• There is no Transport facility on the seat cushion. • Only the pump and mattress combination as indicated by ArjoHuntleigh should be used. The correct function of the product cannot be guaranteed if incorrect pump and mattress combinations are used.
Page 6
The following instructions are important for the safe use of the battery pack and to keep the user (resident/care giver) from harm: ® • The battery pack BBP600 for the Auto Logic pump is rechargeable. For maximum battery life, charge the battery at least once every five months when the product is not in use.
Page 7
ArjoHuntleigh for information on correct disposal. Expected Service Life The Auto Logic pump has an expected service life of seven years. To maintain the condition of the pump, have the pump serviced regularly according to the schedule recommended by ArjoHuntleigh.
1. Introduction About this Manual This manual is your introduction to the Auto Logic Dual ® Mode support systems and the Aura Logic seat cushion. You must read and fully understand this manual before using the system. Use this manual to initially set up the mattress or seat cushion, keep it as a reference for day-to-day routines and as a guide to maintenance.
Page 10
Control Panel Power and Alarm Indicators Hanging Brackets Auto Logic 110 The Auto Logic 110 mattress overlay comprises the following components: Mattress Overlay Detachable Cover 3 Static Head Cells (Dark Blue) Cover Attachment Zips Cable Management CPR Unit 2 Standard Cells (Dark Blue)
Page 11
The standard cover comprises a 2-way stretch PU coated Detachable Cover knitted fabric zipped to a durable polyester base. The zips are protected by flaps to prevent ingress of contaminants and allow easy removal of the cover for cleaning. Alternative covers with advanced properties, such as are also available (Refer to Glide IS and Ventilate IS...
Page 12
Auto Logic 175 & 200 Mattress Replacement The Auto Logic 175 & 200 mattress replacements are of a similar construction to the overlay with the addition of a non-slip base. Within the non-slip base, the 175 mattress includes a foam underlay and the 200 mattress has an air filled zoned sub-mattress both of which replace the need for a mattress on the bed.
Page 13
The Aura Logic seat cushion is an Active (Alternating) Aura Logic Seat pressure redistribution system that can be used on Cushion standard hospital chairs. Detachable Cover Deflation Valves (on underside for packing & storage) Alternating Cells Non-Slip Base Serial Label (Inside Base Cover) Fixed Tubeset Fixing Straps...
2. Clinical Applications Indications The Auto Logic systems are indicated for the prevention and/or management of all categories of pressure ulcer, when combined with an individualised, comprehensive pressure ulcer protocol: for example, repositioning, nutritional support, skin care. Selection should be based upon a holistic assessment of the patient’s individual...
Remove the system from the packaging. You should have the following items: Systems for Use • Auto Logic pump including mains power cord and hanging brackets. • Auto Logic 110 mattress overlay, Auto Logic 175 200 mattress replacement or the Aura Logic seat cushion, which all have integral tubesets and covers.
1. If not already fitted, place the protective cover over the mattress. Ensure that the ArjoHuntleigh logo is uppermost and at the foot end of the mattress. 2. Zip the cover onto the mattress starting from the head end and taking care not to trap any material in the zip.
Installing the Pump 1. Position the pump, feet down, on any convenient horizontal surface or alternatively suspend from the bed foot rail by means of the swing out hooks. 2. Ensure that the mattress/seat tubeset is not “kinked” or twisted and connect it to the pump until it clicks into place.
Installing the The system should be installed as follows: Aura Logic Seat 1. Check that there are no sharp objects on the chair Cushion which may puncture the cushion. 2. Place the cushion on top of the chair surface. From a standing position in front of the chair and facing it, ensure that: •...
Page 19
8. Zip the cover onto the seat cushion, taking care not to trap any material in the zip. 9. Connect to the Auto Logic pump, refer to page 9.
4. Controls, Alarms and Indicators Controls Wait Indicator Run/Standby Low Pressure Indicator Autofirm Alarm Mute Comfort Controls Reactive (CLP) Mode Battery Low Indicator Service Indicator Power Fail Indicator Pressing the Run/Standby button will activate the pump. Run/Standby Button The green indicators on the control panel and the front of the pump will illuminate when the pump is on.
In the Reactive (CLP) mode the patient will not experience the benefits of Active (Alternating) therapy. Autofirm Mode Press and hold the Autofirm button for two seconds to enter Autofirm mode. Once activated, Autofirm creates a temporary firm surface to allow nursing procedures to be performed.
Page 22
will put the mattress into transport mode (Refer to page 17). Service Indicator The yellow Service indicator illuminates and remains on after a set number of running hours. This indicates that the pump is ready for a service. The pump will continue to function normally even when the service indicator is illuminated but you should arrange service as soon as is convenient.
Do not place the patient on the mattress until it is fully inflated and normal operating pressure has been reached. Quick Start Before using any of the Auto Logic mattress or seat systems make sure it has been installed correctly in accordance with “Installation”...
Comfort Control The mattress/seat cell pressure can be manually adjusted for patient comfort using the buttons on the pump control panel. WARNING Autofirm increases the volume of air in the mattress and therefore operation of the CPR function will take longer to deflate the mattress.
This puts the mattress into Transport mode but does not deflate the mattress. To deflate the mattress Refer to page 18. Transport Mode To transport a patient using the Auto Logic mattress, disconnect the tubeset from the pump. This automatically switches the mattress into Transport mode.
