DT-80 Rotation Stage 03.3 Ambient Conditions CONTENTS INSTALLATION ABOUT THIS DOCUMENT General Notes on Installation Objective and Target Group of this User Manual Mounting the Rotation Stage Symbols and Typographic Conventions Affixing the Load Other Applicable Documents START-UP SAFETY General Notes on Start-Up...
• Basic knowledge on servo systems, motion control concepts and The DT-80 is a laboratory device as defined by DIN EN 61010-1. It is applicable safety measures is assumed. intended for indoor use and use in an environment which is free of dirt, oil, •...
Page 5
Improper use of the DT-80 may result in personal injury and/or damage to the DT-80. 2.2.2 Measures during Installation Only use the DT-80 for its intended purpose, and only use it if it is in good The DT-80 may be damaged by excessively long screws and wrongly working order.
2.2.4 Measures during Operation NOTICE • If noise during occurs during operation of the DT-80, check the settings All specifications in this user manual refer only to the of the servo-control parameters of your controller. standard products that are included in the PI-miCos catalog.
DT-80 Rotation Stage 04.7 Safety Instructions PRODUCT DESCRIPTION Features and Application Area Our products are designed specifically for use in the laboratory. Model Overview 6443-9- Order no. DC-B-029 2Phase-041 Product View NOTICE Modifications and servicing on this axis may only be carried out by the manufacturer or persons authorized by the manufacturer.
04.8 DT-80 Rotation Stage Torque constant mNm/A 41.3 Velocity constant rpm/V WARNING n/M slope curve rpm/mNm It is recommended that all persons entrusted with working No load velocity 5300 with this product and who therefore come into contact with Max.continuous velocity at...
DT-80 Rotation Stage 04.9 2Phase-041 Encoder supply voltage DC brush motor + Motor type 2 phase bipolar SH4018L1704 DC brush motor - Phase current Limit forward Step angle ° 1.8 ° Limit reverse Steps LCOM Limit common Coil resistance...
For indoor use only. Prerequisite • The DT-80 was calibrated at an ambient temperature of 20 °C (+/- 3 °C). You have read and understood the general notes on installation (see • The permissible operating temperature is between 20 °C and 40 °C.
M6 DIN 912 3.0 Nm MAINTENANCE Mounting the Additional Part Depending on the operating conditions and the period of use of the DT-80, Select the mounting position so that the existing fixing holes in the slider of the following maintenance measures are required: the DT-80 can be used for the additional part to be affixed.
If the problem that occurred with your system is not listed in the table above or cannot be solved as described, contact our customer service department (see chapter 9). Any old PI miCos equipment can be sent free of charge to the following address: CUSTOMER SERVICE...
04.14 DT-80 Rotation Stage 11. EU DECLARATION OF CONFORMITY An EC Declaration of Conformity has been issued for the DT-80 in accordance with the following European directives: 2004/108/EC, EMC Directive 2011/65/EU, RoHS Directive The applied standards certifying the conformity are listed below.
Need help?
Do you have a question about the dt-80 and is the answer not in the manual?
Questions and answers