ROSS and the COIL trade dress are trademarks of Thermo Fisher Scientific. AQUAfast, Cahn, EZ Flash, Ionalyzer, ionplus, KNIpHE, No Cal, ORION, perpHect, PerpHecT, PerpHecTion, pHISA, pHuture, Pure Water, Sage, Sensing the Future, SensorLink, ROSS Ultra, Sure-Flow, TEA Analyzer, Titrator PLUS, TURBO2 and Wine Master are registered trademarks of Thermo Fisher.
A . “ LO” .annunciator .in .the .LCD .alerts .you .when . the .battery .power .is .running .low . . T he .replacement . batteries .are .4 .alkaline .AAA . Caution: Power off the meter when changing battery. Thermo Scientific Russell RL060C Conductivity Meter User Guide...
You can submerge the cable for brief periods of time, but not continuously. See the Probe Care and Maintenance section for more information. Figure 6: Conductivity Electrode Thermo Scientific Russell RL060C Conductivity Meter User Guide...
Temperature Compensation section . Caution: Do not pull or force on the probe cord or the probe wires might disconnect. Note: Keep the connectors clean. Do not touch the connector with soiled hands. Thermo Scientific Russell RL060C Conductivity Meter User Guide...
Note: The factory default value for normalization temperature is 25 °C. If you need to normalize to a value other than 25 °C, see the Normalization Temperature section. Thermo Scientific Russell RL060C Conductivity Meter User Guide...
Such low conductivity standard will be contaminated as soon as it is exposed to air, therefore you must exercise caution during calibration in the first measurement range (0.00 to 20.0 μS /cm). Thermo Scientific Russell RL060C Conductivity Meter User Guide...
Note: To exit calibration without confirmation, press . ° C again to go back to measurement mode. Note: You can offset the temperature reading up to ±5 °C from the original (default) reading. Thermo Scientific Russell RL060C Conductivity Meter User Guide...
For .ATC, .make .sure .the .phono .jack .of .the .probe .(see Figure 7) .is .securely .inserted . . The .conductivity .reading .displayed .will .be .compensated .for .according .to .the . normalization .temperature .(20 .°C .or .25 .°C) .selected . .See .the .Normalization Temperature section . . Thermo Scientific Russell RL060C Conductivity Meter User Guide...
Page 22
. . T he .[CO] .indicator .will .appear .for .1 .5 .seconds, . and .the .reading .will .stop .flashing . . T he .temperature . calibration .is .successfully .performed . . T he .meter .returns . to .measurement .mode . . °C Thermo Scientific Russell RL060C Conductivity Meter User Guide...
Page 23
Note: When dipping the probe into the sample, take care to ensure that the liquid level is above its upper steel band. Stir the probe gently in the sample to create a homogenous sample. See Figure 6. Thermo Scientific Russell RL060C Conductivity Meter User Guide...
[Or] will appear on the primary display. Press until the correct range is selected. The meter resets to the auto-ranging function once it is turned off. You will have reset the manual ranging function each time you turn the meter off. Thermo Scientific Russell RL060C Conductivity Meter User Guide...
.measurements . . Note: This meter shuts off automatically after 20 minutes µ of non use. If the meter is shut off either automatically or manually, the HOLD value will be lost. Thermo Scientific Russell RL060C Conductivity Meter User Guide...
2 . . Slide .probe .guard .off .end .of .probe . . Note: Remember to re-attach the probe guard prior to taking readings. Failure to do so could result in erroneous readings. Thermo Scientific Russell RL060C Conductivity Meter User Guide...
.motor . e . Broken .probe . e . Replace .probe . Slow .response a . Dirty ./ .Oily .probe . a . Clean .probe . .See .Probe Care & Maintenance . Thermo Scientific Russell RL060C Conductivity Meter User Guide...