To Deflate and Store the Auto Logic Mattress To deflate the 1. Disconnect the tubeset from the pump. mattress: 2. Activate the CPR control to deflate the mattress. To store the mattress Following deflation: 1. Bring the tubeset over the mattress to lie parallel to the foot end of the mattress.
CPR Control IMPORTANT IN THE EVENT OF CARDIAC ARREST In the event of a patient suffering cardiac arrest and CPR needing to be administered: To Activate CPR Press the two CPR release buttons simultaneously. Push the front of the CPR unit until it clicks into place. To Reset CPR...
6. Battery Pack Caution The optional battery pack is not suitable for use in the homecare environment. Installing the Battery Install the battery pack to the pump as follows: Pack 1. Remove the battery socket cover from the base of the pump.
1. Once the battery pack has entered the ‘recycling’ mode it will never show that the battery is fully charged (Note: this does not indicate a fault condition). It is recommended that the battery pack is sent back to an ArjoHuntleigh service centre for disposal (refer to “Battery Storage and Disposal” on page 23).
Make sure the battery pack belongs to the pump by comparing the battery pack label with the “Technical Description” on page 27. If the battery pack type cannot be confirmed, contact ArjoHuntleigh. • Battery pack life depends on many factors. Factors that can influence battery life are: frequency of use, frequency of charging, temperature of operation, storage and storage time.
Battery Pack” on page 22), remove and store the battery pack. If the battery packs are to be put into long term storage they should be recharged at least once every three to six months. Faulty battery packs should be sent back to ArjoHuntleigh for recycling or correct disposal.
Healthcare Facility or the country of use. If you are uncertain, you should seek advice from your local Infection Control Specialist. The Auto Logic system should be routinely decontaminated between patients and at regular intervals while in use; as is good practice for all reusable medical devices.
9. Troubleshooting The following table provides a troubleshooting guide for the Auto Logic and Aura Logic systems in the event of malfunction. Indicator Possible Cause Remedy 1. The pump is inflating the 1. Both indicators will extinguish LOW PRESSURE mattress/seat.
10. Technical Description PUMP Model: Auto Logic 630001 630009KSA Part Numbers: 630009AU Australia 630003 630009IN India 100-230 V 230 V Supply Voltage: 50-60 Hz 60 Hz Supply Frequency: 28-68 VA Power Input: 375mm x 280mm x 125mm (14.8" x 11" x 5 in.) Size: 3.9kg (8.6 lb) (no battery)
PUMP SYMBOLS The operator must read this document Battery Low Do not dispose of in (Instructions for Use) Indicator domestic refuse before use. Note: This symbol is blue on the product label. With respect to electric shock, fire and Run/Standby. mechanical hazards only Note: Unit is not in accordance with...
Page 37
2030 (80") 860 (34") 115 (4.5") Auto Logic 110 (Ventilate IS PXA201ADV PXA282 2030 (80") 780 (30") 115 (4.5") Auto Logic 110 Narrow (Glide IS PXA201DAR PXA280 2030 (80") 780 (30") 115 (4.5") Auto Logic 110 Narrow (Reliant IS PXA201EVE PXA284 2030 (80")
Page 38
2. Chlorine concentrations may vary from 250ppm to 10,000ppm depending on local policy and contamination status. If an alternative disinfectant is selected from the wide variety available, ArjoHuntleigh recommend that suitability for use is confirmed with the chemical supplier prior to use.
Page 39
AUTO LOGIC 200 PRODUCTS WITH Reliant IS RF WELDED COVERS PXB001DAR-RF Auto Logic 200 Mattress with RF Welded Cover PXB080-RF Auto Logic 200 RF Welded Cover MATTRESS CLEANING SYMBOLS Recommended wash temperature: Tumble dry at 60°C (140°F) 15 min at 60°C (140°F)
Page 40
Guidance and manufacturer’s declaration - electromagnetic immunity The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should assure that it is used in such an environment. Immunity Test IEC 60601 Test Compliance Electromagnetic environment - guidance Level...
Page 41
Recommended separation distances between portable and mobile RF communications equipment and the pump The pump is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the pump can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the pump as recommended below, according to the maximum output power of the communications equipment.
Page 42
Guidance and manufacturer’s declaration – electromagnetic immunity The pump is intended for use in the electromagnetic environment specified below. The customer or the user of the pump should assure that it is used in such an environment. IEC 60601 test Electromagnetic environment Immunity test Compliance level...
Page 43
AUSTRALIA FRANCE POLSKA ArjoHuntleigh Pty Ltd ArjoHuntleigh SAS ArjoHuntleigh Polska Sp. z o.o. 78, Forsyth Street 2 Avenue Alcide de Gasperi ul. Ks Piotra Wawrzyniaka 2 O’Connor CS 70133 PL-62-052 KOMORNIKI (Pozna ) AU-6163 Western Australia FR-59436 RONCQ CEDEX Tel: +48 61 662 15 50 Tél: +33 (0) 3 20 28 13 13...
Page 44
Getinge Group is a leading global provider of products and systems that contribute to quality enhancement and cost ef ciency within healthcare and life sciences. We operate under the three brands of ArjoHuntleigh, Getinge and Maquet. Getinge provides solutions for infection control within healthcare and contamination prevention within life sciences. Maquet specializes in solutions, therapies and products for surgical interventions, interventional cardiology and intensive care.