AUTO LOGIC™ AUTOFIRM
Instructions For Use

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AUTO LOGIC and is the answer not in the manual?

Questions and answers

Summary of Contents for Arjohuntleigh AUTO LOGIC

  • Page 1 AUTO LOGIC™ AUTOFIRM Instructions For Use...
  • Page 3: Table Of Contents

    Installation ............7 Preparing the Auto Logic systems for use ......7 Installing the Auto Logic 110, 175 or the 200 Mattress .
  • Page 4 Troubleshooting ...........28 Technical Description .
  • Page 5: General Safety

    • The cover of this product is vapour permeable but not air permeable and may present a suffocation risk. • Only the pump and mattress combination as indicated by ArjoHuntleigh should be used. The correct function of the product cannot be guaranteed if incorrect pump and mattress combinations are used.
  • Page 6 Design Policy and Copyright ® and ™ are trademarks belonging to the ArjoHuntleigh group of companies. As our policy is one of continuous improvement, we reserve the right to modify designs without prior notice.
  • Page 7: Introduction

    Aura Logic™ up the mattress or seat cushion, keep it as a reference for day-to-day routines and as a guide to maintenance. About Auto Logic™ The systems comprise of a mattress Auto Logic replacement, overlay or seat cushion, all operated by the same pump.
  • Page 8 Auto Logic 110 Mattress Overlay mattress overlay comprises the Auto Logic 110 following components: Detachable Cover 3 Static Head Cells (Dark Blue) Cover Attachment Zips Cable Management CPR Unit 2 Standard Cells (Dark Blue) 10 Side Supporting Cell-in-Cells (Light Blue)
  • Page 9 Four corner retention straps are incorporated, which slide under the corners of the base mattress. Auto Logic 175 & 200 Mattress Replacement mattress replacements are of Auto Logic 175 & 200 a similar construction to the overlay with the addition of a non-slip base.
  • Page 10 Detachable Cover 3 Static Head Cells (Dark Blue) Cover Attachment Zips Cable Management CPR Unit 2 Standard Cells (Dark Blue) 10 Side Supporting Cell-in-Cells (Light Blue) 5 Standard Cells (Dark Blue) 175 - Non-slip base including foam underlay Fixed Tube-set Securing Straps 200 - Air Filled Zoned Sub-Mattress (Non-Slip) The Sub-Mattress provides an integral constant pressure...
  • Page 11 Detachable Cover Deflation Valves (on underside for packing & storage) Alternating Cells Non-Slip Base Serial Label (Inside Base Cover) Fixed Tube-set Fixing Straps The Battery Pack is an optional extra and allows the Battery Pack (Optional) pump to operate for a minimum of 8 hours without a mains power supply.
  • Page 12: Clinical Applications

    2. Clinical Applications Indications The systems are indicated for the prevention Auto Logic and management of all categories of pressure ulcer when combined with an individualised monitoring, repositioning and wound care programme. mattress is designed for patients weighing Auto Logic up to 200kg (440 lb).
  • Page 13: Installation

    CPR at the head end. The cells of the mattress must be uppermost. Caution Do not use the Auto Logic 110 mattress overlay directly on the bed frame.
  • Page 14 If you have the mattress Auto Logic 175 or 200 Auto Logic 175 or 200 Mattress Replacement replacement system, it should be installed as follows: 1. Remove the existing mattress from the bed frame and check that there are no protruding bed springs or sharp objects on the bed frame surface.
  • Page 15 1. If not already fitted, place the protective cover over the mattress. Ensure that the ArjoHuntleigh logo is uppermost and at the foot end of the mattress. 2. Zip the cover onto the mattress starting from the head end and taking care not to trap any material in the zip.
  • Page 16: Installing The Pump

    Installing the Pump The pump should be installed as follows: 1. Position the pump, feet down, on any convenient horizontal surface or alternatively suspend from the bed foot rail by means of the swing out hooks. 2. Ensure that the mattress/seat tube-set is not "kinked"...
  • Page 17 If using the mattress replacement or mattress overlay, Cable Management the power cable can be positioned in the cable management flap running down the side of the mattress as follows: 1. Locate the flap running along the mattress on the opposite side to the tubeset and CPR.
  • Page 18: Installing The Aura Logic Seat Cushion

    Installing the Aura Logic Seat cushion The system should be installed as follows: 1. Check that there are no sharp objects on the chair which may puncture the cushion. 2. Place the cushion on top of the chair surface. From a standing position in front of the chair and facing it, ensure that: •...
  • Page 19 7. Place the protective cover over the seat cushion and ensure that the ArjoHuntleigh logo and the orientation icon, printed on the cover, are uppermost and at the front of the seat.
  • Page 20: Controls, Alarms And Indicators

    4. Controls, Alarms and Indicators Controls Wait Indicator Run/Standby Low Pressure Indicator Autofirm Alarm Mute Comfort Controls Constant Low Battery Low Indicator Pressure Mode (CLP) Service Indicator Power Fail Indicator Pressing the Run/Standby button will activate the pump. Run/Standby Button The green indicators on the control panel and the front of the pump will illuminate when the pump is on.
  • Page 21: Alarms And Indicators

