Application; Protein Testing - Miele PG 8527 Operating Instructions Manual

Washer-disinfector
Table of Contents

Advertisement

Application

The washer-disinfector can be fitted
with various mobile units, adapted to
the type and shape of the items that
need to be cleaned and disinfected;
they can be equipped with a wide
variety of modules and inserts.
Mobile units, modules and inserts
should be properly selected to
match the application involved.
On the following pages you will find
loading examples and additional
information.
Before EVERY program start,
conduct the following visual
inspection:
– Are the items properly sorted, loaded
and connected in the
washer-disinfector?
– Are the spray arms clean and can
they rotate freely?
– Is the flat filter free of debris (remove
any coarse material and clean the
filter as needed)?
– Is the flat filter horizontally aligned
with the wash chamber and is the
seal flush with the wash chamber?
– Is the adapter properly connected to
the water supply for the spray arms
and nozzles?
– Are the process chemical dispensers
full (detergent, neutralizer, etc.)?
28
After EVERY program completion,
conduct the following visual
inspection:
– Inspect the labware or instruments
for cleanliness.
– Have any hollow items been
dislodged from their proper position
on the nozzles?
,
Any items that were dislodged
from the adapters during a program
cycle must be washed again.
– Are the cavities and channels in
hollow items cleaned through?
– Are the nozzles and connections
firmly attached to the mobile
units/inserts?

Protein testing

After washing, items used in medical
applications should be checked at
random for proteins.