Current Xb Batch; Xb Batch Means - Beckman Coulter COULTER HmX Operator's Manual

Hematology analyzer
Hide thumbs Also See for COULTER HmX:
Table of Contents

Advertisement

Current XB Batch

Select
Sample Analysis
When XB is ON, the DMS stores the RBC
parameter results of all patient samples as
they are cycled. These results display on the
Current XB Batch screen. When a batch of
20 samples is collected, the DMS performs
XB analysis and calculates the batch mean
for MCV, MCH and MCHC. Partial
aspirations are not included.
Screen-Specific Function Keys:
Ì
Print
Prints the table.
Î
Delete Sample
Deletes a single sample from the current
batch. A maximum of 5 samples may be
deleted.
Ð
Delete Table
Deletes the entire table.
Ñ
Exit
Exits to the Main Menu.

XB Batch Means

Select
Sample Analysis tt XB tt XB Batch Means
Use to view the calculated means for each
batch of 20 samples collected. MCV, MCH
and MCHC means are calculated.
PN 4237519CB
XB
Current XB Batch
tt
tt
STARTUP AND CONTROLS
XB ANALYSIS
2-25
2

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the COULTER HmX and is the answer not in the manual?

Questions and answers

Table of Contents

Save PDF