Explanation Of Error Displays - Hartmann TENSOVAL COMFORT Instructions For Use Manual

Hide thumbs Also See for TENSOVAL COMFORT:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 129
If an irregular heartbeat was
detected during measure-
ment, this information
stored and displayed when recalling
the measured values from the
device memory, together with the
systolic and diastolic blood pressure
value, the pulse rate, the time, date
and year.
Deleting the stored values
You can delete all stored data in
M1 or in M2 for the respective

6. Explanation of error displays

Error which has
occurred
Device will not
turn on.
Cuff will not
inflate.
is also
Possible causes
No batteries, they have
been inserted incorrectly
or are dead.
Mains adapter not
correctly connected or
defective.
Cuff connector is
incorrectly positioned in
the connecting socket of
the device.
Wrong cuff type
connected.
person separately. To do this, press
the memory button of the corre-
sponding person. The average value
will then appear on the display.
Press down on the memory button
again and hold it down for a longer
time. After four seconds the display
will flash and after eight seconds
all data for the corresponding per-
son will be deleted and only M1 or
M2 will be displayed. If you release
the memory button ahead of time,
no data will be deleted.
Remedy
Check batteries and insert
four identical, new batter-
ies, if and when necessary.
Ensure the mains adapter
is plugged in the connect-
ing socket on the rear side
of the device.
Check connection
between the cuff
connector and the
connecting socket.
Check to make sure
that only the approved
Tensoval comfort cuffs
and the corresponding
connectors were used.
140

Hide quick links:

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the TENSOVAL COMFORT and is the answer not in the manual?

Questions and answers

Egger Attiláné
February 18, 2025

Nincs meg a kézikönyv, Err 3?

1 comments:
Mr. Anderson
February 18, 2025

Error 3 on the Hartmann TENSOVAL COMFORT indicates that the measurement signals were not recognized or were incorrectly detected. Possible causes include improper cuff placement, movement, or talking during measurement, or a weak pulse. To resolve the issue, ensure the cuff is correctly positioned, remain still and silent during measurement, and follow the device's usage guidelines.

This answer is automatically generated

Egger Attiláné
February 18, 2025

Hiba jelzés: Err 3

2 comments:
Mr. Anderson
February 18, 2025

The Err 3 error message on the Hartmann TENSOVAL COMFORT means that the measurement signals were not recognized or were incorrectly detected. This can be caused by improper cuff placement, movement, or talking during the measurement. To resolve this issue, ensure the cuff is correctly positioned, remain still, and do not talk during the measurement.

This answer is automatically generated

ALEXANDRA CHIOREAN
April 16, 2025

@Mr. Anderson how u solve error3? I mean i put it again, but it does not work anymore.