIROPRO L...
  • Page 8: Set Supplied

    3 Set supplied Set CHIROPRO L 1x CHIROPRO L control REF 1600613-001 REF 1700298-001 1x Micromotor MX-i LED REF 1600755-001 1x Cable for MX-i LED micromotor REF 1600606-001 1x Pack of 10 disposable sterile lines REF 1500984-010 1x 10 attachment collars for fastening the sterile irrigation line to a cable REF 1303711-010...
  • Page 9: Technical Description

    Contra-angle CA 20:1, without light REF 2100209 Contra-angle CA 20:1 L, with light REF 2100209 Contra-angle CA 20:1 L Micro-Series, with light REF 2100209 Contra-angle CA 20:1 L KM, light REF 2100209 Contra-angle CA 20:1 L KM Micro-Series, light REF 2100209 *only CHIROPRO L...
  • Page 10 Guidance and manu- CHIROPRO L/CHIROPRO is intended for use in the electromagnetic environment specified below. facturer’s declaration The customer or the user of CHIROPRO L/CHIROPRO should ensure that it is used in such an envi- - electromagnetic ronment. emissions...
  • Page 11 CHIROPRO L/CHIROPRO The CHIROPRO L/CHIROPRO is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The custo- mer or the user of the CHIROPRO L/CHIROPRO can help prevent electromagnetic interference by maintaining a minimum distance between por- table and mobile RF communications equipment (transmitters) and the CHIROPRO L/CHIROPRO as recommended below, according to the maxi- mum output power of the communications equipment.
  • Page 12: Installation

    6 Installation Installation Fuse A. CHIROPRO L/CHIROPRO may be positioned on a table, on a trolley or another surface, but in no circumstances on the floor. Power plug is the device for disconnection in case of problems, > ON/OFF and it must be easily accessible at all times.
  • Page 13 6 Installation H. Remove the single-use sterile irrigation line from its pouch. fig. 6 Fitting on the spray tube I. Connect the flexible hose of the irrigation line to the spray tube of the handpiece or contra-angle fig. 7. fig. 7 Installation on the peristaltic pump J.
  • Page 14: Description Of Keys And Elements

    MAINS CONNECTOR (100/115/230 VAC) adjusting Right key (+) Confirmation/selection key > In implantology/endodontics mode: next stage FUSE HOLDER PERISTALTIC PUMP LID MAIN SWITCH ON THE DEVICE MICROMOTOR CONNECTOR LABEL IRRIGATION ON/OFF CONTROL BUTTON BRACKET SUPPORT ON PEDAL MICROMOTOR *only CHIROPRO L...
  • Page 15: Operation

    The changes must always be confirmed with "SAVE", otherwise they will be lost. ORANGE Program key: - Short press next stage - Long press previous stage BLUE Activation/Deactivation of the peristaltic pump. GREY Activation/Deactivation of reversal of motor rotation. Motor speed control pedal. *only CHIROPRO L...
  • Page 16: Start-Up

    Select the system wanted Restore defaults System info and confirm with OK. Select the ratio to be changed OK, then ESC: change Restore defaults change of value with and with then save with SAVE. Continue with OK. ESC: back *only CHIROPRO L...
  • Page 17: Pre-Settings (Setup)

    ROUND BUR 2 System info Zimmer DRILL 1 Restore defaults Apical resection Dentsply Friadent DRILL 2 Tooth extraction Biomet 3i DRILL 3 Sinus Lift Astra Tech DRILL 4 Free Thommen Medical TAPPING System TAP UNSCREWING IMPLANT SCREWING UNSCREWING *only CHIROPRO L...
  • Page 18 10:1 ENDO FILE 3 ENDO FILE 4 ENDO FILE 5 ENDO FILE 6 ENDO FILE 7 ENDO FILE 8 ENDO FILE 9 Implantology Apical resection Endodontics Tooth extraction Surgery Sinus Lift Free Pre-programmed values, see pages 129-130 *only CHIROPRO L...
  • Page 19 Safety Avoid extended use. prevent motor overheating. Equipment initialisation error The following error may occur at start-up of CHIROPRO L/CHIROPRO 1. Check on the integrity of the CHIROPRO L/CHIROPRO memory INIT ERROR 1 The memory is corrupt! The memory data check failed.
  • Page 20: Default Values

    In case of justified claim, Bien-Air Dental or its authorised Serial No. should be clearly indicated. *only CHIROPRO L...
  • Page 21 Default values Implantology: default values All the pre-programmed settings are indicative and MUST be validated by the user before starting the treatment/ operation. Always follow the implant manufacturer recommendations. The table shows the default operating values for the 8 implantology systems pre-programmed in the system, namely: •...
  • Page 22: Only Chiropro L

    250 RPM ENDO FILE 5 30.2 mNm 250 RPM ENDO FILE 6 20.2 mNm 250 RPM ENDO FILE 7 14.9 mNm 250 RPM ENDO FILE 8 14.9 mNm 250 RPM ENDO FILE 9 10.1 mNm 250 RPM *only CHIROPRO L...
  • Page 23 Bien-Air Dental SA Bien-Air USA, Inc. Bien-Air UK Ltd Beijing Bien-Air Länggasse 60 Medical Technologies Arundel House Unit 1 - Ground Floor Medical Instrument Case postale 5 Corporate Park Amberley Court, Whitworth Road Technology Service Co. Ltd. 2500 Bienne 6, Switzerland Suite 160 Crawley, West Sussex, RH11 7XL, England Room 907, The Exchange Beijing,...