    Press and hold the Autofirm button for two seconds to Autofirm Mode enter Autofirm mode. Once activated, Autofirm creates a temporary firm surface to allow nursing procedures to be performed. Autofirm lasts 15 minutes, after which the pump will default back to the previous setting. When the pump is in AUTOFIRM mode the CLP and comfort control LEDs are illuminated.
  • Page 22 continue to function normally even when the service indicator is illuminated. If the yellow indicator flashes, the pump has Service detected an internal fault and a Service Engineer should be called.
  • Page 23: Operation

    Quick Start Before using any of the mattress or seat Auto Logic systems make sure it has been installed correctly in accordance with “Installation” on page 7 and ensure that the CPR unit on the mattress is clicked into the closed position.
  • Page 24: Autofirm

    Autofirm WARNING Autofirm increases the volume of air in the mattress and therefore operation of the CPR function will take longer to deflate the mattress. If there is a risk of adverse patient reactions during clinical procedure then do not use Autofirm. mode allows the mattress to be pumped to Autofirm a steady high pressure to allow nursing procedures that...
  • Page 25: To Disconnect The Tube-Set

    Refer to page 20. Transport Mode To transport a patient using the mattress, Auto Logic disconnect the tube-set from the pump. This will automatically switch the mattress into transport mode. The patient will remain supported by the mattress for up to 12 hours.
  • Page 26: To Deflate And Store The Auto Logic Mattress

    To Deflate and Store the Auto Logic Mattress 1. Disconnect the tube-set from the pump. To deflate the mattress: 2. Activate the CPR control to deflate the mattress. Following deflation: To store the mattress 1. Bring the tubeset over the mattress to lie parallel to the foot end of the mattress.
  • Page 27: Cpr Control

    CPR Control IMPORTANT IN THE EVENT OF CARDIAC ARREST In the event of a patient suffering cardiac arrest and CPR needing to be administered: Press the two CPR release buttons simultaneously. To activate CPR Push the front of the CPR unit until it clicks into place. To reset CPR...
  • Page 28: Battery Pack

    If the battery packs Disposal are to be put into long term storage they should be recharged at least once every 3-6 months. Faulty battery packs should be sent back to a ArjoHuntleigh service centre for recycling.
  • Page 29: Checking The Status Of The Battery Pack

    Checking the Status of the Battery Pack Battery Label LED 1 LED 2 Jack Socket Polarity 80-100% Charge Battery Pack OK Battery Pack 30-80% Charge needs recycling <30% Charge Battery Pack Fault Press Reset / Status Latch To check the battery pack status To check the status of the battery pack: 1.
  • Page 30: Charging The Battery Pack

    If the battery pack is on charge, via the optional battery charger, the LEDs will display one of the following conditions: LED 1 LED 2 Battery Status The Battery Pack is OK. It is fully charged. Green Green The Battery Pack is charging. Green The Battery Pack needs recycling*.
  • Page 31: Decontamination

    Healthcare Facility or the country of use. If you are uncertain, you should seek advice from your local Infection Control Specialist. system should be routinely decontaminated between Auto Logic AUTOFIRM patients and at regular intervals while in use; as is good practice for all reusable medical devices.
  • Page 32 DO NOT WRING/MANGLE, AUTOCLAVE OR USE PHENOLIC BASED SOLUTIONS. For information for the mattress top cover, including Thermal Disinfection laundering guidelines, refer to ‘Cover Specification’ table on page 31.
  • Page 33: Routine Maintenance

    Auto Logic 110, 175 & 200 and Aura Logic Seat Cushion Remove the top cover and inspect for signs of wear or General Care any tears.
  • Page 34: Troubleshooting

    9. Troubleshooting The following table provides a troubleshooting guide for the Auto Logic systems in the event of malfunction. Aura Logic Indicator Possible Cause Remedy LOW PRESSURE and WAIT The pump is inflating the Both indicators will extinguish mattress/seat. when operating pressure is reached.
  • Page 35: Technical Description

    10. Technical Description PUMP Model: Auto Logic AUTOFIRM Part Numbers: 654001 Supply Voltage: 100-240 Vac Supply Frequency: 50-60Hz Power Input: 85VA MAX Size: 375mm x 280mm x 125mm Weight: 2.75kg Case Material: ABS Plastic Plug Fuse Rating: 5A to BS1362 (UK only)
  • Page 36: Accessories

    860 (34") 115 (4.5") PXA001EVE AUTO Logic 110 (eVENT) PXA084 2030 (80") 860 (34") 115 (4.5") PXA201ADV AUTO Logic 110 Narrow (Advantex) PXA282 2030 (80") 780 (30") 115 (4.5") PXA201DAR AUTO Logic 110 Narrow (Dartex) PXA280 2030 (80") 780 (30") 115 (4.5")
  • Page 37: Seat

    SEAT Aura Logic Seat Cushion PXS001 Length: 470mm Width: 455mm Height: 50mm Cell Material: Polyurethane COVER SPECIFICATION Standard Cover ® ® Feature Advantex eVENT Fabric ® (Dartex) Removable Cover Moisture Vapour 12 times higher Permeable Air Permeable Low Friction 18% lower 20% lower Water Resistant / Repellent...
  • Page 38: Cleaning Symbols

    CLEANING SYMBOLS Wash at 80°C (176°F) Do not tumble dry above 50°C. Do not iron Wipe surface with damp cloth Do Not Use Phenol-based cleaning Use solution diluted to 1000 ppm of Available Solutions Chlorine...